Puritan Bennett 980A1ENNISBRecalls & reported problems
Covidien LP · 980 Universal Ventilator · Class 2
All models of Puritan Bennett (119)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Covidien Puritan Bennett 980 ventilator has 86 malfunction reports, with no deaths or injuries. The reports mostly describe loss of ventilation during use and failed self-tests, and none were reported in the last 12 months, down from 6 in the 12 months before. Of its 2 recalls, one is open over alarms that may not sound or display, and one is closed over an incorrectly assembled capacitor.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 86
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 86 adverse event reports that list this exact device (DI 10884521201897), received between May 1, 2017 and Aug 20, 2025 — classified as 86 malfunctions.
The most frequently reported specific device problem is “Output Problem” (29 reports).
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Mar 25, 2022Class IOpen, Classified
The audible alarm may not sound and/or the omni-directional LED visual alarm may not display during alarm states. Delayed awareness of alarm states may lead to delayed response or a delay of treatment, potentially resulting in hypoxia, dyspnea, cardiac arrest, or death.
Z-0966-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Nov 4, 2021Class ITerminated
Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.
Z-0338-2022 · Root cause: Reprocessing Controls · Recall details · FDA record ↗
Most reported problems
Device problems
- Output Problem29
- Improper Flow or Infusion28
- Battery Problem12
- No Flow10
- Smoking9
- Device Displays Incorrect Message6
- Device Inoperable6
- No Audible Alarm4
Patient problems
- No Patient Involvement42
- No Clinical Signs, Symptoms or Conditions35
- No Information8
- No Consequences Or Impact To Patient7
- No Code Available2
Recent adverse event reports
The 25 most recent of 86 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 20, 2025No Flow; Output Problem; No Flow; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR HAD A SYSTEM ERROR. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22854110 ↗MalfunctionJun 12, 2025No Flow; Output Problem; No Flow; Output Problem
IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, THIS 980 VENTILATOR HAD A LOSS OF VENTILATION. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR. THERE WAS NO REPORTED INJURY TO THE PATIENT. POST VENTILATOR TRANSFER, THE DEVICE DID NOT PASS THE POWER ON SELF TEST (POST).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22199396 ↗MalfunctionJun 12, 2025No Flow; No Flow
IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, THIS 980 VENTILATOR EXPERIENCED A LOSS OF VENTILATION (LOV). THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR. THERE WAS NO REPORTED INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22199375 ↗MalfunctionApr 1, 2025Smoking; Smoking
IT WAS REPORTED THAT BEFORE THE PREVENTIVE MAINTENANCE WAS PERFORMED ON THESE 980 VENTILATORS, THE VENTILATOR "HAD TICKING SOUNDS WHEN IT WAS TURNED ON AND WITHIN 10 SECONDS SMOKE WAS OBSERVED COMING FROM THE EXPIRATORY SIDE AND INHALATION FAN". THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21742443 ↗MalfunctionDec 12, 2024Improper Flow or Infusion; Improper Flow or Infusion
IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, THIS 980 VENTILATOR HAD A ¿MALFUNCTION¿ AND DISPLAYED MESSAGES "VENTILATOR - INOPERATIVE, EST REQUIRED, AND DEVICE ALERT." THE PATIENT WAS REMOVED FROM THE VENTILATOR, MANUALLY VENTILATED, AND PLACED ON AN ALTERNATE VENTILATOR WITH NO INJURY. AFTER PATIENT REMOVAL, IT WAS REPORTED THAT THE VENTILATOR FAILED THE POWER ON SELF TEST (POST) WITH THE MESSAGE SHORT SHELF TEST NOT ALLOWED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20920873 ↗MalfunctionNov 13, 2024Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR FAILED THE EXTENDED SELF TEST (EST) CIRCUIT PRESSURE TEST. