Shiley recallZ-1357-2023 · Covidien · initiated Mar 24, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Covidien or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to respiratory failure, dyspnea, treatment delay, tissue injury, or bleeding.

FDA root cause category: Component change control

What the company is doing

On 3/24/23, recall notices were mailed to risk managers and directors of respiratory care. Tracheostomy Tubes that do not securely attach or stay attached to caps or accessories should not be used. 1) All affected unused product should be quarantined and returned to the recalling firm. 2) Complete and then return the customer confirmation form to [email protected] 3) Pass on this notice to all those who need to be aware within your organization or to any organization where the potentially affected product from the specified lots has been transferred or distributed. 4) If purchased from a distributor, contact your distributor directly to arrange for the return of the product back to your distributor. Customers with questions can contact Quality Assurance at 800-255-6774, option 4.

Recall details

Recall number
Z-1357-2023
Status
Open, Classified
Initiated by the company
Mar 24, 2023
Posted by the FDA
Apr 19, 2023
Recalling firm
Covidien
Quantity
599,187
Product code
JOH
FDA submissions
K142296, K150844
Distribution
Worldwide - US Nationwide distribution including in the states of PR, CA, FL, MN, PA, AZ, IL, TN, OH, MI, GA, KS, HI, MD, AK, NY, NE, IA, OK, VA, ND, MO, NC, SC, KY, OR, IN, WA, TX, WI, CT, MA, WV, CO, VT, NJ, SD, AR, AL, MS, LA, DE, MT, WY, DC, UT, NV, NH, RI, ID, NM, ME and the countries of Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Belgium, Bolivia, Plurinational State Of Bosnia And Herzegovina, Brazil, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Cyprus, Czech Republic, Denmark, El Salvador, Estonia, Finland, France, French Guiana, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, Indonesia, Ireland, Israel, Italy, Jordan, Kosovo, Kuwait, Lebanon, Lithuania, Luxembourg, Macao, Malaysia, Malta, Martinique, Mexico, Namibia, Netherlands, New Caledonia, New Zealand, North Macedonia, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Republic Of Tanzania, Thailand, Trinidad And Tobago, Turkey, United Arab Emirates, Viet Nam.

Product description

Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff, and Cuffless: Disposable Inner Cannula or Reusable Inner Cannula (REF, Product Description): 10CN10H, 10.0MM SHILEY CUFFED TRACH CAN; 10CN10R, 10.0MM ADT FLEX TRACH W TG CUFF; 10UN10A, 10.0MM UNCUFF TRACH TUBE X1; 10UN10H, 10.0MM SHILEY CUFFLESS TRACH C; 10UN10R, 10.0MM ADT FLEX TRACH CUFFLESS; 4CN65A, 6.5MM TRACH TUBE W TG CUFF X1; 4CN65H, 6.5MM SHILEY CUFFED TRACH CANN; 4CN65R, 6.5MM ADT FLEX TRACH W TG CUFF X1; 4UN65A, 6.5MM UNCUFF TRACH TUBE X1; 4UN65H, 6.5MM SHILEY CUFFLESS TRACH CA; 4UN65R, 6.5MM ADT FLEX TRACH CUFFLESSX1; 5CN70A, 7.0MM TRACH TUBE W TG CUFF X1; 5CN70H, 7.0MM SHILEY CUFFED TRACH CANN; 5CN70R, 7.0MM ADT FLEX TRACH W TG CUFF X1; 5UN70A, 7.0MM UNCUFF TRACH TUBE X1; 5UN70H, 7.0MM SHILEY CUFFLESS TRACH CA; 5UN70R, 7.0MM ADT FLEX TRACH CUFFLESSX1; 6CN75A, 7.5MM TRACH TUBE W TG CUFF X1; 6CN75H, 7.5MM SHILEY CUFFED TRACH CANN; 6CN75R, 7.5MM ADT FLEX TRACH W TG CUFF X1; 6UN75H, 7.5MM SHILEY CUFFLESS TRACH CA; 6UN75R, 7.5MM ADT FLEX TRACH CUFFLESSX1; 7CN80A, 8.0MM TRACH TUBE W TG CUFF X1; 7CN80H, 8.0MM SHILEY CUFFED TRACH CANN; 7CN80R, 8.0MM ADT FLEX TRACH W TG CUFF X1; 7UN80A, 8.0MM UNCUFF TRACH TUBE X1; 7UN80H, 8.0MM SHILEY CUFFLESS TRACH CA; 7UN80R, 8.0MM ADT FLEX TRACH CUFFLESSX1; 8CN85A, 8.5MM TRACH TUBE W TG CUFF X1; 8CN85H, 8.5MM SHILEY CUFFED TRACH CANN; 8CN85R, 8.5MM ADT FLEX TRACH W TG CUFF X1; 8UN85A, 8.5MM UNCUFF TRACH TUBE X1; 8UN85H, 8.5MM SHILEY CUFFLESS TRACH CA; 8UN85R, 8.5MM ADT FLEX TRACH CUFFLESSX1; 9CN90H, 9.0MM SHILEY CUFFED TRACH CANN; 9CN90R, 9.0MM ADT FLEX TRACH W TG CUFF X1; 9UN90H, 9.0MM SHILEY CUFFLESS TRACH CA; 9UN90R, 9.0MM ADT FLEX TRACH CUFFLESSX1

