Skip to content

TruClear 7209509Recalls & reported problems

Covidien LP · Soft Tissue Shaver Plus · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10884521744103), received between Nov 20, 2023 and May 7, 2026 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Connection Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Appropriate Device Problem Term/Code Not Available1
  • Connection Problem1
  • Detachment of Device or Device Component1
  • Failure to Cut1
  • Mechanical Jam1

Patient problems

  • Insufficient Information2
  • Adhesion(s)1
  • Hemorrhage/Blood Loss/Bleeding1
  • Incomplete Induced Abortion1
  • Internal Organ Perforation1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 7, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT PRIOR TO A HYSTEROSCOPY AND CURETTAGE PROCEDURE, UNIDENTIFIED FRAGMENTS WERE OBSERVED FALLING FROM THE SHAVER TIP. THEY IMMEDIATELY REPLACED THE DEVICE WITH ANOTHER BRAND-NEW SHAVER TO COMPLETE THE PROCEDURE. NOTHING FELL INTO THE PATIENT¿S CAVITY. THERE WAS NO INVASION OF THE PATIENT AT THE TIME THE FRAGMENTS WERE OBSERVED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25115894 ↗
  • MalfunctionFeb 3, 2026Failure to Cut; Mechanical Jam; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A HYSTEROSCOPY POLYPECTOMY THE SOFT TISSUE SHAVER WAS NOT DISSECTING PROPERLY, APPEARED SLOW, AND SEEMED BLUNT. THE DENSE TISSUE SHAVER PLUS WAS THEN USED, BUT THE SITUATION DID NOT IMPROVE. A REUSABLE HAND INSTRUMENT WAS ALSO TRIED, RESULTING IN ONLY A VERY SLIGHT IMPROVEMENT. THE POLYP BEING REMOVED WAS OVER 3 CM IN SIZE. DUE TO THE PERFORMANCE OF THE DEVICES, THE PROCEDURE HAD TO BE COMPLETED USING BIPOLAR RESECTION, WHICH WAS NOT INITIALLY AVAILABLE IN THE HOSPITAL. THIS SWITCH EXTENDED THE SURGICAL TIME BY ONE HOUR.

    Patient problems: Insufficient Information

    MAUDE report 24244684 ↗
  • InjuryJul 29, 2025Connection Problem; Appropriate Device Problem Term/Code Not Available; Connection Problem; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT DURING A PROCEDURE, THE PLUS SHAVER HAD A DETECTION ISSUE AS THEY OPENED A WRONG SCOPE. THEY HAD A MINI SCOPE INSTEAD OF A PLUS SCOPE WHICH WAS NOT COMPATIBLE WITH THE PLUS SHAVER. THE PATIENT WAS ALREADY PLACED UNDER ANESTHESIA AND THE PROCEDURE WAS CANCELLED BECAUSE THERE WAS NO PLUS SCOPE AVAILABLE. THE PROCEDURE WAS RESCHEDULED ON A LATER DATE.

    Patient problems: Insufficient Information

    MAUDE report 22642522 ↗
  • InjuryNov 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE LITERATURE, A PROSPECTIVE STUDY COMPARED INTRAUTERINE ADHESION FORMATION IN PATIENTS WHO UNDERWENT HYSTEROSCOPIC REMOVAL OF RETAINED PRODUCTS OF CONCEPTION OR ULTRASOUND-GUIDED ELECTRIC VACUUM ASPIRATION BETWEEN APRIL 2015 UNTIL JUNE 2022. HYSTEROSCOPIC REMOVAL WAS PERFORMED WITH THE TRUCLEAR 8.0 SYSTEM OR ELITE HYSTEROSCOPE PLUS WITH A SOFT TISSUE SHAVER PLUS OR A COMPETITOR DEVICE. FLUID BALANCE WAS MONITORED USING THE HYSTEROLUX OR A COMPETITOR DEVICE. HYSTEROSCOPES WERE INSERTED UNDER DIRECT VISION AFTER DILATING THE UTERINE CERVIX WITH DILATORS. OF 178 INCLUDED WOMEN, 92 WERE TREATED WITH HYSTEROSCOPY. COMPLICATIONS IN THE HYSTEROSCOPY GROUP INCLUDED HEMORRHAGE, PERFORATION, PERSISTENT RETAINED PRODUCTS OF CONCEPTION, AND ADHESIONS. CONVERSION TO VACUUM ASPIRATION WAS REQUIRED DUE TO HEMORRHAGE (GREATER THAN 500ML) AND BAD VISIBILITY ± POOR DISTENTION/RAPID FLUID LOSS. PERFORATION OCCURRED AT THE CESAREAN SCAR WHICH HAD MINIMAL MYOMETRIUM OVERLAY, AND RESULTED IN A SMALL BLADDER PERFORATION WHICH WAS CONFIRMED WITH CYSTOSCOPY AND TREATED WITH A BLADDER CATHETER FOR 10 DAYS. REPEAT HYSTEROSCOPY WAS REQUIRED FOR PATIENTS WITH PERSISTENT RETAINED CONTENTS AND ADH

    Patient problems: Adhesion(s); Hemorrhage/Blood Loss/Bleeding; Incomplete Induced Abortion; Internal Organ Perforation

    MAUDE report 18175452 ↗

Other HYSTEROSCOPE (AND ACCESSORIES) devices

All HYSTEROSCOPE (AND ACCESSORIES) devices →