Newport HT70PM-JP-NA (UDI 10884521544321)Recalls & reported problems

Newport Medical Instruments, Inc. · HT70 Ventilator with Accessories Plus Model · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Newport HT70 ventilator is a facility-use breathing machine. The 46 reports mostly describe the unit failing to deliver air flow or stopping and shutting down while in use, along with alarms that did not sound when expected;1 death and no injuries were reported. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
46
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 46 adverse event reports that list this exact device (DI 10884521544321), received between Sep 5, 2019 and Apr 29, 2026 — classified as 1 death, 45 malfunctions.

The most frequently reported specific device problem is “No Flow” (32 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 1 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 4 reportsJul 2025: 2 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Flow32
  • Defective Alarm15
  • Accessory Incompatible2
  • Infusion or Flow Problem2
  • Mechanical Problem2
  • Power Problem2
  • Battery Problem1
  • Data Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions39
  • Low Oxygen Saturation4
  • Insufficient Information2
  • No Consequences Or Impact To Patient2
  • Respiratory Failure1
  • Tachycardia1

Recent adverse event reports

The 25 most recent of 46 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2026Defective Alarm

    IT WAS REPORTED THAT DURING SERVICING PERFORMED BY MEDTRONIC, THE HT70 VENTILATOR WAS OBSERVED TO NOT PASS THE SHUTDOWN BUZZER ALARM TEST. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25037963 ↗
  • MalfunctionFeb 12, 2026Defective Alarm; Output Problem

    IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE HT70 VENTILATOR GENERATED AN OXYGEN (O2) CONCENTRATION ALARM. THE VENTILATOR WAS CONTINUED TO BE USED ON THE PATIENT WITHOUT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24327370 ↗
  • MalfunctionDec 9, 2025No Flow; No Flow

    IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE VENTILATION STOPPED ON HIS HT70 VENTILATOR WITH A CONSTANT ALARM SOUNDED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23762880 ↗
  • MalfunctionAug 13, 2025No Flow; No Flow

    IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE VENTILATION STOPPED ON THIS HT70 VENTILATOR, THE SCREEN DISAPPEARED AND A CONTINUOUS ALARM SOUND OCCURRED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22797815 ↗
  • MalfunctionAug 12, 2025No Flow; No Flow

    IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE HT70 VENTILATOR GENERATED A CONTINUOUS ALARM SOUND, THE SCREEN WENT COMPLETELY BLACK AND THE UNIT SHUTDOWN. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND WAS PLACED IN AN ALTERNATE VENTILATOR WITHOUT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22786739 ↗
  • MalfunctionJul 25, 2025Defective Alarm; Defective Alarm

    IT WAS REPORTED THAT THE ALARM RELATED TO THE CAPACITOR OF THIS HT70 VENTILATOR, WHICH WAS SUPPOSED TO ACTIVATE WHEN TURNING HE POWER ON AND OFF, DID NOT SOUND. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22616881 ↗
  • MalfunctionJul 9, 2025No Flow; Power Problem; No Flow; Power Problem

    IT WAS REPORTED THAT DURING SERVICING, THE HT70 VENTILATOR WAS OBSERVED TO SUDDENLY SHUT DOWN FROM STANDBY MODE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT. REPORTABILITY REASSESSED BASED ON THE PRODUCT ANALYSIS FINDINGS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22461358 ↗
  • MalfunctionJun 26, 2025Defective Alarm; Defective Alarm

    IT WAS REPORTED THAT THE HT70 VENTILATOR GENERATED A BACKGROUND CHECK ALERT "ASSERT ERROR" MESSAGE AND THE UNIT'S ALARM SOUNDED FOR A MOMENT, THEN THE DEVICE SILENCED ON ITS OWN WHEN THE POWER WAS SWITCHED OFF. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22337659 ↗
  • MalfunctionJun 6, 2025Defective Alarm; Defective Alarm

    IT WAS REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) PERFORMED BY THE MEDTRONIC PERSONNEL, THIS HT70 VENTILATOR WAS OBSERVED TO HAVE FAILED THE SHUTDOWN BUZZER ALARM TEST. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22161638 ↗
  • MalfunctionJun 6, 2025Defective Alarm; Defective Alarm

    IT WAS REPORTED THAT DURING SERVICING, THE ALARM GENERATED BY THIS HT70 VENTILATOR AFTER SHUTTING DOWN WAS NOTED TO HAVE AN "ABNORMAL" SOUND. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22161487 ↗
  • MalfunctionJun 6, 2025Defective Alarm; Infusion or Flow Problem; Defective Alarm; Infusion or Flow Problem

    IT WAS REPORTED THAT DURING SERVICING BY MEDTRONIC SERVICE PERSONNEL, THE PUMP OF THIS HT70 VENTILATOR WAS NOTED TO HAVE A LEAK, THE POSITIVE END EXPIRATORY PRESSURE (PEEP) WAS FOUND TO BE UNSTABLE AND THE CONTROL BOARD  SHUTDOWN BUZZER WAS NOTED TO HAVE A FAILURE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22160981 ↗
  • MalfunctionApr 16, 2025No Flow; No Flow

    IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE HT70 VENTILATOR GENERATED A CONTINUOUS ALARM AND THE VENTILATION STOPPED. THE OXYGEN (O2) SATURATION OF THE PATIENT TEMPORARILY DECREASED TO 89%. THE PATIENT WAS MANUALLY VENTILATED AND TRANSFERRED TO AN ALTERNATE VENTILATOR WITHOUT INJURY. THE PATIENT RECOVERED FROM THE O2 SATURATION DROP WITHIN TWO MINUTES.

