Puritan Bennett G-760110DIUU-JPRecalls & reported problems
Covidien LP · 700 Series Ventilator · Class 2
All models of Puritan Bennett (119)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 10884521180918), received between Jul 3, 2018 and Aug 3, 2021 — classified as 19 malfunctions.
The most frequently reported specific device problem is “No Flow” (17 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- No Flow17
- Output Problem2
- Defective Alarm1
- Defective Component1
- Failure to Power Up1
- Infusion or Flow Problem1
Patient problems
- No Consequences Or Impact To Patient8
- No Code Available7
- No Clinical Signs, Symptoms or Conditions6
- No Patient Involvement3
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 3, 2021No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE 760 VENTILATOR GENERATED AN ALARM, STOPPED WORKING, FAILED TO DELIVER BREATHS AND THE SAFETY VALVE OPENED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND WAS PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY. ADDITIONALLY, WHEN THE POWER WAS TURNED ON AGAIN THE DEVICE DISPLAYED THE MESSAGE "POWER ON SELF-TEST (POST) FAILED".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12268893 ↗MalfunctionMay 24, 2021No Flow; Output Problem; No Flow; Output Problem
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THIS 760 VENTILATOR GENERATED A SAFETY VALVE OPEN ALARM AND VENTILATION STOPPED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITHOUT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11872534 ↗MalfunctionJan 13, 2021No Flow; No Flow
IT WAS REPORTED THAT DURING MAINTENANCE, THE 760 VENTILATOR FAILED THE POWER ON SELF TEST (POST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11167226 ↗MalfunctionJan 7, 2021Infusion or Flow Problem; No Flow; Infusion or Flow Problem; No Flow
IT WAS REPORTED THAT DURING MAINTENANCE, THE 760 VENTILATOR GENERATED AN OXYGEN(O2) SUPPLY LOW PRESSURE ALARM EVEN THOUGH THE O2 PIPE WAS CONNECTED. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11136152 ↗MalfunctionJul 10, 2020No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE 760 VENTILATOR GENERATED AN ALARM AND THE SAFETY VALVE OPENING INDICATOR WAS FOUND TO BE ILLUMINATED. IT WAS REPORTED THAT THE VENTILATION STOPPED. THE PATIENT WAS PLACED ON AN ALTERNATE VENTILATOR AND WAS NOT HARM AS RESULT OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10260962 ↗MalfunctionJun 29, 2020No Flow; No Flow
IT WAS REPORTED THAT DURING USE ON A PATIENT, A 760 VENTILATOR GENERATED AN ALARM AND DISPLAYED POST (POWER ON SELF TEST) FAILURE. MANUAL VENTILATION HAD TO BE PERFORMED DUE TO LOSS OF VENTILATION AND THE PATIENT WAS TRANSFERRED TO AN ALTERNATE VENTILATOR WITHOUT ANY HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10206720 ↗MalfunctionMar 17, 2020No Flow; Output Problem; No Flow; Output Problem
IT WAS REPORTED THAT DURING SERVICING, A 760 VENTILATOR FAILED ITS POWER ON SELF TEST (POST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Patient Involvement
MAUDE report 9845686 ↗MalfunctionOct 31, 2019No Flow; No Flow
IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, A 760 VENTILATOR GENERATED AN ALARM AND THE SAFETY VALVE OPENED, WHICH STOPPED VENT ILLATION. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITH NO HARM OR INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9260605 ↗MalfunctionOct 25, 2019No Flow; No Flow
IT WAS REPORTED WHILE IN USE THIS 760 VENTILATOR ALARMED AND VENTILATION STOPPED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITH NO HARM REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9241554 ↗MalfunctionSep 23, 2019No Flow; No Flow
IT WAS REPORTED THAT DURING USE ON A PATIENT, A 760 VENTILATOR STOPPED VENTILATION AND SAFETY VALVE WENT OPEN WITH AN ALARM. PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITH NO HARM REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9106451 ↗MalfunctionAug 19, 2019No Flow; No Flow
IT WAS REPORTED THAT DURING USE ON A PATIENT, THE SAFETY VALVE OPENED AND VENTILATION STOPPED. THE PATIENT WAS TRANSFERRED TO AN ALTERNATE VENTILATOR WITH NO PATIENT HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8908051 ↗MalfunctionFeb 11, 2019No Flow; No Flow
IT WAS REPORTED DURING USE ON A PATIENT, THE 760-VENTILATOR GENERATED A LOW OXYGEN SUPPLY PRESSURE ALARM APPROXIMATELY FIFTEEN MINUTES AFTER THE CANNULA WAS CHANGED. ADDITIONALLY, THE POWER ON SELF-TEST FAILED, SAFETY VALVE OPENED, AND VENTILATION STOPPED. MANUAL VENTILATION WAS IMMEDIATELY PERFORMED AND THE PATIENT WAS THEN TRANSFERRED TO AN ALTERNATE VENTILATOR WITHOUT HARM OR INJURY.
