Newport HT70M-JP-NA (UDI 10884521542075)Recalls & reported problems
Newport Medical Instruments, Inc. · HT70 Ventilator With Accessories;O2 Monitor · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Newport HT70 ventilator is a facility-use breathing machine for patients. The 89 reports mostly describe airflow stopping, the unit shutting down unexpectedly, and alarms that were too quiet or not visible, sometimes while the ventilator was in use on a patient. There are no deaths, 1 injury, and no recalls name this device. No reports were filed in the past 12 months, compared with 21 in the period before that.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 89
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 89 adverse event reports that list this exact device (DI 10884521542075), received between Aug 18, 2017 and Oct 16, 2025 — classified as 1 injury, 88 malfunctions.
The most frequently reported specific device problem is “No Flow” (66 reports).
4 other recalls cover devices cleared under the same FDA submission (K111146); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- No Flow66
- Defective Alarm20
- No Audible Alarm3
- Alarm Not Visible2
- Decrease in Pressure2
- Device Inoperable2
- Battery Problem1
- Device Displays Incorrect Message1
Patient problems
- No Clinical Signs, Symptoms or Conditions53
- No Consequences Or Impact To Patient17
- No Code Available14
- Low Oxygen Saturation9
- Insufficient Information3
- Cardiac Arrest1
- No Information1
Recent adverse event reports
The 25 most recent of 89 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 16, 2025No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE HT70 VENTILATOR GENERATED A CONTINUOUS ALARM AND VENTILATION STOPPED. IT WAS ALSO REPORTED THAT THE UNIT HAD NO DISPLAY ON THE SCREEN AND THE UNIT COULD NOT BE RESTARTED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND WAS MANUALLY VENTILATED VIA AN AMBU BAG WITHOUT INJURY. REVIEW OF THE VENTILATOR LOGS CONFIRMED REPORTED VENTILATOR SHUTDOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23316553 ↗MalfunctionJun 9, 2025Defective Alarm; Defective Alarm
IT WAS REPORTED THAT DURING SERVICING, THE SHUTDOWN ALARM "HARDLY SOUNDED" ON THIS HT70 VENTILATOR. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22172550 ↗MalfunctionJun 9, 2025Defective Alarm; Defective Alarm
IT WAS REPORTED THAT DURING SERVICING BY MEDTRONIC SERVICE PERSONNEL, THIS HT70 VENTILATOR WAS NOTED TO HAVE A SHUT DOWN BUZZER ALARM TEST FAILURE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22168135 ↗MalfunctionJun 9, 2025Defective Alarm; Defective Alarm
IT WAS REPORTED THAT DURING SERVICING, THE SHUTDOWN ALARM OF THIS HT70 VENTILATOR "ALMOST" DID NOT SOUND WHEN THE POWER WAS TURNED OFF. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22167848 ↗MalfunctionJun 9, 2025Defective Alarm; Defective Alarm
IT WAS REPORTED THAT DURING SERVICING, THIS HT70 VENTILATOR WAS NOTED TO HAVE A SHUTDOWN ALARM THAT HARDLY SOUNDED. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22167680 ↗MalfunctionJun 9, 2025Defective Alarm; Defective Alarm
IT WAS REPORTED THAT DURING SERVICING, THE SHUTDOWN ALARM OF THIS HT70 VENTILATOR "HARDLY SOUNDED" WHEN THE POWER WAS TURNED OFF. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22167495 ↗MalfunctionJun 9, 2025Defective Alarm; Defective Alarm
IT WAS REPORTED THAT DURING SERVICING, IT WAS OBSERVED THAT THE SHUTDOWN ALARM OF THIS HT70 VENTILATOR BARELY SOUNDED WHEN THE POWER WAS TURNED OFF. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22167386 ↗MalfunctionJun 7, 2025Defective Alarm; Defective Alarm
IT WAS REPORTED THAT DURING SERVICING, THE GENERATED SHUTDOWN ALARM OF THIS HT70 VENTILATOR WAS SHORT. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22163162 ↗MalfunctionJun 7, 2025Defective Alarm; Defective Alarm
IT WAS REPORTED THAT DURING SERVICING, THE SHUTDOWN ALARM GENERATED BY THIS HT70 VENTILATOR SOUNDED IN "SHORT TIME" WHEN THE UNIT WAS POWERED OFF. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22162380 ↗MalfunctionJun 6, 2025Defective Alarm; Defective Alarm
IT WAS REPORTED THAT DURING SERVICING, THE HT70 VENTILATOR'S SHUTDOWN ALARM WAS OBSERVED TO BE TOO SHORT. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22161844 ↗MalfunctionJun 6, 2025Defective Alarm; Defective Alarm
IT WAS REPORTED THAT DURING SERVICING, THE SHUTDOWN ALARM OF THIS HT70 VENTILATOR BARELY SOUNDED WHEN THE POWER WAS TURNED OFF. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22161590 ↗MalfunctionJun 6, 2025Defective Alarm; Defective Alarm
IT WAS REPORTED THAT DURING SERVICING, IT WAS OBSERVED THAT THE SHUTDOWN ALARM HARDLY SOUNDED ON THIS HT70 VENTILATOR. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22161417 ↗MalfunctionJun 5, 2025No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE HT70 VENTILATOR GENERATED AN ABNORMAL ALARM THEN UNIT SHUT DOWN AND THE VENTILATION STOPPED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22155118 ↗MalfunctionMay 15, 2025No Flow; Power Problem; No Flow; Power Problem
IT WAS REPORTED THAT THE HT70 VENTILATOR WOULD NOT START. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22032097 ↗MalfunctionMar 26, 2025No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE HT70 VENTILATOR GENERATED A SHUTDOWN ALARM WITH A LOW SOUND, FOLLOWED BY A SYSTEM ERROR DISPLAYED ON THE SCREEN. ALTHOUGH THE UNIT'S SCREEN REMAINED OPERATIONAL, THE VENTILATION STOPPED, AND WAS NOT ABLE TO PERFORM VENTILATION. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21701167 ↗MalfunctionMar 19, 2025No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE HT70 VENTILATOR WAS CONTINUOUSLY GENERATING A SHUTDOWN ALARM, THE SCREEN DISPLAYED A SYSTEM ERROR, APPEARED THAT NO AIR WAS BEING AND THE PATIENT'S OXYGEN (O2) SATURATION DROPPED. THERE WAS NO PATIENT INJURY.
