Emprint CAGEN1ZDRecalls & reported problems
Covidien LP · Ablation Generator with Thermospher Technology · Class 2
All models of Emprint (19)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10884521552579), received between Feb 22, 2021 and Dec 30, 2021 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem4
- Device Displays Incorrect Message3
- Failure to Power Up1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Insufficient Information1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 30, 2021Device Displays Incorrect Message; Adverse Event Without Identified Device or Use Problem; Device Displays Incorrect Message; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING ABLATION PROCEDURE, THEY ATTEMPTED ABLATION FOR 75 WATTS AT 4 MINUTES. THE GENERATOR HAD THE ANTENNA TEMPERATURE SENSOR ERROR WHILE DOING AN ABLATION THAT ENDING AT 3 MINUTES INSTEAD OF 4 MINUTES. CONFIRMED SALINE WAS FLOWING. CUSTOMER WAS NOT ABLE TO COMPLETE THE PROCEDURE AND IT WAS ABORTED AFTER THE HARD REST (UNPLUGGING ALL POWER CORDS, ANTENNAS AND THEN PLUGGING EVERYTHING BACK IN TO POWER ON). THEY PRESS THE RESET BUTTON AND IT STILL CONTINUED TO SHOW THE TEMP ALERT ERROR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13117846 ↗MalfunctionDec 1, 2021Device Displays Incorrect Message; Adverse Event Without Identified Device or Use Problem; Device Displays Incorrect Message; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE UNIT HAD AN ERROR AND HAD TO BE RESET SEVERAL TIMES. THE PROCEDURE WAS CANCELLED. THE PATIENT WAS NOT UNDER ANESTHESIA AT THE TIME BUT WAS UNDER CONSCIOUS SEDATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12908340 ↗MalfunctionMar 11, 2021Device Displays Incorrect Message; Adverse Event Without Identified Device or Use Problem; Device Displays Incorrect Message; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING MICROWAVE ABLATION OF A LIVER TUMOR, AN ERROR MESSAGE "FATAL ERROR" SHOWED IN THE UNIT. IT WAS THE ERROR THE HAND AND THE BOX WITH AN ARROW COMING OUT AND IT WOULD NOT ALLOW TO RESET IT. THE PROCEDURE WAS ABORTED AND RE-SCHEDULED ON THE NEXT DAY WITH NEW GENERATOR. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11463705 ↗MalfunctionFeb 22, 2021Failure to Power Up; Adverse Event Without Identified Device or Use Problem; Failure to Power Up; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, TURNED ON WATER PUMP AND GENERATOR PRIOR TO THE PATIENT GETTING ON THE TABLE AND BOTH WORKED. WHEN THE CASE STARTED AND THEY WENT TO HOOK UP THE ANTENNA TO THE WATER PUMP AND THE GENERATOR, THE WATER PUMP DID NOT TURN ON, THEY TRIED A DIFFERENT CORD AND A DIFFERENT OUTLET BUT NEITHER WORKED. THEY HAD TO WAIT AT LEAST 45 MIN TO FIND ANOTHER PUMP. IN THE MEAN TIME THE PATIENT WAS ON THE TABLE AND SEDATED. NO PATIENT INJURY.
Patient problems: Insufficient Information
MAUDE report 11358592 ↗
Other SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES devices
- Emprint CA20L2Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES2 recalls1 death43 reports
- Emprint CA15L2Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES2 recalls1 death41 reports
- Emprint CA30L2Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES2 recalls2 reports
- Emprint CA15L1Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES1 recall49 reports
- Solero 700106001-USAngiodynamics, Inc. · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES92 reports
- NEUWAVE PR15XTNeuwave Medical, Inc. · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES90 reports
- Emprint CA20L1Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES1 recall18 reports
- Emprint CA30L1Covidien LP · SYSTEM, ABLATION, MICROWAVE AND ACCESSORIES1 recall1 death9 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page