Puritan Bennett 980A1ENNISBRecalls & reported problems

Covidien LP · 980 Universal Ventilator · Class 2

All models of Puritan Bennett (119)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Covidien Puritan Bennett 980 ventilator has 86 malfunction reports, with no deaths or injuries. The reports mostly describe loss of ventilation during use and failed self-tests, and none were reported in the last 12 months, down from 6 in the 12 months before. Of its 2 recalls, one is open over alarms that may not sound or display, and one is closed over an incorrectly assembled capacitor.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
86
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 86 adverse event reports that list this exact device (DI 10884521201897), received between May 1, 2017 and Aug 20, 2025 — classified as 86 malfunctions.

The most frequently reported specific device problem is “Output Problem” (29 reports).

2 FDA recall records name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 4 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 25, 2022Class IOpen, Classified

    The audible alarm may not sound and/or the omni-directional LED visual alarm may not display during alarm states. Delayed awareness of alarm states may lead to delayed response or a delay of treatment, potentially resulting in hypoxia, dyspnea, cardiac arrest, or death.

    Z-0966-2022 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Nov 4, 2021Class ITerminated

    Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.

    Z-0338-2022 · Root cause: Reprocessing Controls · Recall details · FDA record ↗

Most reported problems

Device problems

  • Output Problem29
  • Improper Flow or Infusion28
  • Battery Problem12
  • No Flow10
  • Smoking9
  • Device Displays Incorrect Message6
  • Device Inoperable6
  • No Audible Alarm4

Patient problems

  • No Patient Involvement42
  • No Clinical Signs, Symptoms or Conditions35
  • No Information8
  • No Consequences Or Impact To Patient7
  • No Code Available2

Recent adverse event reports

The 25 most recent of 86 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 20, 2025No Flow; Output Problem; No Flow; Output Problem

    IT WAS REPORTED THAT THIS 980 VENTILATOR HAD A SYSTEM ERROR. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22854110 ↗
  • MalfunctionJun 12, 2025No Flow; Output Problem; No Flow; Output Problem

    IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, THIS 980 VENTILATOR HAD A LOSS OF VENTILATION. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR. THERE WAS NO REPORTED INJURY TO THE PATIENT. POST VENTILATOR TRANSFER, THE DEVICE DID NOT PASS THE POWER ON SELF TEST (POST).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22199396 ↗
  • MalfunctionJun 12, 2025No Flow; No Flow

    IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, THIS 980 VENTILATOR EXPERIENCED A LOSS OF VENTILATION (LOV). THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR. THERE WAS NO REPORTED INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22199375 ↗
  • MalfunctionApr 1, 2025Smoking; Smoking

    IT WAS REPORTED THAT BEFORE THE PREVENTIVE MAINTENANCE WAS PERFORMED ON THESE 980 VENTILATORS, THE VENTILATOR "HAD TICKING SOUNDS WHEN IT WAS TURNED ON AND WITHIN 10 SECONDS SMOKE WAS OBSERVED COMING FROM THE EXPIRATORY SIDE AND INHALATION FAN". THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21742443 ↗
  • MalfunctionDec 12, 2024Improper Flow or Infusion; Improper Flow or Infusion

    IT WAS REPORTED THAT, WHILE IN USE ON A PATIENT, THIS 980 VENTILATOR HAD A ¿MALFUNCTION¿ AND DISPLAYED MESSAGES "VENTILATOR - INOPERATIVE, EST REQUIRED, AND DEVICE ALERT." THE PATIENT WAS REMOVED FROM THE VENTILATOR, MANUALLY VENTILATED, AND PLACED ON AN ALTERNATE VENTILATOR WITH NO INJURY. AFTER PATIENT REMOVAL, IT WAS REPORTED THAT THE VENTILATOR FAILED THE POWER ON SELF TEST (POST) WITH THE MESSAGE SHORT SHELF TEST NOT ALLOWED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20920873 ↗
  • MalfunctionNov 13, 2024Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem

    IT WAS REPORTED THAT THIS 980 VENTILATOR FAILED THE EXTENDED SELF TEST (EST) CIRCUIT PRESSURE TEST. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20679240 ↗
  • MalfunctionSep 27, 2024Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem

    IT WAS REPORTED THAT THIS 980 VENTILATOR GENERATED AN EXTENDED SELF TEST (EST) FAILURE DIAGNOSTIC CODE WITH AN "INSPIRATORY AUTO ZERO SOLENOID NOT OPERATIONAL" MESSAGE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20322981 ↗
  • MalfunctionSep 13, 2024Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem

    IT WAS REPORTED THAT THIS 980 VENTILATOR FAILED THE EXTENDED SELF TEST (EST) WITH AN "INSPIRATORY AUTO ZERO SOLENOID NOT OPERATIONAL" MESSAGE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20214480 ↗
  • MalfunctionSep 11, 2024Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem

    IT WAS REPORTED THAT THE 980 VENTILATOR FAILED THE CIRCUIT PRESSURE TEST WITH "INSPIRATORY AUTO ZERO SOLENOID NOT OPERATIONAL MESSAGE" DURING EXTENDED SELF TEST (EST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20197281 ↗
  • MalfunctionJan 2, 2024Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem

