Puritan Bennett 980S3ESDIUARecalls & reported problems
Covidien LP · 980 Pediatric-Adult Ventilator DC Compressor Configuration · Class 2
All models of Puritan Bennett (119)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Mar 25, 2022Class IOpen, Classified
The audible alarm may not sound and/or the omni-directional LED visual alarm may not display during alarm states. Delayed awareness of alarm states may lead to delayed response or a delay of treatment, potentially resulting in hypoxia, dyspnea, cardiac arrest, or death.
Z-0966-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Oct 16, 2020Class IIOpen, Classified
In-house lab observations and customer reports of overheating and/or thermal damage to the battery pack, create the risk of exposure to smoke or hazardous materials, which may result in burns and/or bronchospasm,
Z-0425-2021 · Root cause: Device Design · Recall details · FDA record ↗
Other VENTILATOR, CONTINUOUS, FACILITY USE devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page