Newport HT70PM-JP-NA (UDI 10884521544321)Recalls & reported problems
Newport Medical Instruments, Inc. · HT70 Ventilator with Accessories Plus Model · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Newport HT70 ventilator is a facility-use breathing machine. The 46 reports mostly describe the unit failing to deliver air flow or stopping and shutting down while in use, along with alarms that did not sound when expected;1 death and no injuries were reported. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 46
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 46 adverse event reports that list this exact device (DI 10884521544321), received between Sep 5, 2019 and Apr 29, 2026 — classified as 1 death, 45 malfunctions.
The most frequently reported specific device problem is “No Flow” (32 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- No Flow32
- Defective Alarm15
- Accessory Incompatible2
- Infusion or Flow Problem2
- Mechanical Problem2
- Power Problem2
- Battery Problem1
- Data Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions39
- Low Oxygen Saturation4
- Insufficient Information2
- No Consequences Or Impact To Patient2
- Respiratory Failure1
- Tachycardia1
Recent adverse event reports
The 25 most recent of 46 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 29, 2026Defective Alarm
IT WAS REPORTED THAT DURING SERVICING PERFORMED BY MEDTRONIC, THE HT70 VENTILATOR WAS OBSERVED TO NOT PASS THE SHUTDOWN BUZZER ALARM TEST. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25037963 ↗MalfunctionFeb 12, 2026Defective Alarm; Output Problem
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE HT70 VENTILATOR GENERATED AN OXYGEN (O2) CONCENTRATION ALARM. THE VENTILATOR WAS CONTINUED TO BE USED ON THE PATIENT WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24327370 ↗MalfunctionDec 9, 2025No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE VENTILATION STOPPED ON HIS HT70 VENTILATOR WITH A CONSTANT ALARM SOUNDED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23762880 ↗MalfunctionAug 13, 2025No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE VENTILATION STOPPED ON THIS HT70 VENTILATOR, THE SCREEN DISAPPEARED AND A CONTINUOUS ALARM SOUND OCCURRED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22797815 ↗MalfunctionAug 12, 2025No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE HT70 VENTILATOR GENERATED A CONTINUOUS ALARM SOUND, THE SCREEN WENT COMPLETELY BLACK AND THE UNIT SHUTDOWN. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND WAS PLACED IN AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22786739 ↗MalfunctionJul 25, 2025Defective Alarm; Defective Alarm
IT WAS REPORTED THAT THE ALARM RELATED TO THE CAPACITOR OF THIS HT70 VENTILATOR, WHICH WAS SUPPOSED TO ACTIVATE WHEN TURNING HE POWER ON AND OFF, DID NOT SOUND. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22616881 ↗MalfunctionJul 9, 2025No Flow; Power Problem; No Flow; Power Problem
IT WAS REPORTED THAT DURING SERVICING, THE HT70 VENTILATOR WAS OBSERVED TO SUDDENLY SHUT DOWN FROM STANDBY MODE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT. REPORTABILITY REASSESSED BASED ON THE PRODUCT ANALYSIS FINDINGS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22461358 ↗MalfunctionJun 26, 2025Defective Alarm; Defective Alarm
IT WAS REPORTED THAT THE HT70 VENTILATOR GENERATED A BACKGROUND CHECK ALERT "ASSERT ERROR" MESSAGE AND THE UNIT'S ALARM SOUNDED FOR A MOMENT, THEN THE DEVICE SILENCED ON ITS OWN WHEN THE POWER WAS SWITCHED OFF. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22337659 ↗MalfunctionJun 6, 2025Defective Alarm; Defective Alarm
IT WAS REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) PERFORMED BY THE MEDTRONIC PERSONNEL, THIS HT70 VENTILATOR WAS OBSERVED TO HAVE FAILED THE SHUTDOWN BUZZER ALARM TEST. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22161638 ↗MalfunctionJun 6, 2025Defective Alarm; Defective Alarm
IT WAS REPORTED THAT DURING SERVICING, THE ALARM GENERATED BY THIS HT70 VENTILATOR AFTER SHUTTING DOWN WAS NOTED TO HAVE AN "ABNORMAL" SOUND. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22161487 ↗MalfunctionJun 6, 2025Defective Alarm; Infusion or Flow Problem; Defective Alarm; Infusion or Flow Problem
IT WAS REPORTED THAT DURING SERVICING BY MEDTRONIC SERVICE PERSONNEL, THE PUMP OF THIS HT70 VENTILATOR WAS NOTED TO HAVE A LEAK, THE POSITIVE END EXPIRATORY PRESSURE (PEEP) WAS FOUND TO BE UNSTABLE AND THE CONTROL BOARD SHUTDOWN BUZZER WAS NOTED TO HAVE A FAILURE. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22160981 ↗MalfunctionApr 16, 2025No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE HT70 VENTILATOR GENERATED A CONTINUOUS ALARM AND THE VENTILATION STOPPED. THE OXYGEN (O2) SATURATION OF THE PATIENT TEMPORARILY DECREASED TO 89%. THE PATIENT WAS MANUALLY VENTILATED AND TRANSFERRED TO AN ALTERNATE VENTILATOR WITHOUT INJURY. THE PATIENT RECOVERED FROM THE O2 SATURATION DROP WITHIN TWO MINUTES.
Patient problems: Low Oxygen Saturation
MAUDE report 21848672 ↗DeathFeb 12, 2025No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON AN ELDERLY, TERMINALLY ILL PATIENT, THE HT70 VENTILATOR GENERATED AN ALARM AND VENTILATION STOPPED. WHILE BEING TRANSFERRED TO AN ALTERNATE VENTILATOR, THE PATIENT EXPERIENCED CARDIAC ARREST AND CARDIOPULMONARY RESUSCITATION (CPR) WAS PERFORMED, BUT THE PATIENT DID NOT SURVIVE.
