Puritan Bennett 980U3ESDIUURecalls & reported problems

Covidien LP · 980 Universal Ventilator DC Compressor Configuration · Class 2

All models of Puritan Bennett (119)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
2

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10884521201545), received between Jun 7, 2017 and Feb 4, 2026 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Device Inoperable” (3 reports).

2 FDA recall records name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 25, 2022Class IOpen, Classified

    The audible alarm may not sound and/or the omni-directional LED visual alarm may not display during alarm states. Delayed awareness of alarm states may lead to delayed response or a delay of treatment, potentially resulting in hypoxia, dyspnea, cardiac arrest, or death.

    Z-0966-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Nov 4, 2021Class ITerminated

    Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.

    Z-0344-2022 · Root cause: Reprocessing Controls · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Inoperable3
  • No Flow2
  • Smoking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Code Available2
  • No Information2
  • No Patient Involvement1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 4, 2026No Flow

    IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THIS 980 VENTILATOR HAD A LOSS OF VENTILATION. THE PATIENT WAS MANUALLY VENTILATED USING AN ARTIFICIAL MANUAL BREATHING UNIT (AMBU) BAG. THERE WAS NO INJURY TO THE PATIENT. A REVIEW OF THE VENTILATOR LOGS INDICATE THE VENTILATOR EXPERIENCED A LOSS OF VENTILATION AFTER AN ATTEMPT TO ENTER INTO BACKUP VENTILATION (BUV).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24259096 ↗
  • MalfunctionNov 12, 2021No Flow; No Flow

    IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, THIS 980 VENTILATOR ¿SHUT DOWN¿. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR. IT WAS ADDITIONALLY REPORTED THAT THE PATIENT WAS IN POOR HEALTH AND DIED LATER DUE TO THEIR "SERIOUS CONDITION". THE HOSPITAL STAFF REPORTED THAT THE VENTILATOR DID NOT CAUSE OR CONTRIBUTE TO THE PATIENT DEATH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12804618 ↗
  • MalfunctionOct 4, 2021Smoking; Smoking

    IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, THIS 980 VENTILATOR EMITTED SMOKE FROM THE BATTERY AREA AND SHUT DOWN. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND MANUALLY VENTILATED VIA AN AMBU BAG BEFORE BEING PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12568094 ↗
  • MalfunctionJul 26, 2017Device Inoperable; Device Inoperable

    IT WAS REPORTED THAT, THE COMPRESSOR ON A 980 VENTILATOR WAS INOPERABLE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE EVENT OCCURRED. THE SERVICE ENGINEER (SE) INSPECTED THE DEVICE AND VERIFIED THE REPORTED CONDITION. THE SE RECONNECTED ALL OF THE ELECTRICAL CONNECTIONS FROM THE COMPRESSOR AND THE UNIT RESUMED NORMAL OPERATION. THE SE PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED.

    Patient problems: No Patient Involvement

    MAUDE report 6745155 ↗
  • InjuryJun 7, 2017Device Inoperable; Device Inoperable

    ADDITIONAL INFORMATION WAS RECEIVED THAT THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR.

    Patient problems: No Code Available; No Information

    MAUDE report 6620565 ↗
  • InjuryJun 7, 2017Device Inoperable; Device Inoperable

    IT WAS REPORTED THAT A 980 VENTILATOR WENT INTO INOPERABLE CONDITION AND STOPPED CYCLING WHILE IN USE ON A PATIENT. ALTHOUGH REQUESTED, INFORMATION REGARDING THE INTERVENTION TAKEN ON THE BEHALF OF THE PATIENT HAS NOT BEEN PROVIDED. THE CUSTOMER REPORTED THERE WAS NO INJURY TO THE PATIENT.

    Patient problems: No Code Available; No Information

    MAUDE report 6620449 ↗

Other VENTILATOR, CONTINUOUS, FACILITY USE devices