Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884521181014), received between Apr 25, 2017 and Aug 14, 2017 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Device Alarm System” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Device Alarm System1
  • Insufficient Device Problem Information1
  • Tidal Volume Fluctuations1

Patient problems

  • No Information2
  • No Patient Involvement2
  • No Code Available1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 14, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT AN 840 VENTILATOR HAD AN OXYGEN SENSOR ISSUE. ALTHOUGH REQUESTED, IT IS UNKNOWN IF THE PATIENT WAS CONNECTED TO THE VENTILATOR AT THE TIME OF THE EVENT. THE SERVICE ENGINEER REPLACED THE OXYGEN SENSOR TO ADDRESS THE ISSUE. THE UNIT PASSED ALL TESTING AND OPERATED WITHIN THE MANUFACTURING SPECIFICATIONS.

    Patient problems: No Information; No Patient Involvement

    MAUDE report 6792181 ↗
  • MalfunctionJul 31, 2017Tidal Volume Fluctuations; Tidal Volume Fluctuations

    ADDITIONAL INFORMATION WAS RECEIVED THAT THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Information; No Patient Involvement

    MAUDE report 6755829 ↗
  • InjuryApr 25, 2017Device Alarm System; Device Alarm System

    IT WAS REPORTED THAT, AN 840 VENTILATOR SYSTEM ERROR ALARM WHILE IN USE ON A PATIENT. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE SERVICE ENGINEER (SE) VERIFIED THE MALFUNCTION AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB) AND UPGRADED THE TWO INVERTER BACKLIGHTS. THE UNIT PASSED ALL TESTING AND OPERATES WITHIN THE MANUFACTURING SPECIFICATIONS.

    Patient problems: No Code Available; No Consequences Or Impact To Patient

    MAUDE report 6518514 ↗

Other VENTILATOR, CONTINUOUS, FACILITY USE devices

All VENTILATOR, CONTINUOUS, FACILITY USE devices →