Code information (models, lots, serial numbers)

Show the full code information (13k characters)

REF/UDI-DI/Lot: 10CN10H/10884521172371/21C0402JZX, 21D0461JZX, 21G0457JZX, 21I0293JZX, 21I0294JZX, 21I0552JZX, 21J1435JZX, 21J1436JZX, 21K0275JZX, 21K0276JZX, 21K0277JZX, 21K0278JZX; 10CN10R/10884521205475/202107218X, 202107296X, 202108128X, 202108285X, 21H0792JZX, 21H0793JZX, 21H0794JZX, 22I0596JZX; 10UN10A/20884521175690/21D0463JZX; 10UN10H/20884521176291/21E0661JZX, 21G0473JZX; 10UN10R/10884521205543/202105307X, 202105308X, 202106148X, 202209335X; 4CN65A/10884521172456/21C0432JZX, 21I0369JZX; 4CN65H/20884521176307/21C0358JZX, 21C0359JZX, 21C0360JZX, 21C0361JZX, 21D0440JZX, 21D0441JZX, 21E0512JZX, 21E0513JZX, 21E0514JZX, 21E0515JZX, 21E0516JZX, 21E0517JZX, 21E0518JZX, 21E0519JZX, 21E0520JZX, 21E0521JZX, 21E0522JZX, 21E0523JZX, 21E0524JZX, 21E0525JZX, 21E0526JZX, 21E0527JZX, 21E1773JZX, 21E1774JZX, 21E1775JZX, 21E1776JZX, 21E1777JZX, 21E1778JZX, 21E1779JZX, 21E1780JZX, 21E1781JZX, 21E1782JZX, 21E1783JZX, 21H1324JZX, 21J0181JZX, 21J1094JZX, 21K0237JZX, 21K0239JZX, 21K0639JZX, 21K0643JZX, 21K0645JZX, 22C1465JZX, 22C1466JZX, 22C1467JZX, 22D0798JZX, 22D0799JZX, 22D0800JZX, 22D0801JZX, 22D0891JZX, 22D0892JZX, 22D0893JZX, 22D0897JZX, 22D0898JZX, 22D0899JZX, 22D0900JZX, 22D0901JZX, 22D0909JZX, 22F0918JZX, 22F0919JZX, 22F0920JZX, 22F0921JZX, 22F0922JZX, 22F0923JZX, 22F0924JZX, 22F0925JZX, 22F0926JZX, 22F0927JZX, 22F0928JZX, 22F0929JZX, 22J0073JZX, 22J0074JZX, 22J0075JZX, 22J0076JZX, 22J0077JZX, 22J0080JZX, 22J0081JZX, 22J0083JZX; 4CN65R/10884521205024/202105082X, 202105310X, 202106267X, 202106268X, 202107294X, 202107295X, 202111088X, 202111270X, 202111399X, 202112135X, 202201303X, 202202186X; 4UN65A/20884521175706/21D0466JZX, 21E0659JZX, 21E0660JZX; 4UN65H/20884521176239/21C0413JZX, 21C0414JZX, 21C0416JZX, 21C0417JZX, 21D0898JZX, 21D0899JZX, 21D0900JZX, 21D0901JZX, 21E1784JZX, 21E1785JZX, 22D0930JZX, 22D0931JZX, 22D0932JZX, 22D0933JZX, 22E0679JZX, 22E0680JZX; 4UN65R/10884521205482/202107297X, 202107298X, 202107299X, 202107300X, 202108122X, 202108123X, 202108124X, 202108125X, 202108139X, 202108140X, 202108228X, 202108229X, 21E1139JZX; 5CN70A/10884521172463/22D0927JZX, 22D0928JZX, 22D0929JZX, 22H0031JZX, 22H0032JZX, 