VersaOne NONB12STFRecalls & reported problems
Covidien LP · Bladeless Trocar with Fixation Cannula · Class 2
All models of VersaOne (47)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Covidien VersaOne trocar is a surgical port used to create an entry point for laparoscopic instruments during keyhole surgery. Of its 129 reports, 125 are malfunctions and 4 are injuries, with no deaths; many describe gas leaking from the device, and others describe torn seals, packaging problems and parts coming loose. Reports fell to 9 in the last 12 months from 20 in the 12 months before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 128
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 128 adverse event reports that list this exact device (DI 10884521546882), received between Dec 9, 2016 and Mar 3, 2026 — classified as 4 injuries, 124 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (61 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash61
- Material Split, Cut or Torn35
- Packaging Problem35
- Detachment of Device or Device Component32
- Break24
- Decrease in Pressure15
- Detachment Of Device Component7
- Gas/Air Leak6
Patient problems
- No Clinical Signs, Symptoms or Conditions74
- No Known Impact Or Consequence To Patient26
- No Consequences Or Impact To Patient19
- Insufficient Information8
- No Patient Involvement4
- Foreign Body In Patient3
- Blood Loss1
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 25 most recent of 128 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 3, 2026Leak/Splash; Decrease in Pressure
IT WAS REPORTED THAT DURING A LAPAROSCOPIC HYSTERECTOMY, THE DEVICE EXHIBITED MATERIAL LEAKAGE THAT COMPROMISED PNEUMOPERITONEUM. IT WAS NOTED THAT THERE WAS A RAPID LOSS OF PNEUMOPERITONEUM NECESSITATING IMMEDIATE PRODUCT REPLACEMENT TO RESOLVE THE ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24497342 ↗MalfunctionFeb 18, 2026Leak/Splash; Decrease in Pressure
IT WAS REPORTED THAT DURING A PROCEDURE, THE TROCAR LEAKED WHICH RESULTED IN A RAPID LOSS OF PNEUMOPERITONEUM. TO RESOLVE THE ISSUE, A NEW DEVICE OF THE SAME TYPE WAS USED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24379826 ↗MalfunctionDec 22, 2025Leak/Splash; Decrease in Pressure; Leak/Splash; Decrease in Pressure
IT WAS REPORTED THAT DURING A VENTRAL LAPAROSCOPIC (VLP) RECTOSIGMOIDECTOMY, HEAVY LEAKING FROM THE TROCAR OCCURRED AT THE BEGINNING OF SURGERY WHEN THE TROCAR WAS INSERTED. THIS RESULTED IN A RAPID LOSS OF PNEUMOPERITONEUM. THE TROCAR WAS REPLACED TO ALLOW THE PR OCEDURE TO CONTINUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23865659 ↗MalfunctionDec 17, 2025Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING A PROCEDURE, THE TROCAR WAS LEAKING GAS, AND THE SEAL ON THE DEVICE WAS DAMAGED.
Patient problems: Insufficient Information
MAUDE report 23826413 ↗MalfunctionDec 5, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, INTRA-OPERATIVELY, AIR OR GAS LEAKED FROM THE SEAL OF MULTIPLE TROCARS. THE LEAKING SOUND COULD BE HEARD FROM ONE OF THE TROCARS AND ONE TROCAR HAD A LOOSE SEAL. THERE WAS NO PATIENT INJURY.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23730384 ↗MalfunctionDec 5, 2025Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE TROCAR EXHIBITED A DEFECT IN THE CARBON DIOXIDE (CO2) CONNECTOR, RESULTING IN A CO2 LEAK FROM THE TROCAR. THE CANNULA MEMBRANE WAS FOUND TO BE PERFORATED, WHICH CONTRIBUTED TO THE CO2 LEAK. UPON TESTING, THE CANNULA MEMBRANE PERFORATION WAS VERIFIED. THE TROCAR WAS REPLACED WITH A COMPLETE NEW UNIT TO RESOLVE THE ISSUE AND ALLOW THE PROCEDURE TO CONTINUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23726078 ↗MalfunctionNov 5, 2025Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING A LAPAROSCOPIC GASTROPLASTY BYPASS, THE TROCAR WAS LEAKING PROFUSELY AT THE BEGINNING OF THE PROCEDURE AND THE SEAL WAS DAMAGED. IT WAS NECESSARY TO OPEN ANOTHER TROCAR TO COMPLETE THE SURGERY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23473996 ↗MalfunctionNov 4, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING A LAPAROSCOPIC SLEEVE GASTRECTOMY, WHILE THE DEVICE WAS BEING PLACED IN PATIENT, THE SURGEON REALIZED AN AIR OR GAS LEAKED FROM THE SEAL DUE TO A