Patient problems: Respiratory Failure
MAUDE report 21368834 ↗MalfunctionDec 12, 2024No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT AT HOME, THIS HT70 VENTILATOR GENERATED A "SYSTEM ERROR" AND THE VENTILATION STOPPED. A CONTINUOUS ALARM WAS GENERATED AND THE POWER WAS UNABLE TO BE TURNED OFF. THERE WAS NO INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20919294 ↗MalfunctionAug 14, 2024No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THIS HT70 VENTILATOR HAD A SYSTEM ERROR, THE LIQUID CRYSTAL DISPLAY (LCD) BECAME DARK, AND VENTILATION STOPPED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND WAS PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19983957 ↗MalfunctionJun 26, 2024No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THIS HT70 VENTILATOR HAD A SYSTEM ERROR. THERE WAS NO INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19615209 ↗MalfunctionJun 20, 2024No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THE HT70 VENTILATOR GENERATED A SHUTDOWN ALARM AND VENTILATION STOPPED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND WAS PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19578385 ↗MalfunctionJun 18, 2024No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THIS HT70 VENTILATOR GENERATED A BEEPING ALARM SOUNDS AND A SYSTEM ERROR WAS DISPLAYED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19564070 ↗MalfunctionMay 8, 2024Defective Alarm; No Flow; Defective Alarm; No Flow
ADDITIONAL INFORMATION RECEIVED AS FOLLOWS: IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THIS HT70 VENTILATOR DID NOT ALARM, SHUTDOWN AND VENTILATION STOPPED. THE PATIENT WAS PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19274868 ↗MalfunctionApr 22, 2024No Flow; No Flow
IT WAS REPORTED THAT VENTILATION STOPPED WHILE THIS HT70 VENTILATOR WAS IN USE ON A PATIENT. THE PATIENT'S PERIPHERAL CAPILLARY OXYGEN SATURATION (SPO2) LEVEL DROPPED TO 50% AND THEN IMPROVED AFTER PLACING THE PATIENT ON AN ALTERNATE VENTILATOR.
Patient problems: Low Oxygen Saturation
MAUDE report 19153662 ↗MalfunctionApr 11, 2024Device Sensing Problem; No Flow; Power Problem; Device Sensing Problem; No Flow; Power Problem
IT WAS REPORTED THAT WHILE VENTILATING ON THE TEST LUNG, THE HT70 VENTILATOR GENERATED AN ALARM AND THE POWER WAS FOUND TO HAVE SHUT DOWN. ADDITIONALLY, THE MUTE AND RESTART BUTTONS WERE NOT WORKING. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19093637 ↗MalfunctionDec 8, 2023No Flow; No Flow
IT WAS REPORTED THAT THE HT70 VENTILATOR GENERATED A SYSTEM ERROR AND VENTILATION STOPPED WHILE THE PATIENT WAS CONNECTED. THE OXYGEN SATURATION OF THE PATIENT DROPPED TO 91% AND HEART RATE WAS 131 BPM. ADDITIONALLY, THE ALARM COULD NOT BE SILENCED AND THE POWER COULD NOT BE TURNED OFF. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND WAS PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: Low Oxygen Saturation; Tachycardia
MAUDE report 18293937 ↗MalfunctionOct 9, 2023No Flow; No Flow
IT WAS REPORTED THAT THE HT70 VENTILATOR GENERATED A CONTINUOUS BEEPING SOUND, THE DISPLAY ON THE MONITOR DISAPPEARED AND THE VENTILATION STOPPED WHILE THE PATIENT WAS CONNECTED. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND WAS PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17898016 ↗MalfunctionAug 18, 2023No Flow; No Flow
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, THIS HT70 VENTILATOR HAD A BLACK SCREEN, GENERATED AN ALARM AND VENTILATION STOPPED THEN SHUTDOWN. THE PATIENT WAS PLACED ON AN ALTERNATE VENTILATOR WITHOUT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17577476 ↗MalfunctionJul 31, 2023Defective Alarm; Defective Alarm
IT WAS REPORTED THAT DURING SERVICING, THE SHUTDOWN BUZZER ALARM OF THIS HT70 VENTILATOR DID NOT SOUND. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17426537 ↗
Other VENTILATOR, CONTINUOUS, FACILITY USE devices
- Trilogy Trilogy100 Ventilator, United StatesRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE92 deaths8,161 reports
- Trilogy Evo Trilogy Evo, USARespironics, Inc. · Ventilator, continuous, facility use6 recalls50 deaths4,393 reports
- Trilogy 100 Trilogy 100 W/ Bluetooth, USARespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE65 deaths4,569 reports
- CARESCAPE R860Datex-Ohmeda Inc · VENTILATOR, CONTINUOUS, FACILITY USE1 recall1 death4,346 reports
- HAMILTON-C1 161001Hamilton Medical AG · Ventilator, Continuous, Facility Use2 recalls12 deaths1,849 reports
- Trilogy Trilogy O2 Ventilator JapanRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE7 deaths1,900 reports
- Trilogy Trilogy 202 Ventilator InternationalRespironics, Inc. · VENTILATOR, CONTINUOUS, FACILITY USE11 deaths1,564 reports
- HAMILTON-C6 160021Hamilton Medical AG · Ventilator, Continuous, Facility Use5 recalls10 deaths1,175 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page