22H0033JZX, 22H0860JZX, 22H0861JZX, 22H0863JZX, 22H0865JZX, 22J0071JZX, 22J0072JZX; 5CN70H/20884521176314/21C0404JZX, 21C0405JZX, 21C0406JZX, 21D0462JZX, 21G1130JZX, 21I0283JZX, 21I0284JZX, 21I0285JZX, 21I0286JZX, 21I0519JZX, 21I0520JZX, 21I0521JZX, 21I0522JZX, 21I0544JZX, 21I0545JZX, 21I0546JZX, 21I0630JZX, 21I0631JZX, 22D0924JZX, 22D0925JZX, 22D0926JZX, 22H0152JZX; 5CN70R/10884521205420/202111135X, 202208294X, 202208295X, 202210208X; 5UN70A/20884521175713/21B0062JZX, 21E0645JZX, 21E0646JZX, 21E0647JZX; 5UN70H/20884521176246/21E0655JZX, 21E0656JZX, 21G1008JZX, 21G1009JZX, 21H0735JZX; 5UN70R/10884521205499/202208296X; 6CN75A/10884521172470/22A0137JZX, 22H0868JZX; 6CN75H/20884521176321/21D0420JZX, 21D0421JZX, 21D0422JZX, 21D0423JZX, 21D0424JZX, 21D0425JZX, 21D0426JZX, 21D0427JZX, 21D0428JZX, 21D0429JZX, 21D0430JZX, 21D0431JZX, 21D0432JZX, 21D0433JZX, 21D0434JZX, 21D0435JZX, 21D0436JZX, 21D0437JZX, 21D0438JZX, 21D0439JZX, 21D0741JZX, 21D0742JZX, 21D0743JZX, 21D0744JZX, 21D0745JZX, 21D0746JZX, 21D0747JZX, 21D0748JZX, 21D0749JZX, 21E0632JZX, 21E0633JZX, 21E0634JZX, 21E0635JZX, 21E0636JZX, 21E0637JZX, 21E0638JZX, 21E0639JZX, 21E0640JZX, 21E0641JZX, 21E0642JZX, 21E0643JZX, 21E0644JZX, 21E1100JZX, 21E1101JZX, 21E1673JZX, 21E1674JZX, 21E1675JZX, 21E1676JZX, 21E1677JZX, 21E1678JZX, 21E1679JZX, 21E1680JZX, 21E1681JZX, 21E1682JZX, 21E1683JZX, 21E1684JZX, 21E1685JZX, 21E1686JZX, 21E1687JZX, 21E1688JZX, 21E1689JZX, 21E1690JZX, 21E1691JZX, 21E1694JZX, 21E1695JZX, 21E1696JZX, 21E1697JZX, 21E1769JZX, 21F0692JZX, 21F0693JZX, 21F0694JZX, 21G0296JZX, 21G0297JZX, 21G0298JZX, 21G0299JZX, 21G0426JZX, 21G0427JZX, 21G0428JZX, 21G0429JZX, 21G0430JZX, 21H0624JZX, 21I0260JZX, 21I0261JZX, 21I0262JZX, 21I0263JZX, 21I0264JZX, 21I0265JZX, 21I0266JZX, 21I0267JZX, 21I0268JZX, 21I0269JZX, 21I0270JZX, 21I0271JZX, 21I0491JZX, 21I0492JZX, 21I0680JZX, 21I0681JZX, 21I0682JZX, 21I0683JZX, 21J0090JZX, 21J0168JZX, 21J0169JZX, 21J0170JZX, 21J0171JZX, 21J0172JZX, 21J0176JZX, 21J0177JZX, 21J1088JZX, 21J1410JZX, 21J1411JZX, 21J1417JZX, 21J1418JZX, 21L0549JZX, 21L0550JZX, 21L0551JZX, 21L0552JZX, 21L0553JZX, 21L0756JZX, 21L0777JZX, 21L0778JZX, 21L0779JZX, 21L0796JZX, 21L0797JZX, 21L0801JZX, 21L0802JZX, 21L0803JZX, 21L0804JZX, 21L0805JZX, 21L0806JZX, 