Newport Medical Instruments, Inc. · e360 Ventilator with Accessories;E Model · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884521540958), received between Apr 20, 2017 and May 30, 2017 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Battery Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Battery Problem1
  • Defective Component1
  • Insufficient Device Problem Information1

Patient problems

  • No Information2
  • No Patient Involvement1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 30, 2017Defective Component; Defective Component

    IT WAS REPORTED THAT THE E360 VENTILATOR'S AIR SENSOR OF THE MOTHERBOARD WAS DEFECTIVE. THERE WAS NO PATIENT INVOLVEMENT. THE MOTHERBOARD WAS REPLACED AND A 5000K PREVENTATIVE MAINTENANCE (PM) WAS PERFORMED ALONG WITH CALIBRATIONS AND ALL TESTS PASSED.

    Patient problems: No Patient Involvement

    MAUDE report 6600107 ↗
  • MalfunctionApr 20, 2017Battery Problem; Battery Problem

    IT WAS REPORTED THAT E360 VENTILATOR HAD A BAD BATTERY. PATIENT INVOLVEMENT IS UNKNOWN. THE SERVICE ENGINEER (SE) INSPECTED THE DEVICE AND REPLACED THE BATTERY. THE SE PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED. THE SERIAL NUMBER OF THE VENTILATOR WAS NOT PROVIDED THEREFORE THE DEVICE MANUFACTURE DATE IS UNKNOWN.

    Patient problems: No Information

    MAUDE report 6508000 ↗
  • MalfunctionApr 20, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT E360 VENTILATOR HAD A DEFECTIVE OXYGEN SENSOR. PATIENT INVOLVEMENT IS UNKNOWN. THE SERVICE ENGINEER (SE) INSPECTED THE DEVICE AND REPLACED THE OXYGEN SENSOR. THE SE PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED. THE SERIAL NUMBER OF THE VENTILATOR WAS NOT PROVIDED THEREFORE THE DEVICE MANUFACTURE DATE IS UNKNOWN.

    Patient problems: No Information

    MAUDE report 6507967 ↗