BROKEN VALVE SEAL. THE NURSE OPENED ANOTHER NEW TROCAR TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23468845 ↗MalfunctionOct 20, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING A LAPAROSCOPIC GALLBLADDER SURGERY, TWO TROCARS WERE FOUND TO HAVE A DAMAGED SEAL AND THE VALVE WAS NOT TIGHT, RESULTING IN GAS LEAKAGE DURING THE INSUFFLATION STEP OF THE PROCEDURE. THE ISSUE WAS RESOLVED BY REPLACING THE PRODUCT WITH ANOTHER MANUFACTURER'S DEVICE, AND THE CASE WAS COMPLETED WITHOUT ADDITIONAL OPERATIONS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23341259 ↗MalfunctionOct 20, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING A LAPAROSCOPIC GALLBLADDER SURGERY, TWO TROCARS WERE FOUND TO HAVE A DAMAGED SEAL AND THE VALVE WAS NOT TIGHT, RESULTING IN GAS LEAKAGE DURING THE INSUFFLATION STEP OF THE PROCEDURE. THE ISSUE WAS RESOLVED BY REPLACING THE PRODUCT WITH ANOTHER MANUFACTURER'S DEVICE, AND THE CASE WAS COMPLETED WITHOUT ADDITIONAL OPERATIONS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23341161 ↗MalfunctionOct 15, 2025Leak/Splash; Detachment of Device or Device Component; Material Split, Cut or Torn; Leak/Splash; Detachment of Device or Device Component; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, IN A LAPAROSCOPIC APPENDECTOMY, THE SEAL OF THE PORT WAS DAMAGED AND DISENGAGED WHICH CAUSED AIR OR GAS TO LEAK FROM THE SEAL. THE CANNULA ALSO DETACHED FROM THE TOP PORTION OF THE PORT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23297557 ↗MalfunctionSep 8, 2025Difficult to Insert; Difficult to Insert
IT WAS REPORTED THAT THERE WAS DIFFICULTY PENETRATING AND FAILURE OF THE SEALING MECHANISM WHEN PASSING THE CLIPPER. THE CLAMP WOULD FALL DUE TO THE FORCE USED TO PASS THE CLIPPER THROUGH THE TROCAR, AND THE SEAL WOULD BECOME OBLITERATED. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22999479 ↗MalfunctionSep 8, 2025Break; Break
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC HERNIA, AT THE START OF THE PROCEDURE, THE SURGEON NOTICED THAT THE TROCAR WAS CRACKED. A NEW DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22993870 ↗MalfunctionAug 5, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A LAPAROSCOPIC APPENDICECTOMY, THE TROCAR BROKE AND A MATERIAL FROM THE TROCAR COMPONENT DEVICE FELL INTO THE CAVITY OF THE PATIENT. THE MATERIAL WAS RETRIEVED AND AN X-RAY WAS NOT NECESSARY TO LOCATE OR CONFIRM RETRIEVAL OF THE COMPONENT. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22717391 ↗MalfunctionJun 19, 2025Difficult to Insert; Difficult to Insert
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC GASTROPLASTY BYPASS, THE SURGEON USED A CLIPPER BUT IT DID NOT FIT IN THE TROCAR. ANOTHER DEVICE FROM OTHER MANUFACTURER WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22294324 ↗MalfunctionJun 10, 2025Break; Break
IT WAS REPORTED THAT DURING THE LAPAROSCOPIC CHOLECYSTECTOMY, AT THE START OF THE SURGERY, THE TROCAR WAS BROKEN THE ISSUE WAS RESOLVED BY REPLACING THE TROCAR. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22175382 ↗MalfunctionMay 21, 2025Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING A VIDEO LAPAROSCOPIC TOTAL SPLENECTOMY, THERE WAS A GAS LEAK FROM THE TROCAR AND THE SEAL WAS DAMAGED. THE ISSUE WAS RESOLVED BY REPLACING THE DEVICE WITH ANOTHER IDENTICAL ITEM. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22058970 ↗MalfunctionMay 12, 2025Detachment of Device or Device Component; Material Split, Cut or Torn; Detachment of Device or Device Component; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING A LAPAROSCOPIC PROCEDURE, WHILE PASSING A FORCEPS THROUGH THE DEVICE, THE EXTERNAL VALVE WAS LOOSE AND TORN. IT THEN FELL INTO THE ABDOMINAL CAVITY OF THE PATIENT AND WAS REMOVED WITHOUT FURTHER DAMAGE. THE PRODUCT WAS REPLACED TO RESOLVE THE ISSUE AND COMPLETE THE CASE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22006886 ↗MalfunctionMay 5, 2025Leak/Splash; Decrease in Pressure; Leak/Splash; Decrease in Pressure