21L0807JZX, 21L0808JZX, 21L0809JZX, 21L0810JZX, 21L0811JZX, 21L0812JZX, 21L0813JZX, 21L0814JZX, 21L0815JZX, 21L0816JZX, 21L0817JZX, 21L0818JZX, 21L0819JZX 22A0665JZX, 22A0666JZX, 22A0670JZX, 22A0671JZX, 22A0672JZX, 22A0673JZX, 22A0674JZX, 22A0771JZX, 22A0772JZX, 22A0900JZX, 22A0901JZX, 22A0902JZX, 22A0903JZX, 22A0904JZX, 22A0905JZX, 22A0906JZX, 22A0907JZX, 22A0908JZX, 22A0909JZX, 22A0910JZX, 22A0911JZX, 22A0912JZX, 22A0913JZX, 22A0914JZX, 22A0915JZX, 22A0916JZX, 22B0277JZX, 22B0280JZX, 22B0288JZX, 22B0289JZX, 22B0290JZX, 22B0291JZX, 22B0334JZX, 22B0335JZX, 22B0336JZX, 22B0337JZX, 22C0758JZX, 22C0759JZX, 22D0936JZX, 22D0937JZX, 22D0938JZX, 22D0939JZX, 22D0940JZX, 22D0941JZX, 22D0942JZX, 22D0943JZX, 22D0944JZX, 22D0945JZX, 22D0946JZX, 22D0947JZX, 22D0948JZX, 22D0951JZX, 22D0952JZX, 22D0953JZX, 22D0954JZX, 22D0955JZX, 22D0956JZX, 22D0957JZX, 22D0958JZX, 22D0960JZX, 22D0961JZX, 22D0962JZX, 22D0963JZX, 22D0964JZX, 22D0965JZX, 22D0966JZX, 22D0967JZX, 22D0968JZX, 22D0969JZX, 22D0970JZX, 22D0971JZX, 22D0972JZX, 22G0403JZX, 22G0404JZX, 22G0405JZX, 22G0406JZX, 22G0407JZX, 22G0408JZX, 22G0414JZX, 22G0420JZX, 22H0872JZX, 22H0873JZX, 22H0874JZX, 22H0875JZX, 22H0877JZX, 22H0881JZX, 22H0886JZX, 22H0887JZX, 22J0084JZX, 22J0085JZX, 22J0086JZX, 22J0093JZX, 22J0094JZX, 22J0095JZX, 22J0096JZX, 22J0097JZX, 22J0098JZX, 22J0099JZX, 22J0100JZX, 22J0101JZX, 22J0102JZX, 22J0103JZX, 22J0104JZX, 22J0187JZX, 22J0189JZX, 22J0190JZX, 22J0191JZX, 22J0192JZX, 22J0193JZX, 22J0194JZX, 22J0196JZX, 22J0197JZX, 22J0198JZX, 22J0206JZX, 22K0244JZX, 22K1921JZX, 22K1922JZX, 22K1983JZX; 6CN75R/10884521205437/202110171X, 202110231X, 202110310X, 202110387X, 202111268X, 202203160X, 202203304X, 202203305X, 202203306X, 202203406X, 202204167X, 202206076X, 202206077X, 202206078X, 202207117X, 202207266X, 202207267X, 202207268X, 202208201X, 202208202X, 202209220X, 202210429X, 21G0479JZX, 21G0480JZX, 21G0481JZX, 21G0482JZX, 21L0522JZX, 21L0523JZX, 21L0524JZX, 21L0526JZX; 6UN75H/20884521176253/21C0420JZX, 21C0421JZX, 21C0422JZX, 21C0423JZX, 21C0424JZX, 21C0425JZX, 21C0426JZX, 21C0427JZX, 21C0428JZX, 21C0429JZX, 21C0430JZX, 21E1758JZX, 21E1759JZX, 21E1760JZX, 21E1761JZX, 21E1762JZX, 21E1763JZX, 21E1764JZX, 21E1765JZX, 21E1766JZX, 21E1767JZX, 21F0225JZX, 21G1158JZX, 