Newport E360S-WW-INRecalls & reported problems
Newport Medical Instruments, Inc. · e360 Ventilator with Accessories;S Model · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10884521541122), received between Apr 10, 2017 and Aug 9, 2017 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Leak/Splash2
- Appropriate Device Problem Term/Code Not Available1
- Device Displays Incorrect Message1
- Device Inoperable1
- Device Sensing Problem1
- Failure to Calibrate1
- No Display/Image1
Patient problems
- No Patient Involvement6
- No Information3
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 9, 2017No Display/Image; Device Displays Incorrect Message; No Display/Image; Device Displays Incorrect Message
IT WAS REPORTED THAT E360 VENTILATOR HAD A BLANK DISPLAY. ALTHOUGH REQUESTED, INFORMATION PERTAINING TO THE USE OF THE DEVICE AT THE TIME OF THE REPORTED EVENT AND PATIENT OUTCOME (IF APPLICABLE) HAS NOT BEEN PROVIDED. MEDTRONIC WAS NOT AUTHORIZED TO EVALUATE/REPAIR THE DEVICE AS THE PRODUCT IS NOT IN MEDTRONIC CONTROL. THE REPAIR WAS COMPLETED AT A NON-MEDTRONIC LOCATION AND THE SERVICE PROVIDER (SP) FOUND THE SINGLE BOARD COMPUTER AND AN O2 SENSOR ERROR NEED TO BE REPLACED. THE REPORTED COMPLAINT COULD NOT BE CONFIRMED WITHOUT THE PRODUCT RETURN. SHOULD NEW INFORMATION BECOME AVAILABLE OR IF THE PRODUCT IS RETURNED TO THE MANUFACTURER FOR INVESTIGATION THE COMPLAINT WILL BE RE-EVALUATED.
Patient problems: No Information; No Patient Involvement
MAUDE report 6778103 ↗MalfunctionAug 8, 2017Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE E360 VENTILATOR HAD A LEAKAGE. THERE WAS NO PATIENT INVOLVEMENT AT THE TIME OF THE EVENT. MEDTRONIC WAS NOT AUTHORIZED TO EVALUATE/REPAIR THE DEVICE AS THE PRODUCT IS NOT IN MEDTRONIC CONTROL. THE REPAIR WAS COMPLETED AT A NON-MEDTRONIC LOCATION AND THE SERVICE PROVIDER (SP) FOUND A LEAKAGE IN AIR TUBING. THE SP RECONNECTED ALL TUBING TO ADDRESS THE ISSUE. THE REPORTED COMPLAINT COULD NOT BE CONFIRMED WITHOUT THE PRODUCT RETURN. SHOULD NEW INFORMATION BECOME AVAILABLE OR IF THE PRODUCT IS RETURNED TO THE MANUFACTURER FOR INVESTIGATION THE COMPLAINT WILL BE RE-EVALUATED.
Patient problems: No Patient Involvement
MAUDE report 6774343 ↗MalfunctionJul 31, 2017Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED THAT THE E360 VENTILATOR FAILED CALIBRATION FOR FLOW SENSOR. THERE WAS NO PATIENT INVOLVEMENT AT THE TIME OF THE EVENT. MEDTRONIC WAS NOT AUTHORIZED TO EVALUATE/REPAIR THE DEVICE AS THE PRODUCT IS NOT IN MEDTRONIC CONTROL. THE REPAIR WILL BE COMPLETED AT A NON-MEDTRONIC LOCATION. THE REPORTED COMPLAINT COULD NOT BE CONFIRMED WITHOUT THE PRODUCT RETURN. SHOULD NEW INFORMATION BECOME AVAILABLE OR IF THE PRODUCT IS RETURNED TO THE MANUFACTURER FOR INVESTIGATION THE COMPLAINT WILL BE RE-EVALUATED.
Patient problems: No Patient Involvement
MAUDE report 6754150 ↗MalfunctionJun 16, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE E360 VENTILATOR GENERATED A OXYGEN SUPPLY LOSS ALARM. THERE WAS NO PATIENT INVOLVEMENT AT THE TIME OF THE REPORTED CONDITION. COVIDIEN/MEDTRONIC WAS NOT AUTHORIZED TO EVALUATE/SERVICE THE DEVICE SO THE REPORTED COMPLAINT COULD NOT BE CONFIRMED. A NON-COVIDIEN /MEDTRONIC SERVICE PROVIDER CHECKED THE VENTILATOR AND FOUND REPORTED PROBLEM. THE SERVICE PROVIDER CROSSCHECKED SERVO VALVE WITH WORKING VENTILATOR AND CONFIRMED IT WAS DETERMINED THAT THE SERVO VALVE NEEDS TO BE REPLACED.
Patient problems: No Patient Involvement
MAUDE report 6648935 ↗MalfunctionJun 7, 2017Leak/Splash; Leak/Splash
ADDITIONAL INFORMATION WAS RECEIVED THAT THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT
Patient problems: No Information; No Patient Involvement
MAUDE report 6618635 ↗MalfunctionMay 3, 2017Device Inoperable; Device Inoperable
IT WAS REPORTED THAT THE E360 VENTILATOR HAD INCOMPATIBLE DISPLAY BOARD SOFTWARE AND DOES NOT ACCEPT ANY COMMAND. PATIENT INVOLVEMENT INFORMATION WAS NOT PROVIDED BY THE CUSTOMER. MEDTRONIC WAS NOT AUTHORIZED TO EVALUATE/SERVICE THE DEVICE SO THE REPORTED COMPLAINT COULD NOT BE CONFIRMED. THE VENTILATOR WAS EVALUATED AND IT WAS REPORTED THAT THE SINGLE BOARD COMPUTER AND ITS CONNECTIONS WERE CLEANED. THE VENTILATOR IS IN WORKING CONDITION.
Patient problems: No Information
MAUDE report 6541252 ↗MalfunctionApr 10, 2017Failure to Calibrate; Failure to Calibrate
IT WAS REPORTED THAT THERE WAS A MALFUNCTION OF THE OXYGEN SENSOR. THE VENTILATOR WAS NOT ON A PATIENT AT THE TIME OF THE EVENT. REPAIR WILL BE COMPLETED AT A NON-MEDTRONIC LOCATION AND THE PRODUCT IS NOT IN MEDTRONIC CONTROL. THE REPORTED COMPLAINT COULD NOT BE CONFIRMED WITHOUT THE PRODUCT RETURN. THE COMPLAINT WILL BE CLOSED AND REOPENED/UPDATED SHOULD NEW INFORMATION BECOME AVAILABLE OR IF THE PRODUCT IS RETURNED TO THE MANUFACTURER FOR INVESTIGATION. THE INFORMATION HAS BEEN ADDED TO THE DATABASE FOR TRENDING PURPOSES AND TRENDS WILL CONTINUE TO BE MONITORED. A CAPA IS NOT REQUIRED AT THIS TIME.
Patient problems: No Patient Involvement
MAUDE report 6475592 ↗
Other VENTILATOR, CONTINUOUS, FACILITY USE devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page