1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
16
Deaths reported
0
Injuries reported
6
Recalls naming it
1

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 10884521205475), received between Dec 27, 2021 and Dec 3, 2024 — classified as 6 injuries, 10 malfunctions.

The most frequently reported specific device problem is “Gas/Air Leak” (9 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K150844); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 24, 2023Open, Classified

    A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to respiratory failure, dyspnea, treatment delay, tissue injury, or bleeding.

    Z-1357-2023 · Root cause: Component change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Gas/Air Leak9
  • Use of Device Problem7
  • Defective Component2
  • Material Split, Cut or Torn1

Patient problems

  • Unintended Extubation14
  • Discomfort2
  • Hemorrhage/Blood Loss/Bleeding2
  • Insufficient Information2
  • Low Oxygen Saturation2
  • Unspecified Tissue Injury1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 3, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    ACCORDING TO THE REPORTER, DURING USE, THE CANNULA INSERTED, BROKE THE CUFF ON THE MORNING WHERE THE TEAM OPTED TO INSERT A COMMON CANNULA DUE TO THE PATIENT'S SEVERITY AND RISKS.

    Patient problems: Unintended Extubation

    MAUDE report 20832904 ↗
  • MalfunctionNov 27, 2024Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, DURING USE, THE CANNULA INSERTED SHOWED AIR LEAKAGE THROUGH THE CUFF, REQUIRING A NEW TRACHEAL MOLD UNTIL NEW INSERTION.

    Patient problems: Unintended Extubation

    MAUDE report 20801309 ↗
  • MalfunctionNov 4, 2024Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, UPON INSERTION OF THE TRACHEOSTOMY CANNULA, IT WAS NOTED THAT THERE WAS AN AIR LEAKAGE SEEN IN THE CUFF. THE TRACH WAS REPLACED.

    Patient problems: Unintended Extubation

    MAUDE report 20597502 ↗
  • InjurySep 15, 2024Use of Device Problem; Gas/Air Leak; Use of Device Problem; Gas/Air Leak

    ACCORDING TO THE REPORTER, THE CUFF OF THE TRACHEOSTOMY TUBE WAS DEFLATED PREMATURELY. THE PATIENT HAD TO CHANGE THE TRACH TUBE EVERY WEEK OR EVERY 15 DAYS INSTEAD OF ONCE A MONTH RESULTED TO DISCOMFORT. UPON VISUAL EXAMINATION, THE NURSE REPORTED THAT THE CUFF WAS 'WRINKLED' AND WHEN IT WAS INFLATED WITH SYRINGE, A PUNCTURED CUFF WAS OBSERVED.

    Patient problems: Discomfort; Unintended Extubation

    MAUDE report 20220984 ↗
  • InjurySep 15, 2024Use of Device Problem; Gas/Air Leak; Use of Device Problem; Gas/Air Leak

    ACCORDING TO THE REPORTER, THE BALLOON OF THE TRACHEOSTOMY TUBE WAS DEFLATED PREMATURELY. THE PATIENT HAD TO CHANGE THE TRACH TUBE EVERY WEEK OR EVERY 15 DAYS INSTEAD OF ONCE A MONTH RESULTED TO DISCOMFORT. UPON VISUAL EXAMINATION, THE NURSE REPORTED THAT THE BALLOON WAS 'WRINKLED' AND WHEN IT WAS INFLATED WITH SYRINGE, A LEAKED BALLOON WAS OBSERVED.

    Patient problems: Discomfort; Unintended Extubation

    MAUDE report 20220973 ↗
  • InjuryJan 10, 2024Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, BOTH TRACHEOSTOMY TUBE'S CUFF HAD A LEAK DURING PATIENT USE. THE PATIENT DESATURATED DOWN TO 70% DURING THE 3RD INSERTING TENTATIVE. TUBES WERE REPLACED TWICE FROM THE PATIENT.

    Patient problems: Low Oxygen Saturation; Unintended Extubation

    MAUDE report 18486593 ↗
  • MalfunctionJan 8, 2024Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, BOTH TRACHEOSTOMY TUBE'S CUFF HAD A LEAK DURING PATIENT USE. THE PATIENT DESATURATED DOWN TO 70% DURING THE 3RD INSERTING TENTATIVE. TUBES WERE REPLACED TWICE FROM THE PATIENT.

    Patient problems: Low Oxygen Saturation; Unintended Extubation

    MAUDE report 18472372 ↗
  • MalfunctionMay 9, 2023Defective Component; Defective Component

    ACCORDING TO THE REPORTER, PERFORMED SEVERAL EXCHANGES OF CANNULA BECAUSE THE CUFF WAS NOT KEPT INFLATED. AFTER TESTING IN A SINK WITH WATER, IT WAS NOTICED THAT THE CUFF HAD HOLES IN IT. THERE WAS NO REPORTED PATIENT OUTCOME.