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20679240 ↗MalfunctionSep 27, 2024Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR GENERATED AN EXTENDED SELF TEST (EST) FAILURE DIAGNOSTIC CODE WITH AN "INSPIRATORY AUTO ZERO SOLENOID NOT OPERATIONAL" MESSAGE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20322981 ↗MalfunctionSep 13, 2024Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR FAILED THE EXTENDED SELF TEST (EST) WITH AN "INSPIRATORY AUTO ZERO SOLENOID NOT OPERATIONAL" MESSAGE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20214480 ↗MalfunctionSep 11, 2024Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THE 980 VENTILATOR FAILED THE CIRCUIT PRESSURE TEST WITH "INSPIRATORY AUTO ZERO SOLENOID NOT OPERATIONAL MESSAGE" DURING EXTENDED SELF TEST (EST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20197281 ↗MalfunctionJan 2, 2024Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THE 980 VENTILATOR FAILED THE EXTENDED SELF TEST WITH AN 'INSPIRATORY SOLENOID NOT OPERATIONAL' MESSAGE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18427108 ↗MalfunctionJan 2, 2024Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR FAILED THE EXTENDED SELF TEST (EST) WITH AN "INSPIRATORY AUTOZERO SOLENOID NOT OPERATIONAL" MESSAGE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18427090 ↗MalfunctionJan 2, 2024Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR FAILED THE EXTENDED SELF TEST (EST) WITH AN "INSPIRATORY AUTOZERO SOLENOID NOT OPERATIONAL" MESSAGE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18427075 ↗MalfunctionJan 2, 2024Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR FAILED THE EXTENDED SELF TEST (EST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18427047 ↗MalfunctionSep 12, 2023Smoking; Smoking
IT WAS REPORTED THAT THIS 980 VENTILATOR WAS EMITTING SMOKE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17733141 ↗MalfunctionAug 15, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR FAILED THE CIRCUIT PRESSURE TEST WITH AN "INSPIRATORY AUTO ZERO SOLENOID NOT OPERATIONAL" DURING THE EXTENDED SELF-TEST (EST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17549696 ↗MalfunctionAug 2, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR EXTENDED SELF TEST (EST) FAILED AT THE "CIRCUIT PRESSURE STEP "INSPIRATORY AUTO ZERO SOLENOID NOT OPERATIONAL". THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17445530 ↗MalfunctionJul 10, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR FAILED THE CIRCUIT PRESSURE TEST WITH AN "INSPIRATORY AUTO ZERO SOLENOID NOT OPERATIONAL" DURING THE EXTENDED SELF-TEST (EST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17293384 ↗MalfunctionJul 10, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR FAILED THE CIRCUIT PRESSURE TEST DURING THE EXTENDED SELF-TEST (EST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17292745 ↗MalfunctionJun 19, 2023Product Quality Problem; Improper Flow or Infusion; Product Quality Problem; Improper Flow or Infusion
IT WAS REPORTED DURING INSTALLATION THAT THIS 980 VENTILATOR HAD A FAILURE OUT OF BOX (FOOB) DUE TO FAILING THE CIRCUIT PRESSURE TEST DURING THE EXTENDED SELF-TEST (EST) AND DISPLAYED THE ERROR MESSAGE "INSPIRATORY AUTOZERO SOLENOID NOT OPERATIONAL" AND CORRESPONDING DIAGNOSTIC CODE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17155637 ↗MalfunctionFeb 2, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THE 980 VENTILATOR FAILED THE CIRCUIT PRESSURE PORTION OF THE SHORT SELF TESTING (SST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16287881 ↗MalfunctionJan 30, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR DID NOT PASS THE INSPIRATORY AUTO ZERO SOLENOID PORTION OF THE SHORT SELF TEST (SST) . THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16264974 ↗MalfunctionJan 25, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR DID NOT PASS THE CIRCUIT PRESSURE TEST PORTION OF THE EXTENDED SELF TEST (EST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16240849 ↗MalfunctionJan 25, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR DID NOT PASS THE CIRCUIT PRESSURE TEST PORTION OF THE EXTENDED SELF TEST (EST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16240843 ↗MalfunctionJan 17, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT A 980 VENTILATOR FAILED THE EXTENDED SELF TEST (EST) AND SHORT SELF TEST (SST) WITH DIAGNOSTIC CODES. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16186709 ↗MalfunctionJan 4, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem
IT WAS REPORTED THAT THIS 980 VENTILATOR DISPLAYED A "INSPIRATORY AUTOZERO SOLENOID NOT OPERATIONAL" MESSAGE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16096230 ↗
Other VENTILATOR, CONTINUOUS, FACILITY USE devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page