Puritan Bennett 980U3ESDIUURecalls & reported problems
Covidien LP · 980 Universal Ventilator DC Compressor Configuration · Class 2
All models of Puritan Bennett (119)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10884521201545), received between Jun 7, 2017 and Feb 4, 2026 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Device Inoperable” (3 reports).
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 25, 2022Class IOpen, Classified
The audible alarm may not sound and/or the omni-directional LED visual alarm may not display during alarm states. Delayed awareness of alarm states may lead to delayed response or a delay of treatment, potentially resulting in hypoxia, dyspnea, cardiac arrest, or death.
Z-0966-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Nov 4, 2021Class ITerminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
Z-0344-2022 · Root cause: Reprocessing Controls · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Inoperable3
- No Flow2
- Smoking1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Code Available2
- No Information2
- No Patient Involvement1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 4, 2026No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THIS 980 VENTILATOR HAD A LOSS OF VENTILATION. THE PATIENT WAS MANUALLY VENTILATED USING AN ARTIFICIAL MANUAL BREATHING UNIT (AMBU) BAG. THERE WAS NO INJURY TO THE PATIENT. A REVIEW OF THE VENTILATOR LOGS INDICATE THE VENTILATOR EXPERIENCED A LOSS OF VENTILATION AFTER AN ATTEMPT TO ENTER INTO BACKUP VENTILATION (BUV).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24259096 ↗MalfunctionNov 12, 2021No Flow; No Flow
IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, THIS 980 VENTILATOR ¿SHUT DOWN¿. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR. IT WAS ADDITIONALLY REPORTED THAT THE PATIENT WAS IN POOR HEALTH AND DIED LATER DUE TO THEIR "SERIOUS CONDITION". THE HOSPITAL STAFF REPORTED THAT THE VENTILATOR DID NOT CAUSE OR CONTRIBUTE TO THE PATIENT DEATH.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12804618 ↗MalfunctionOct 4, 2021Smoking; Smoking
IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, THIS 980 VENTILATOR EMITTED SMOKE FROM THE BATTERY AREA AND SHUT DOWN. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND MANUALLY VENTILATED VIA AN AMBU BAG BEFORE BEING PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12568094 ↗MalfunctionJul 26, 2017Device Inoperable; Device Inoperable
IT WAS REPORTED THAT, THE COMPRESSOR ON A 980 VENTILATOR WAS INOPERABLE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE EVENT OCCURRED. THE SERVICE ENGINEER (SE) INSPECTED THE DEVICE AND VERIFIED THE REPORTED CONDITION. THE SE RECONNECTED ALL OF THE ELECTRICAL CONNECTIONS FROM THE COMPRESSOR AND THE UNIT RESUMED NORMAL OPERATION. THE SE PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED.
Patient problems: No Patient Involvement
MAUDE report 6745155 ↗InjuryJun 7, 2017Device Inoperable; Device Inoperable
ADDITIONAL INFORMATION WAS RECEIVED THAT THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR.
Patient problems: No Code Available; No Information
MAUDE report 6620565 ↗InjuryJun 7, 2017Device Inoperable; Device Inoperable
IT WAS REPORTED THAT A 980 VENTILATOR WENT INTO INOPERABLE CONDITION AND STOPPED CYCLING WHILE IN USE ON A PATIENT. ALTHOUGH REQUESTED, INFORMATION REGARDING THE INTERVENTION TAKEN ON THE BEHALF OF THE PATIENT HAS NOT BEEN PROVIDED. THE CUSTOMER REPORTED THERE WAS NO INJURY TO THE PATIENT.
Patient problems: No Code Available; No Information
MAUDE report 6620449 ↗
Other VENTILATOR, CONTINUOUS, FACILITY USE devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.