Shiley 9UN90RRecalls & reported problems

Covidien LP · Adult Flexible Tracheostomy Tube Cuffless Reusable Inner Cannula · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10884521205536), received between Jun 25, 2024 and Jun 25, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Human-Device Interface Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K150844); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 24, 2023Open, Classified

    A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to respiratory failure, dyspnea, treatment delay, tissue injury, or bleeding.

    Z-1357-2023 · Root cause: Component change control · FDA record ↗

Most reported problems

Device problems

  • Human-Device Interface Problem1

Patient problems

  • Airway Obstruction1
  • Discomfort1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 25, 2024Human-Device Interface Problem; Human-Device Interface Problem

    ACCORDING TO THE REPORTER, DURING USE, THE SIZE IN THE BOX OF THE FIRST DEVICE WAS INCORRECT. THE PATIENT USED A CALIPER TO CHECK THE SIZE AND FOUND THAT THE SIZES LISTED ON THE BOX WERE INCORRECT. THE PATIENT MENTIONED THAT IT WAS MUCH SMALLER THAN THE PREVIOUS VERSION, AND THE INNER AND OUTER DIAMETERS WERE NOT THE SAME. THE PATIENT AIRWAY WAS RESTRICTED AND NOISY. THE PATIENT WENT UP TO A 9-MM TUBE, WHICH WAS CLOSER TO THE SIZE OF THE FIRST DEVICE, BUT STILL FELT MORE NARROWED, WHICH CAUSED ADDED NOISE AND VOLUME RESTRICTION, AND PREFERRED TO HAVE THE OLD STYLE BACK.

    Patient problems: Airway Obstruction; Discomfort

    MAUDE report 19607292 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K150844. One submission often covers many models, so check the product and lot details in each record.

  • Z-1390-2025 · Feb 26, 2025 · Open, Classified
    Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula, REF: 7CN80R

Other Tube tracheostomy and tube cuff devices