Puritan Bennett 980X3ENDIUURecalls & reported problems
Covidien LP · 980 Ventilator Compressor Configuration · Class 2
All models of Puritan Bennett (119)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Puritan Bennett 980 ventilator is a breathing machine. Its 826 reports include 3 deaths and 2 injuries, with many describing improper flow or output problems. Reports fell to 16 in the last 12 months, down from 84 in the 12 months before. 2 recalls name this device: 1 open, about alarms that may not sound or show, and 1 closed, about a capacitor assembled incorrectly.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 827
- Deaths reported
- 3
- Injuries reported
- 2
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 827 adverse event reports that list this exact device (DI 10884521172524), received between Aug 24, 2017 and Sep 28, 2026 — classified as 3 deaths, 2 injuries, 822 malfunctions.
The most frequently reported specific device problem is “Improper Flow or Infusion” (580 reports).
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 25, 2022Class IOpen, Classified
The audible alarm may not sound and/or the omni-directional LED visual alarm may not display during alarm states. Delayed awareness of alarm states may lead to delayed response or a delay of treatment, potentially resulting in hypoxia, dyspnea, cardiac arrest, or death.
Z-0966-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Nov 4, 2021Class ITerminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
Z-0352-2022 · Root cause: Reprocessing Controls · Recall details · FDA record ↗
Most reported problems
Device problems
- Improper Flow or Infusion580
- Output Problem445
- No Flow165
- Smoking31
- Defective Alarm14
- Device Inoperable14
- Battery Problem7
- Power Problem6
Patient problems
- No Clinical Signs, Symptoms or Conditions667
- No Consequences Or Impact To Patient67
- No Code Available51
- No Patient Involvement29
- Insufficient Information13
- Low Oxygen Saturation6
- No Information6
- No Known Impact Or Consequence To Patient5
Recent adverse event reports
The 25 most recent of 827 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Smoking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50926498 ↗MalfunctionJun 18, 2026Output Problem
IT WAS REPORTED THAT THE 980 VENTILATOR DID NOT PASS THE SHORT SELF TEST (SST) AND EXTENDED SELF TEST (EST) FOR THE AUTO ZERO TRANSDUCER ON THE EXHALATION SIDE. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25546589 ↗MalfunctionMay 27, 2026Improper Flow or Infusion
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THIS 980 VENTILATOR WAS IN AN INOPERATIVE STATUS. THERE WAS NO REPORTED INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25300335 ↗MalfunctionMay 14, 2026Mechanical Problem; No Flow
IT WAS REPORTED THAT DURING ANNUAL PREVENTATIVE MAINTENANCE PERFORMED BY MEDTRONIC, THIS 980 VENTILATOR WAS FOUND TO HAVE DAMAGE TO THE DIRECT CURRENT (DC) - DC PRINTED CIRCUIT BOARD ASSEMBLY (PCBA). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25178428 ↗MalfunctionApr 29, 2026Improper Flow or Infusion
IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, THIS 980 VENTILATOR GENERATED A BACKGROUND ALERT INDICATING THE VENTILATOR ENTERED INTO BACKUP VENTILATION. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITH NO INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25032335 ↗MalfunctionApr 14, 2026Improper Flow or Infusion
IT WAS REPORTED THAT THE 980 VENTILATOR GENERATED A MIX AIR FLOW OUT OF RANGE BACKGROUND DIAGNOSTIC CODE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24877893 ↗MalfunctionMar 31, 2026Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT THE 980 VENTILATOR GENERATED A MESSAGE INDICATING THAT A EXTENDED SELF TEST (EST) WAS REQUIRED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITH NO INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24744959 ↗MalfunctionMar 24, 2026No Flow
IT WAS REPORTED THAT WHILE SETTING UP THE 980 VENTILATOR TO RUN SST, THE VENTILATOR WENT INOPERATIVE WITH DIAGNOSTIC CODES INDICATING AN ISSUE WITH THE DIRECT CURRENT (DC)-DC CONVERTER PRINTED CIRCUIT BOARD ASSEMBLY (PCBA). THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24681492 ↗MalfunctionMar 18, 2026Improper Flow or Infusion
IT WAS REPORTED THAT THIS 980 VENTILATOR DID NOT PASS THE EXHALATION VALVE (EV) PERFORMANCE TEST PORTION OF THE SHORT SELF TEST (SST) WITH A MESSAGE INDICATING THE PRESSURE WAS OUT OF RANGE (OOR). THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24627348 ↗MalfunctionMar 5, 2026No Flow