Patient problems: Low Oxygen Saturation
MAUDE report 21649046 ↗MalfunctionMar 12, 2025No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE SCREEN OF THIS HT70 VENTILATOR WAS FROZEN, THE UNIT GENERATED AN "OPERATIONAL MALFUNCTION" ALARM AND THE VENTILATION STOPPED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND WAS PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY. A REVIEW OF THE VENTILATOR LOGS CONFIRM THE REPORTED LOSS OF VENTILATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21587286 ↗MalfunctionMar 12, 2025No Flow; No Flow
IT WAS REPORTED THAT THE HT70 VENTILATOR GENERATED A SHUTDOWN ALARM AND THE VENTILATION STOPPED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21586730 ↗MalfunctionFeb 25, 2025No Flow; No Flow
IT WAS REPORTED THAT THE HT70 VENTILATOR GENERATED A SYSTEM ERROR ALARM AND THE VENTILATION STOPPED WHILE THE PATIENT WAS CONNECTED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND WAS PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21461036 ↗MalfunctionFeb 13, 2025Defective Alarm; No Flow; Pressure Problem; Defective Alarm; No Flow; Pressure Problem
IT WAS REPORTED THAT DURING A POSITION CHANGE, THE PEAK AIRWAY PRESSURE OF THIS HT70 VENTILATOR WAS LOWER THAN USUAL, NO ALARM SOUND WAS GENERATED AND THE UNIT STOPPED VENTILATING. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21379782 ↗MalfunctionJan 8, 2025No Flow; No Flow
UPDATED: IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE HT70 VENTILATOR ALARMED AND GENERATED A SYSTEM ERROR MESSAGE AND VENTILATION STOPPED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND WAS PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21107343 ↗MalfunctionOct 14, 2024No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE VENTILATION STOPPED ON THIS HT70 VENTILATOR AND IT COULD NOT BE ACTIVATED AND THE UNIT ALSO GENERATED AND DISPLAYED A "SYSTEM ERROR" MESSAGE. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20447306 ↗MalfunctionAug 14, 2024No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE GRAPHICAL USER INTERFACE OF THIS HT70 VENTILATOR FROZE AND VENTILATION STOPPED. THERE WAS NO INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19986610 ↗MalfunctionMar 26, 2024No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE VENTILATION SUDDENLY STOPPED ON THIS HT70 VENTILATOR AND THE UNIT SHUT DOWN. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18980578 ↗MalfunctionJan 26, 2024No Flow; No Flow
IT WAS REPORTED THAT THE HT70 VENTILATOR GENERATED A SHUTDOWN ALARM, THE UNIT SCREEN WENT OFF AND THE VENTILATION STOPPED WHILE THE PATIENT WAS CONNECTED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND WAS PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18588575 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111146. One submission often covers many models, so check the product and lot details in each record.
- Z-1915-2025 · May 21, 2025 · Open, ClassifiedNewport HT70 Ventilator, REF: HT70M-JP-NA, HT70M-CN-NA, HT70M-ES-EU, HT70M-PT-BR, HT70M-SY-EU, HT70M-WW-EU, HT70M-WW-NA, HT70-SY-EU. Newport HT70Plus Ventilator, REF: HT70PM-ES-EU
- Z-1874-2017 · Mar 30, 2017 · TerminatedNewport Medical HT70 and HT70- Plus Ventilators, Rx Only Product Usage: The Newport HT70 family of ventilators is intended to provide continuous or intermittent positive pressure
- Z-1171-2013 · Apr 5, 2013 · TerminatedNewport Medical Instruments HT70 and HT70 Plus ventilator Power Pac batteries Model number: HT70 and HT70 Plus. The HT70 and HT70 Plus ventilator systems are intended to provide
- Z-2250-2012 · Aug 3, 2012 · TerminatedNewport HT70 and HT70 Plus Ventilators, Model Number: HT70 and HT70 Plus. Product Usage: The HT70 ventilator system is intended to provide continuous or intermittent positive pre
Other VENTILATOR, CONTINUOUS, FACILITY USE devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page