    Patient problems: Low Oxygen Saturation

    MAUDE report 21848672 ↗
  • DeathFeb 12, 2025No Flow; No Flow

    IT WAS REPORTED THAT WHILE IN USE ON AN ELDERLY, TERMINALLY ILL PATIENT, THE HT70 VENTILATOR GENERATED AN ALARM AND VENTILATION STOPPED. WHILE BEING TRANSFERRED TO AN ALTERNATE VENTILATOR, THE PATIENT EXPERIENCED CARDIAC ARREST AND CARDIOPULMONARY RESUSCITATION (CPR) WAS PERFORMED, BUT THE PATIENT DID NOT SURVIVE.

    Patient problems: Respiratory Failure

    MAUDE report 21368834 ↗
  • MalfunctionDec 12, 2024No Flow; No Flow

    IT WAS REPORTED THAT WHILE IN USE ON A PATIENT AT HOME, THIS HT70 VENTILATOR GENERATED A "SYSTEM ERROR" AND THE VENTILATION STOPPED. A CONTINUOUS ALARM WAS GENERATED AND THE POWER WAS UNABLE TO BE TURNED OFF. THERE WAS NO INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20919294 ↗
  • MalfunctionAug 14, 2024No Flow; No Flow

    IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THIS HT70 VENTILATOR HAD A SYSTEM ERROR, THE LIQUID CRYSTAL DISPLAY (LCD) BECAME DARK, AND VENTILATION STOPPED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND WAS PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19983957 ↗
  • MalfunctionJun 26, 2024No Flow; No Flow

    IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THIS HT70 VENTILATOR HAD A SYSTEM ERROR. THERE WAS NO INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19615209 ↗
  • MalfunctionJun 20, 2024No Flow; No Flow

    IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE HT70 VENTILATOR GENERATED A SHUTDOWN ALARM AND VENTILATION STOPPED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND WAS PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19578385 ↗
  • MalfunctionJun 18, 2024No Flow; No Flow

    IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THIS HT70 VENTILATOR GENERATED A BEEPING ALARM SOUNDS AND A SYSTEM ERROR WAS DISPLAYED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19564070 ↗
  • MalfunctionMay 8, 2024Defective Alarm; No Flow; Defective Alarm; No Flow

    ADDITIONAL INFORMATION RECEIVED AS FOLLOWS: IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THIS HT70 VENTILATOR DID NOT ALARM, SHUTDOWN AND VENTILATION STOPPED. THE PATIENT WAS PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19274868 ↗
  • MalfunctionApr 22, 2024No Flow; No Flow

    IT WAS REPORTED THAT VENTILATION STOPPED WHILE THIS HT70 VENTILATOR WAS IN USE ON A PATIENT. THE PATIENT'S PERIPHERAL CAPILLARY OXYGEN SATURATION (SPO2) LEVEL DROPPED TO 50% AND THEN IMPROVED AFTER PLACING THE PATIENT ON AN ALTERNATE VENTILATOR.

    Patient problems: Low Oxygen Saturation

    MAUDE report 19153662 ↗
  • MalfunctionApr 11, 2024Device Sensing Problem; No Flow; Power Problem; Device Sensing Problem; No Flow; Power Problem

    IT WAS REPORTED THAT WHILE VENTILATING ON THE TEST LUNG, THE HT70 VENTILATOR GENERATED AN ALARM AND THE POWER WAS FOUND TO HAVE SHUT DOWN. ADDITIONALLY, THE MUTE AND RESTART BUTTONS WERE NOT WORKING. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19093637 ↗
  • MalfunctionDec 8, 2023No Flow; No Flow

    IT WAS REPORTED THAT THE HT70 VENTILATOR GENERATED A SYSTEM ERROR AND VENTILATION STOPPED WHILE THE PATIENT WAS CONNECTED. THE OXYGEN SATURATION OF THE PATIENT DROPPED TO 91% AND HEART RATE WAS 131 BPM. ADDITIONALLY, THE ALARM COULD NOT BE SILENCED AND THE POWER COULD NOT BE TURNED OFF. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND WAS PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.

    Patient problems: Low Oxygen Saturation; Tachycardia

    MAUDE report 18293937 ↗
  • MalfunctionOct 9, 2023No Flow; No Flow

    IT WAS REPORTED THAT THE HT70 VENTILATOR GENERATED A CONTINUOUS BEEPING SOUND, THE DISPLAY ON THE MONITOR DISAPPEARED AND THE VENTILATION STOPPED WHILE THE PATIENT WAS CONNECTED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND WAS PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17898016 ↗
  • MalfunctionAug 18, 2023No Flow; No Flow

    IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THIS HT70 VENTILATOR HAD A BLACK SCREEN, GENERATED AN ALARM AND VENTILATION STOPPED THEN SHUTDOWN. THE PATIENT WAS PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17577476 ↗
  • MalfunctionJul 31, 2023Defective Alarm; Defective Alarm

    IT WAS REPORTED THAT DURING SERVICING, THE SHUTDOWN BUZZER ALARM OF THIS HT70 VENTILATOR DID NOT SOUND. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17426537 ↗

Other VENTILATOR, CONTINUOUS, FACILITY USE devices

All VENTILATOR, CONTINUOUS, FACILITY USE devices →