21G1160JZX, 21G1161JZX, 21G1162JZX, 21G1163JZX, 21H0749JZX, 21H1009JZX, 21I0022JZX, 22B1035JZX, 22B1036JZX, 22B1039JZX, 22B1040JZX, 22B1041JZX, 22B1042JZX, 22B1043JZX, 22B1044JZX, 22B1045JZX, 22B1046JZX, 22B1047JZX, 22B1048JZX, 22B1049JZX, 22B1052JZX, 22B1053JZX, 22B1054JZX, 22C1511JZX, 22C1512JZX, 22C1513JZX, 22C1514JZX, 22C1515JZX, 22C1516JZX, 22C1518JZX, 22C1519JZX, 22D0792JZX, 22D0793JZX, 22D0794JZX, 22D0795JZX, 22D0796JZX, 22D0902JZX, 22D0903JZX, 22D0904JZX, 22D0905JZX, 22D0906JZX, 22D0907JZX, 22D0908JZX, 22D0913JZX, 22D0914JZX, 22D0915JZX, 22D0916JZX, 22D0919JZX, 22D0920JZX, 22D0921JZX, 22D0922JZX, 22E1002JZX, 22E1003JZX, 22E1004JZX, 22E1005JZX, 22E1008JZX, 22E1009JZX, 22E1010JZX, 22E1012JZX, 22E1013JZX, 22E1014JZX, 22E1015JZX, 22E1016JZX, 22E1017JZX, 22K0254JZX, 22K0255JZX, 22K0256JZX, 22K0257JZX, 22K0258JZX, 22K0259JZX, 22K0260JZX, 22K0261JZX, 22K0262JZX, 22K0263JZX, 22K0264JZX, 22K0265JZX, 22K0267JZX, 22K0268JZX, 22K0269JZX, 22K0270JZX, 22K0271JZX, 22K0272JZX, 22K0273JZX, 22K0274JZX, 22K0275JZX; 6UN75R/10884521205505/202107305X, 202107306X, 202108126X, 202108178X, 202108179X, 202109214X, 202109215X, 202109340X, 202111221X, 202111222X, 202111266X, 202201019X, 21G0487JZX, 21H0958JZX, 22B0728JZX, 22C1156JZX, 22C1157JZX, 22C1158JZX, 22C1159JZX, 22C1160JZX, 22E1001JZX, 22H0064JZX, 22H0065JZX, 22H0067JZX, 22H0068JZX, 22H0069JZX, 22H0072JZX, 22H0073JZX, 22H0075JZX, 22H0076JZX; 7CN80A/10884521172487/21C0355JZX, 22D0975JZX, 22D0976JZX, 22E0964JZX; 7CN80H/20884521176338/21C0381JZX, 21C0382JZX, 21C0383JZX, 21C0384JZX, 21D0442JZX, 21D0443JZX, 21E1698JZX, 21G0454JZX, 21I0402JZX, 21K0651JZX, 21K0652JZX, 21K0653JZX, 21K0654JZX, 21K0655JZX; 7CN80R/10884521205444/202104382X, 202104383X, 202105228X, 202105284X, 202106225X, 202203162X, 202203309X, 202204065X, 202206207X, 202206208X, 202206209X, 202206297X, 21G0791JZX; 7UN80A/20884521175737/20J0050JZX, 21C0356JZX, 21E1744JZX, 22C0313JZX, 22C0314JZX; 7UN80H/20884521176260/21C0431JZX, 21E1768JZX, 22C1527JZX, 22C1528JZX; 7UN80R/10884521205512/202203307X, 21G0774JZX; 8CN85A/10884521172494/21C0434JZX, 21C0435JZX, 21C0436JZX, 21I0287JZX, 21I0366JZX, 22A0136JZX, 22E1151JZX, 22E1152JZX, 22E1156JZX, 22E1157JZX; 8CN85H/20884521176345/21C0386JZX, 21C0387JZX, 21C0388JZX, 