Skip to content

Versaport ONB11LGFRecalls & reported problems

Covidien LP · Bladeless Optical Trocar with Fixation Cannula · Class 2

All models of Versaport (21)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10884521181380), received between Jul 9, 2019 and Apr 14, 2022 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Detachment of Device or Device Component1
  • Leak/Splash1
  • Material Split, Cut or Torn1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 14, 2022Break; Leak/Splash; Adverse Event Without Identified Device or Use Problem; Material Split, Cut or Torn; Break; Leak/Splash; Adverse Event Without Identified Device or Use Problem; Material Split, Cut or Torn

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC PYELOTOMY PROCEDURE, WHEN CHANGING THE DOOR ANGLE TO DISSECTION CLAMP, THE TROCAR BEGAN TO SPILL OUT. WHEN OBSERVED, THE MEMBRANE WAS BROKEN. THE PRODUCT WAS CHANGED TO COMPLETE THE CASE AND RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14112500 ↗
  • MalfunctionJul 9, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC GASTRIC BYPASS PROCEDURE WHEN DOING THE SURGICAL STAPLING, THE SEAL OF THE TROCAR WAS DISENGAGED AND THE RUBBER ENDED UP BURSTING. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8773703 ↗