Skip to content

Newport HT70PM-SY-EU (UDI 10884521544369)Recalls & reported problems

Newport Medical Instruments, Inc. · HT70 Ventilator with Accessories Plus Model · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10884521544369), received between Aug 11, 2017 and Aug 11, 2023 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Defective Alarm” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. May 21, 2025Class IOpen, Classified

    Ventilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation, or the shutdown alert alarm fails to alarm effectively during shut down, which may result in respiratory failure, hypoventilation, low oxygen saturation, hypoxia, treatment delay.

    Z-1915-2025 · Root cause: Component design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Defective Alarm1
  • No Display/Image1
  • No Flow1

Patient problems

  • Coma1
  • No Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 11, 2023Defective Alarm; No Flow; Defective Alarm; No Flow

    IT WAS REPORTED THAT WHILE IN USE ON A HOME CARE PATIENT, THIS HT70 VENTILATOR SUDDENLY STOPPED WORKING AND THERE WAS NO ALARM OR CODES GENERATED. THE PATIENT TURNED BLUE, AND WAS UNCONSCIOUS AFTERWARDS. THE PATIENT WAS REMOVED FROM THE VENTILATOR AND MANUALLY VENTILATED VIA AN AMBU BAG. WHEN THE AMBULANCE ARRIVED, THE PATIENT'S OXYGEN (O2) SATURATION WAS AT 50 %, AND THE END-TIDAL CARBON DIOXIDE (ETCO2) WAS AT 80 MMHG. THE PATIENT WAS PLACED ON AN ALTERNATE VENTILATOR AND THE O2 SATURATION AND CARBON DIOXIDE (CO2) IMPROVED BUT THE PATIENT REMAINED UNRESPONSIVE. THE PATIENT WAS ADMITTED TO THE HOSPITAL DUE TO DROP IN GLASGOW COMA SCALE.

    Patient problems: Coma

    MAUDE report 17514570 ↗
  • MalfunctionAug 11, 2017No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE HT70 PLUS VENTILATOR HAD BLANK SCREEN. PATIENT INVOLVEMENT WAS NOT REPORTED. MEDTRONIC WAS NOT AUTHORIZED TO EVALUATE/REPAIR THE DEVICE AS THE PRODUCT IS NOT IN MEDTRONIC CONTROL. THE REPAIR WILL BE COMPLETED AT A NON-MEDTRONIC LOCATION. THE REPORTED COMPLAINT COULD NOT BE CONFIRMED WITHOUT THE PRODUCT RETURN. SHOULD NEW INFORMATION BECOME AVAILABLE OR IF THE PRODUCT IS RETURNED TO THE MANUFACTURER FOR INVESTIGATION THE COMPLAINT WILL BE RE-EVALUATED.

    Patient problems: No Information

    MAUDE report 6789169 ↗

Other VENTILATOR, CONTINUOUS, FACILITY USE devices

All VENTILATOR, CONTINUOUS, FACILITY USE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page