21C0389JZX, 21C0390JZX, 21C0391JZX, 21C0392JZX, 21C0393JZX, 21C0394JZX, 21C0395JZX, 21C0396JZX, 21C0397JZX, 21C0398JZX, 21C0399JZX, 21C0400JZX, 21C0401JZX, 21D0444JZX, 21D0445JZX, 21D0446JZX, 21D0447JZX, 21D0448JZX, 21D0449JZX, 21D0450JZX, 21D0451JZX, 21D0452JZX, 21D0453JZX, 21D0454JZX, 21D0455JZX, 21D0456JZX, 21D0457JZX, 21D0460JZX, 21E0623JZX, 21E0624JZX, 21E0625JZX, 21E0626JZX, 21E0627JZX, 21E0628JZX, 21E0629JZX, 21E0630JZX, 21E0631JZX, 21E1094JZX, 21E1095JZX, 21E1142JZX, 21E1143JZX, 21E1146JZX, 21E1147JZX, 21E1550JZX, 21E1551JZX, 21E1552JZX, 21E1711JZX, 21E1714JZX, 21E1715JZX, 21E1716JZX, 21E1717JZX, 21E1718JZX, 21E1719JZX, 21E1720JZX, 21E1721JZX, 21E1722JZX, 21E1723JZX, 21E1724JZX, 21E1725JZX, 21E1726JZX, 21E1727JZX, 21E1728JZX, 21E1729JZX, 21E1730JZX, 21E1731JZX, 21E1732JZX, 21E1733JZX, 21E1734JZX, 21E1735JZX, 21E1736JZX, 21E1737JZX, 21E1738JZX, 21F0524JZX, 21G0287JZX, 21G0288JZX, 21G0289JZX, 21G0290JZX, 21G0291JZX, 21G1149JZX, 21G1150JZX, 21H0796JZX, 21H0797JZX, 21I0330JZX, 21I0331JZX, 21I0332JZX, 21I0333JZX, 21I0334JZX, 21I0367JZX, 21I0368JZX, 21J0186JZX, 21J0187JZX, 21J0188JZX, 21J0189JZX, 21J0190JZX, 21J0191JZX, 21J0193JZX, 21L0333JZX, 22A0694JZX, 22A0695JZX, 22A0779JZX, 22A0780JZX, 22A0781JZX, 22A0782JZX, 22A0783JZX, 22A0784JZX, 22A0785JZX, 22A0786JZX, 22A0787JZX, 22A0788JZX, 22A0820JZX, 22A0821JZX, 22A0822JZX, 22A0823JZX, 22A0824JZX, 22A0825JZX, 22A0826JZX, 22A0827JZX, 22A0828JZX, 22A0829JZX, 22A0830JZX, 22A0831JZX, 22A0832JZX, 22A0833JZX, 22A0835JZX, 22A0836JZX, 22C0795JZX, 22C0797JZX, 22C0798JZX, 22C0799JZX ,22C0800JZX ,22C0802JZX 22C0813JZX, 22C0814JZX, 22C0815JZX, 22C0819JZX, 22C0820JZX, 22E1231JZX, 22E1232JZX, 22E1233JZX, 22E1234JZX, 22E1235JZX, 22E1236JZX, 22E1237JZX, 22E1238JZX, 22E1239JZX, 22E1240JZX, 22E1243JZX, 22E1244JZX, 22E1245JZX, 22E1246JZX, 22E1247JZX, 22E1248JZX, 22E1249JZX, 22E1260JZX, 22E1261JZX, 22E1262JZX, 22E1263JZX, 22E1264JZX, 22E1265JZX, 22E1266JZX, 22E1267JZX, 22F0476JZX, 22F0477JZX, 22F0478JZX, 22F0479JZX, 22F0480JZX, 22F0481JZX, 22F0482JZX, 22F0483JZX, 22F0484JZX, 22F0485JZX, 22F0486JZX, 22F0487JZX, 22F0488JZX, 22F0489JZX, 22F0490JZX, 22F0491JZX, 22F0492JZX, 22F0493JZX, 22F0494JZX, 22F0495JZX, 22F0496JZX, 22F0497JZX, 22F0498JZX, 22F0499JZX, 22F0500JZX, 22F0501JZX, 