IT WAS REPORTED THAT THE 980 VENTILATOR WOULD NOT STAY POWERED ON WHILE OPERATING ON BATTERY POWER AND WOULD SHUT OFF AS SOON AS IT WAS UNPLUGGED FROM THE WALL OUTLET. THERE WAS NO PATIENT INVOLVED. HOWEVER, A REVIEW OF THE VENTILATOR LOGS INDICATE THAT THERE WAS A LOSS OF VENTILATION (LOV) WHILE IN USE. THERE WAS NO REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24521551 ↗MalfunctionFeb 27, 2026Smoking; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THIS 980 VENTILATOR'S SCREEN WENT BLACK, THEN WHEN THE SCREEN CAME ON CLINICIANS SAW SMOKE COMING FROM THE BACK OF THE VENTILATOR. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24475247 ↗MalfunctionJan 29, 2026No Flow
IT WAS REPORTED THAT THIS 980 VENTILATOR WAS IN AN INOPERATIVE CONDITION. THERE WAS NO PATIENT INVOLVED. A REVIEW OF THE VENTILATOR LOGS SHOW A LOSS OF VENTILATION OCCURRING AFTER AN ATTEMPT TO ENTER INTO BACKUP VENTILATION (BUV) WHICH APPEARS TO HAVE OCCURRED DURING PATIENT USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24210750 ↗MalfunctionJan 28, 2026Improper Flow or Infusion
IT WAS REPORTED THAT THIS 980 VENTILATOR WAS IN AN INOPERATIVE CONDITION. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: Insufficient Information
MAUDE report 24203368 ↗MalfunctionJan 8, 2026Improper Flow or Infusion
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT THIS 980 VENTILATOR "SWITCHED OFF". THERE WAS NO REPORTED PATIENT INJURY. AFTER A REVIEW OF THE VENTILATOR LOGS THERE WAS NO EVIDENCE THE VENTILATOR SHUT DOWN DURING USE. THERE WAS HOWEVER, EVIDENCE IN THE VENTILATOR LOGS THAT THE VENTILATOR ENTERED INTO BACKUP VENTILATION (BUV).
Patient problems: Insufficient Information
MAUDE report 24011925 ↗MalfunctionDec 17, 2025Improper Flow or Infusion; Improper Flow or Infusion
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THIS 980 VENTILATOR GENERATED ERROR MESSAGES. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23834726 ↗MalfunctionNov 26, 2025Improper Flow or Infusion; Improper Flow or Infusion
IT WAS REPORTED THAT DURING USE OF THE 980 VENTILATOR ON A PATIENT, A BACKGROUND FAULT WAS DETECTED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23657361 ↗MalfunctionNov 20, 2025Device Emits Odor; No Flow; Output Problem; Noise, Audible; Device Emits Odor; No Flow; Output Problem; Noise, Audible
IT WAS REPORTED THAT THE 980 VENTILATOR MADE A LOUD NOISE, HAD A BURNT SMELL, AND DID NOT PASS THE "MIX SUBSYSTEM TEST", "INSPIRATORY SUBSYSTEM TEST" AND EXHALATION SUBSYSTEM TEST" DURING THE POWER ON SELF TEST (POST). THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23607581 ↗MalfunctionOct 30, 2025Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THE 980 VENTILATOR DID NOT PASS THE SHORT SELF TEST (SST). THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23430204 ↗MalfunctionOct 20, 2025Improper Flow or Infusion; Improper Flow or Infusion
IT WAS REPORTED THAT THIS 980 VENTILATOR DID NOT PASS THE INSPIRATORY AUTOZERO OFFSET DURING THE EXTENDED SELF TEST (EST). THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23343068 ↗MalfunctionOct 16, 2025Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THE 980 VENTILATOR DID NOT PASS THE EXTENDED SELF TEST (EST) AND SHORT SELF TEST (SST). THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23318569 ↗MalfunctionOct 16, 2025Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THE EXTENDED SELF TEST (EST) PERFORMED ON THE 980 VENTILATOR DID NOT PASS, DISPLAYING THE MESSAGE "INSPIRATORY AUTOZERO SOLENOID NOT OPERATIONAL." THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23317395 ↗MalfunctionOct 15, 2025No Flow; Output Problem; No Flow; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR DID NOT PASS THE SHORT SELF TEST (SST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23306188 ↗MalfunctionOct 15, 2025Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THE 980 VENTILATOR FAILED THE CIRCUIT PRESSURE TEST WITH AN "INSPIRATORY AUTO ZERO SOLENOID NOT OPERATIONAL" MESSAGE, DURING THE EXTENDED SELF TEST (EST) AND SHORT SELF TEST (SST). THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23286526 ↗MalfunctionSep 19, 2025Improper Flow or Infusion; Improper Flow or Infusion
IT WAS REPORTED THAT DURING THE PREVENTIVE MAINTENANCE OF THIS 980 VENTILATOR, THE VENTILATOR DID NOT PASS THE EXTENDED SELF TEST (EST) AS THE INSPIRATORY SOLENOID WAS READING OUT OF RANGE (OOR) DURING THE CIRCUIT PRESSURE TEST. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23100313 ↗MalfunctionSep 9, 2025Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THE 980 VENTILATOR'S AUTO ZERO SOLENOID WAS NOT OPERATIONAL AND THE VENTILATOR DID NOT PASS THE EXTENDED SELF TEST (EST) OR THE SHORT SELF TEST (SST). THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23011821 ↗
Other VENTILATOR, CONTINUOUS, FACILITY USE devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record