    IT WAS REPORTED THAT THE 980 VENTILATOR FAILED THE EXTENDED SELF TEST WITH AN 'INSPIRATORY SOLENOID NOT OPERATIONAL' MESSAGE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18427108 ↗
  • MalfunctionJan 2, 2024Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem

    IT WAS REPORTED THAT THIS 980 VENTILATOR FAILED THE EXTENDED SELF TEST (EST) WITH AN "INSPIRATORY AUTOZERO SOLENOID NOT OPERATIONAL" MESSAGE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18427090 ↗
  • MalfunctionJan 2, 2024Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem

    IT WAS REPORTED THAT THIS 980 VENTILATOR FAILED THE EXTENDED SELF TEST (EST) WITH AN "INSPIRATORY AUTOZERO SOLENOID NOT OPERATIONAL" MESSAGE.  THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18427075 ↗
  • MalfunctionJan 2, 2024Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem

    IT WAS REPORTED THAT THIS 980 VENTILATOR FAILED THE EXTENDED SELF TEST (EST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18427047 ↗
  • MalfunctionSep 12, 2023Smoking; Smoking

    IT WAS REPORTED THAT THIS 980 VENTILATOR WAS EMITTING SMOKE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17733141 ↗
  • MalfunctionAug 15, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem

    IT WAS REPORTED THAT THIS 980 VENTILATOR FAILED THE CIRCUIT PRESSURE TEST WITH AN "INSPIRATORY AUTO ZERO SOLENOID NOT OPERATIONAL" DURING THE EXTENDED SELF-TEST (EST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17549696 ↗
  • MalfunctionAug 2, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem

    IT WAS REPORTED THAT THIS 980 VENTILATOR EXTENDED SELF TEST (EST) FAILED AT THE "CIRCUIT PRESSURE STEP "INSPIRATORY AUTO ZERO SOLENOID NOT OPERATIONAL".  THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17445530 ↗
  • MalfunctionJul 10, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem

    IT WAS REPORTED THAT THIS  980 VENTILATOR FAILED THE CIRCUIT PRESSURE TEST WITH AN "INSPIRATORY AUTO ZERO SOLENOID NOT OPERATIONAL" DURING THE EXTENDED SELF-TEST (EST).  THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17293384 ↗
  • MalfunctionJul 10, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem

    IT WAS REPORTED THAT THIS 980 VENTILATOR FAILED THE CIRCUIT PRESSURE TEST DURING THE EXTENDED SELF-TEST (EST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17292745 ↗
  • MalfunctionJun 19, 2023Product Quality Problem; Improper Flow or Infusion; Product Quality Problem; Improper Flow or Infusion

    IT WAS REPORTED DURING INSTALLATION THAT THIS 980 VENTILATOR HAD A FAILURE OUT OF BOX (FOOB) DUE TO FAILING THE CIRCUIT PRESSURE TEST DURING THE EXTENDED SELF-TEST (EST) AND DISPLAYED THE ERROR MESSAGE "INSPIRATORY AUTOZERO SOLENOID NOT OPERATIONAL" AND CORRESPONDING DIAGNOSTIC CODE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17155637 ↗
  • MalfunctionFeb 2, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem

    IT WAS REPORTED THAT THE 980 VENTILATOR FAILED THE CIRCUIT PRESSURE PORTION OF THE SHORT SELF TESTING (SST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16287881 ↗
  • MalfunctionJan 30, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem

    IT WAS REPORTED THAT THIS 980 VENTILATOR DID NOT PASS THE INSPIRATORY AUTO ZERO SOLENOID PORTION OF THE SHORT SELF TEST (SST) . THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16264974 ↗
  • MalfunctionJan 25, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem

    IT WAS REPORTED THAT THIS 980 VENTILATOR DID NOT PASS THE CIRCUIT PRESSURE TEST PORTION OF THE EXTENDED SELF TEST (EST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16240849 ↗
  • MalfunctionJan 25, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem

    IT WAS REPORTED THAT THIS 980 VENTILATOR DID NOT PASS THE CIRCUIT PRESSURE TEST PORTION OF THE EXTENDED SELF TEST (EST). THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16240843 ↗
  • MalfunctionJan 17, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem

    IT WAS REPORTED THAT A 980 VENTILATOR FAILED THE EXTENDED SELF TEST (EST) AND SHORT SELF TEST (SST) WITH DIAGNOSTIC CODES. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16186709 ↗
  • MalfunctionJan 4, 2023Improper Flow or Infusion; Output Problem; Improper Flow or Infusion; Output Problem

    IT WAS REPORTED THAT THIS 980 VENTILATOR DISPLAYED A "INSPIRATORY AUTOZERO SOLENOID NOT OPERATIONAL" MESSAGE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16096230 ↗

Other VENTILATOR, CONTINUOUS, FACILITY USE devices

All VENTILATOR, CONTINUOUS, FACILITY USE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page