Newport HT70M-WW-EURecalls & reported problems
Newport Medical Instruments, Inc. · HT70 Ventilator With Accessories;O2 Monitor · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10884521542136), received between Aug 17, 2017 and Aug 17, 2017 — classified as 1 malfunction.
The most frequently reported specific device problem is “Leak/Splash” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K111146); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- May 21, 2025Class IOpen, Classified
Ventilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation, or the shutdown alert alarm fails to alarm effectively during shut down, which may result in respiratory failure, hypoventilation, low oxygen saturation, hypoxia, treatment delay.
Z-1915-2025 · Root cause: Component design/selection · Recall details · FDA record ↗
Most reported problems
Device problems
- Leak/Splash1
Patient problems
- No Patient Involvement1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 17, 2017Leak/Splash; Leak/Splash
IT WAS REPORTED THAT DURING TESTING THE HT70 VENTILATOR HAD AN INTERNAL LEAKAGE. THERE WAS NO PATIENT INVOLVEMENT AT THE TIME OF THE EVENT. MEDTRONIC WAS NOT AUTHORIZED TO EVALUATE/REPAIR THE DEVICE AS THE PRODUCT IS NOT IN MEDTRONIC CONTROL. THE REPAIR WILL BE COMPLETED AT A NON-MEDTRONIC LOCATION. THE REPORTED COMPLAINT COULD NOT BE CONFIRMED WITHOUT THE PRODUCT RETURN. SHOULD NEW INFORMATION BECOME AVAILABLE OR IF THE PRODUCT IS RETURNED TO THE MANUFACTURER FOR INVESTIGATION THE COMPLAINT WILL BE RE-EVALUATED.
Patient problems: No Patient Involvement
MAUDE report 6802391 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K111146. One submission often covers many models, so check the product and lot details in each record.
- Z-1874-2017 · Mar 30, 2017 · TerminatedNewport Medical HT70 and HT70- Plus Ventilators, Rx Only Product Usage: The Newport HT70 family of ventilators is intended to provide continuous or intermittent positive pressure
- Z-1171-2013 · Apr 5, 2013 · TerminatedNewport Medical Instruments HT70 and HT70 Plus ventilator Power Pac batteries Model number: HT70 and HT70 Plus. The HT70 and HT70 Plus ventilator systems are intended to provide
- Z-2250-2012 · Aug 3, 2012 · TerminatedNewport HT70 and HT70 Plus Ventilators, Model Number: HT70 and HT70 Plus. Product Usage: The HT70 ventilator system is intended to provide continuous or intermittent positive pre
Other VENTILATOR, CONTINUOUS, FACILITY USE devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page