ACCORDING TO THE REPORTER, IN A LAPAROSCOPIC GASTROPLASTY BYPASS, THE TROCAR LEAKED AFTER THE ENDOSCOPIC STAPLER WAS PASSED. IT WAS IMPOSSIBLE TO CONTINUE USING THE TROCAR, BECAUSE WHEN USING THE 5MM FORCEPS, THE PNEUMOPERITONEUM WAS LOWERED AND LOST RAPIDLY. ANOTHER TYPE OF TROCAR WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21958209 ↗MalfunctionApr 30, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, IN A VIDEO LAPAROSCOPIC GASTROPLASTY BYPASS WAS, WHEN TRYING TO PASS THE FORCEPS THROUGH THE TROCAR, THE RUBBER SEAL TORE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21939869 ↗MalfunctionApr 30, 2025Decrease in Pressure; Detachment of Device or Device Component; Decrease in Pressure; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A LAPAROSCOPIC APPENDECTOMY, THERE WAS A LOSS OF PNEUMOPERITONEUM DUE TO A LOOSE EXTERNAL SEAL WHILE USING THE TROCAR. THE ISSUE WAS IDENTIFIED WHEN PASSING THE ENDOSTAPLER, AND AIR OR GAS LEAKED FROM THE SEAL. THE PRODUCT WAS REPLACED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21934497 ↗MalfunctionApr 22, 2025Break; Break
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE MATERIAL BROKE, NECESSITATING THE USE OF ANOTHER TROCAR TO CONTINUE THE SURGERY, WHICH RESULTED IN A DELAY OF THE ENTIRE SURGICAL PROCESS.
Patient problems: Insufficient Information
MAUDE report 21886338 ↗MalfunctionApr 22, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, THE RUBBER OF THE TROCAR CAME OFF.
Patient problems: Insufficient Information
MAUDE report 21884748 ↗MalfunctionApr 21, 2025Leak/Splash; Detachment of Device or Device Component; Leak/Splash; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A LAPAROSCOPIC PROCEDURE, THE RUBBER ON THE TROCAR CAME LOOSE WHEN THE CLIPPER PASSED THROUGH, AND AIR OR GAS LEAKED FROM THE SEAL. THE PRODUCT WAS REPLACED BY ANOTHER ONE TO RESOLVE THE ISSUE AND COMPLETE THE CASE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21874739 ↗MalfunctionMar 7, 2025Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, IN A LAPAROSCOPIC GASTRIC BYPASS, AFTER ABOUT 40 MINUTES, TWO TROCARS BEGAN TO LEAK, CAUSING THE LOSS OF THE PNEUMOPERITONEUM. THE SEALS OF TWO TROCARS WERE DAMAGED. THE DEVICES WERE NOT REPLACED. THE SURGICAL TIME EXTENDED 30 MINUTES OR MORE DUE TO THE LOSS OF PNEUMOPERITONEUM. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21547531 ↗
FDA clearance and classification
Cleared through Special 510(k) K151548 on Jul 2, 2015 (decision: Substantially Equivalent). Submitted by Covidien as “VersaOne Bladed Trocar, VersaOne Bladeless Trocar”. The FDA review took 23 days. FDA record ↗
The FDA regulates product code GCJ (laparoscope, general & plastic surgery) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Common questions about the VersaOne NONB12STF
Has the VersaOne NONB12STF been recalled?
No FDA recall record in our data names this exact device identifier (DI 10884521546882), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the VersaOne NONB12STF?
The FDA's MAUDE database lists 128 adverse event reports for this device, received between Dec 9, 2016 and Mar 3, 2026. In those reports, the most common specific device problems are “Leak/Splash” (61), “Material Split, Cut or Torn” (35) and “Packaging Problem” (35); the most common patient problems are “Foreign Body In Patient” (3), “Blood Loss” (1) and “Device Embedded In Tissue or Plaque” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the VersaOne NONB12STF?
None of the reports are classified as deaths, 4 as injuries and 124 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the VersaOne NONB12STF FDA approved?
It was cleared through Special 510(k) K151548 on Jul 2, 2015. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the VersaOne NONB12STF, and is it still sold?
The FDA's device database lists Covidien LP as the company responsible for it (brand VersaOne, model NONB12STF). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,913 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,857 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,068 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall945 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery850 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery770 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall694 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery569 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page