22F0502JZX, 22F0503JZX, 22F0504JZX, 22F0505JZX, 22F0506JZX, 22F0862JZX, 22F0863JZX, 22F0864JZX, 22F0897JZX, 22F0898JZX, 22F0899JZX, 22F0900JZX, 22I0530JZX, 22I0531JZX, 22I0532JZX, 22I0533JZX, 22I0534JZX, 22J0120JZX, 22J0121JZX, 22J0122JZX, 22J0123JZX, 22J0124JZX, 22J0125JZX, 22J0126JZX, 22J0127JZX, 22J0128JZX, 22J0129JZX, 22J0130JZX, 22J0131JZX, 22J0132JZX, 22J1084JZX, 22J1085JZX, 22J1096JZX, 22J1097JZX, 22J1098JZX, 22J1099JZX, 22J1100JZX, 22J1101JZX, 22J1102JZX, 22J1103JZX, 22J1104JZX, 22J1105JZX, 22J1107JZX, 22J1108JZX, 22J1109JZX, 22J1115JZX; 8CN85R/10884521205451/202106382X, 202107194X, 202107195X, 202107196X, 202109037X, 202109124X, 202109168X, 202110189X, 202111087X, 202111106X, 202111115X, 202201302X, 202202117X, 202202118X, 202202120X, 202202323X, 202202438X, 202204062X, 21H0762JZX, 21H0763JZX, 21H0764JZX, 21H0795JZX, 21H0951JZX, 21H0952JZX, 21J0382JZX, 21J0383JZX, 22A0732JZX, 22F0889JZX, 22H0040JZX; 8UN85A/20884521175744/22C1488JZX; 8UN85H/20884521176277/21C0408JZX, 21C0409JZX, 21C0410JZX, 21E0653JZX, 21E0654JZX, 21E1745JZX, 21E1746JZX, 21E1747JZX, 21E1748JZX, 21E1749JZX, 21E1750JZX, 21E1751JZX, 21E1752JZX, 21E1753JZX, 21F0523JZX, 21H0479JZX, 21L0296JZX, 21L0297JZX, 21L0298JZX, 21L0299JZX, 21L0300JZX, 21L0301JZX, 21L0302JZX, 21L0303JZX, 21L0304JZX, 21L0305JZX, 21L0380JZX, 22B1059JZX, 22B1060JZX, 22C1484JZX, 22C1485JZX; 8UN85R/10884521205529/202105199X, 202105200X, 202105201X, 202105383X, 202105384X, 202105385X, 202106264X, 202106265X, 202106313X, 202106314X, 202106315X, 202106316X, 202107063X, 202107064X, 202107065X, 202107113X, 202107114X, 202107115X, 202107116X, 202107202X, 202108121X, 202108187X, 202108188X, 202108230X, 202202326X, 202203066X, 202203409X, 202208021X, 202209336X, 21G0922JZX, 22C1430JZX, 22C1432JZX; 9CN90H/20884521176352/21C0411JZX, 21C0412JZX, 21D0464JZX, 21D0465JZX, 21D0601JZX, 21I0555JZX, 21I0556JZX, 21K0262JZX, 21K0263JZX; 9CN90R/10884521205468/202105281X, 202110114X, 202202327X, 202204166X, 202205514X; 9UN90H/20884521176284/21E0657JZX, 21E0658JZX; 9UN90R/10884521205536/202203308X, 21F0328JZX

Devices named in this recall

38 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

11 devices cleared under K142296, K150844 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.