Skip to content

Shiley 4UN65RRecalls & reported problems

Covidien LP · Adult Flexible Tracheostomy Tube Cuffless Reusable Inner Cannula · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10884521205482), received between Aug 12, 2022 and Jun 28, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Human-Device Interface Problem” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K150844); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 24, 2023Open, Classified

    A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to respiratory failure, dyspnea, treatment delay, tissue injury, or bleeding.

    Z-1357-2023 · Root cause: Component change control · FDA record ↗

Most reported problems

Device problems

  • Human-Device Interface Problem2
  • Use of Device Problem2
  • Patient-Device Incompatibility1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Unintended Extubation2
  • Insufficient Information1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 28, 2024Human-Device Interface Problem; Human-Device Interface Problem

    ACCORDING TO THE REPORTER, DURING USE, THE CANNULA WAS CLEANED WITH A 10 MM NYLON BRUSH WITH A SHARP METAL END. IT WAS FOUND OUT THAT THERE WAS A SCRATCH AND TEAR AT THE TIP OF THE INNER CANNULA, AS SHOWN IN THE ATTACHED PHOTO. THE REPLACEMENT OF THE TUBE WAS DONE.

    Patient problems: Unintended Extubation

    MAUDE report 19638427 ↗
  • MalfunctionNov 28, 2023Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE'S FLANGE BROKE DURING USED. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18222935 ↗
  • MalfunctionOct 12, 2023Human-Device Interface Problem; Human-Device Interface Problem

    ACCORDING TO THE REPORTER, THE SPEAKING VALVE WOULD NOT WORK ON THE DEVICE. IT DID NOT FIT SECURELY. THERE WAS NO ALLEGATION OF PATIENT DEATH OR SERIOUS INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17922822 ↗
  • MalfunctionAug 9, 2023Patient-Device Incompatibility; Patient-Device Incompatibility

    ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE WAS NOT HOLDING UP AND KEPT FALLING. THERE WAS NO REPORTED PATIENT OUTCOME.

    Patient problems: Insufficient Information

    MAUDE report 17493222 ↗
  • MalfunctionAug 12, 2022Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, DURING USE, THE INNER CANNULA WAS CRACKED ON THE SECOND CLEANING. THE PATIENT WAS DECANNULATED.

    Patient problems: Unintended Extubation

    MAUDE report 15218533 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K150844. One submission often covers many models, so check the product and lot details in each record.

  • Z-1390-2025 · Feb 26, 2025 · Open, Classified
    Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula, REF: 7CN80R

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.