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 8324981 ↗MalfunctionNov 6, 2018Defective Alarm; Defective Alarm
IT WAS REPORTED THAT THE 700 SERIES VENTILATOR FAILED POST (POWER ON SELF TEST) WITH AN ERROR MESSAGE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Patient Involvement
MAUDE report 8044522 ↗MalfunctionOct 22, 2018No Flow; No Flow
IT WAS REPORTED THAT DURING USE ON A PATIENT, THE 740 VENTILATOR GENERATED AN ALARM AND VENTILATION STOPPED. THE PATIENT WAS TRANSFERRED TO AN ALTERNATE VENTILATOR WITH NO PATIENT HARM.
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 7992856 ↗MalfunctionSep 27, 2018No Flow; No Flow
IT WAS REPORTED THAT DURING USE ON PATIENT THE 760 VENTILATOR HAD GENERATED VENTILATION FAILURE ALARM. THE DEVICE TURNED ON AGAIN THEN POST(POWER ON SELF TEST) FAILURE OCCURRED WITH ERROR CODES. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITH NO HARM OR INJURY.
Patient problems: No Code Available
MAUDE report 7916047 ↗MalfunctionAug 31, 2018Failure to Power Up; No Flow; Failure to Power Up; No Flow
ADDITIONAL INFORMATION RECEIVED AS FOLLOWS: IT WAS REPORTED THAT DURING SETTING UP THE 700 SERIES VENTILATOR PASSED POST (POWER ON SELF-TEST) WITH AN ERROR MESSAGE "IT COULD NOT BE USED". THE PATIENT WAS NOT ATTACHED TO THE VENTILATOR AT THE TIME OF THE EVENT. IMMEDIATELY ANOTHER VENTILATOR WAS PREPARED, AND IT WAS USED. NO HARM WAS REPORTED.
Patient problems: No Code Available; No Patient Involvement
MAUDE report 7839994 ↗MalfunctionAug 24, 2018No Flow; No Flow
IT WAS REPORTED THAT DURING USE ON A PATIENT, THE SAFETY VALVE OPENED AND A ERROR MESSAGE WAS GENERATED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR.
Patient problems: No Code Available
MAUDE report 7814465 ↗MalfunctionAug 13, 2018No Flow; No Flow
IT WAS REPORTED THAT DURING USE ON PATIENT THE 700 SERIES VENTILATOR SUDDEN SHUTDOWN. MANUAL VENTILATION WAS PERFORMED. THERE WAS ONLY ONE DEVICE THE CUSTOMER COULD USE SO THE TECHNICIAN RECHARGED THE BATTERY. THE VENTILATOR PASSED THE BATTERY LOAD TEST AFTER CHARGING. THE DEVICE IS CONTINUOUSLY BEING USED. NO PATIENT HARM WAS REPORTED.
Patient problems: No Code Available
MAUDE report 7778999 ↗MalfunctionJul 3, 2018Defective Component; Defective Component
IT WAS REPORTED THAT DURING USE ON PATIENT 760 VENTILATOR HAD POST FAILED. THE DEVICE WAS REPLACED. NO PATIENT HARM REPORTED.
Patient problems: No Code Available
MAUDE report 7661179 ↗
Other VENTILATOR, CONTINUOUS, FACILITY USE devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page