Skip to content

Puritan Bennett 4-076326-00Recalls & reported problems

Covidien LP · 800 Series Ventilator Compressor,Model 806 · Class 2

All models of Puritan Bennett (119)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884521181168), received between Aug 22, 2017 and Mar 13, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Smoking” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Smoking2
  • Calibration Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • No Information1
  • No Patient Involvement1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 13, 2026Smoking

    IT WAS REPORTED THAT THERE WAS A SMOKE COMING FROM THE 806 COMPRESSOR OF THIS 840 VENTILATOR. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24590414 ↗
  • MalfunctionFeb 3, 2020Smoking; Smoking

    IT WAS REPORTED THAT DURING USE THE 840 VENTILATOR COMPRESSOR HAD SMOKE COMING OUT AND THE CIRCUIT BREAKER TRIPPED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITHOUT HARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9662149 ↗
  • MalfunctionAug 22, 2017Calibration Problem; Calibration Problem

    THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE EVENT. THE OXYGEN O2 SENSOR WAS REPLACED AND THE VENTILATOR WAS FOUND TO BE IN WORKING CONDITION.

    Patient problems: No Information; No Patient Involvement

    MAUDE report 6811637 ↗

Other VENTILATOR, CONTINUOUS, FACILITY USE devices

All VENTILATOR, CONTINUOUS, FACILITY USE devices →

Sources and updates