    Patient problems: Insufficient Information

    MAUDE report 16897993 ↗
  • MalfunctionMay 9, 2023Defective Component; Defective Component

    ACCORDING TO THE REPORTER, PERFORMED SEVERAL EXCHANGES OF CANNULA BECAUSE THE CUFF WAS NOT KEPT INFLATED. AFTER TESTING IN A SINK WITH WATER, IT WAS NOTICED THAT THE CUFF HAD HOLES IN IT. THERE WAS NO REPORTED PATIENT OUTCOME.

    Patient problems: Insufficient Information

    MAUDE report 16897650 ↗
  • MalfunctionOct 20, 2022Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, INTRA-OPERATIVELY, THE DEVICE DEVELOPED CUFF LEAK. TRACHEOTOMY REQUIRED UNPLANNED REPLACEMENT WITH RESPIRATORY BACKUP AND ANESTHESIA. DEVICE CUFF WOULD NOT STAY INFLATED LEADING TO DEFLATION WHEN IT WAS TESTED AFTER IT WAS REMOVED.

    Patient problems: Unintended Extubation

    MAUDE report 15637364 ↗
  • InjuryAug 10, 2022Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE CUFFS WERE CHECKED PRIOR TO INSERTION BUT UNFORTUNATELY THEY BOTH POPPED ONCE INSERTED. PATIENT THEN NEEDED FURTHER SURGERY TO CONTROL BLEEDING.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Unintended Extubation

    MAUDE report 15202280 ↗
  • InjuryJun 23, 2022Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE CUFFS WERE CHECKED PRIOR TO INSERTION BUT UNFORTUNATELY THEY BOTH POPPED ONCE INSERTED. PATIENT THEN NEEDED FURTHER SURGERY TO CONTROL BLEEDING.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Unintended Extubation

    MAUDE report 14789674 ↗
  • InjuryFeb 14, 2022Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, THE DEVICE WAS INSERTED A DAY LATER OR SO DEVICE HAD A SLOW LEAK. AFTER INSPECTION, THE TRACH DID NOT SHOW ANY VISIBLE DAMAGE AND WILL HOLD AIR, HOWEVER, OVERTIME IT SLOWLY DEFLATED. IT WAS REPORTED THAT THERE WAS PATIENT HARM AS RESULT OF THE ISSUE.

    Patient problems: Unintended Extubation; Unspecified Tissue Injury

    MAUDE report 13522742 ↗
  • MalfunctionDec 27, 2021Use of Device Problem; Gas/Air Leak; Use of Device Problem; Gas/Air Leak

    ACCORDING TO THE REPORTER, THE PHYSICIAN PLACED THE FIRST CANNULA BUT FIGURED OUT IMMEDIATELY AFTER PLACEMENT THAT THE CUFF WAS BROKEN. THE SAME WITH THE SECOND CUFF. BEFORE PLACING THE THIRD, THE PHYSICIAN INFLATED THE CUFF, AND AGAIN THE CUFF BROKE. A LEAK WAS FOUND. AFTER THIS, THE PHYSICIAN PLACED A SMALLER ONE, SIZE 8. UNTIL TODAY THERE ARE NO ISSUES.

    Patient problems: Unintended Extubation

    MAUDE report 13090688 ↗
  • MalfunctionDec 27, 2021Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE PHYSICIAN PLACED THE FIRST CANNULA BUT FIGURED OUT IMMEDIATELY AFTER PLACEMENT THAT THE CUFF WAS BROKEN AND THE SAME WITH THE SECOND CUFF. BEFORE PLACING THE THIRD, THE PHYSICIAN INFLATED THE CUFF, AND AGAIN THE CUFF BROKE. AFTER THIS, THE PHYSICIAN PLACED A SMALLER ONE, SIZE 8. UNTIL TODAY THERE ARE NO ISSUES.

    Patient problems: Unintended Extubation

    MAUDE report 13090679 ↗
  • MalfunctionDec 27, 2021Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE PHYSICIAN PLACED THE FIRST CANNULA BUT FIGURED OUT IMMEDIATELY AFTER PLACEMENT THAT THE CUFF WAS BROKEN AND THE SAME WITH THE SECOND CUFF. BEFORE PLACING THE THIRD, THE PHYSICIAN INFLATED THE CUFF, AND AGAIN THE CUFF BROKE. AFTER THIS, THE PHYSICIAN PLACED A SMALLER ONE, SIZE 8. UNTIL TODAY THERE ARE NO ISSUES.

    Patient problems: Unintended Extubation

    MAUDE report 13090594 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K150844. One submission often covers many models, so check the product and lot details in each record.

  • Z-1390-2025 · Feb 26, 2025 · Open, Classified
    Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula, REF: 7CN80R

Other Tube tracheostomy and tube cuff devices