TruClear 7209807SRecalls & reported problems
Covidien LP · Handpiece, Refurbished · Class 2
All models of TruClear (20)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10884521749900), received between Feb 26, 2018 and May 5, 2025 — classified as 2 injuries, 8 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break5
- Adverse Event Without Identified Device or Use Problem1
- Appropriate Device Problem Term/Code Not Available1
- Connection Problem1
- Device Damaged Prior to Use1
- Device Displays Incorrect Message1
- Intermittent Energy Output1
- Material Deformation1
Patient problems
- No Patient Involvement5
- No Information2
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 5, 2025Overheating of Device; Smoking; Device Displays Incorrect Message; Adverse Event Without Identified Device or Use Problem; Overheating of Device; Smoking; Device Displays Incorrect Message; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A HYSTEROSCOPY/POLYPECTOMY PROCEDURE FOR TISSUE REMOVAL, THE HANDPIECE BECAME EXTREMELY HOT AND DISPLAYED AN E6 ERROR CODE. THE DEVICE EMITTED A BURNING SMELL AND WAS SUBSEQUENTLY REMOVED FROM THE OPERATING ROOM, CAUSING AN HOUR DELAYED TO THE PROCEDURE. THERE WAS NO MEDICAL OR SURGICAL INTERVENTION NEEDED TO PREVENT A PERMANENT IMPAIRMENT OF A FUNCTION. ANOTHER HANDPIECE WAS USED, AND IT WORKED FINE. THE PROCEDURE WAS COMPLETED AS IS WITH DILATION AND CURETTAGE (DC).
Patient problems: Insufficient Information
MAUDE report 21961078 ↗InjuryJul 19, 2023Appropriate Device Problem Term/Code Not Available; Intermittent Energy Output; Appropriate Device Problem Term/Code Not Available; Intermittent Energy Output
ACCORDING TO THE REPORTER, DURING A HYSTEROSCOPY MYOMECTOMY, THE HANDPIECE AND BLADE JUST STOPPED WORKING AND CONTROL UNIT SAID "STALLED MOTOR, TRY NEW HANDPIECE". THE SURGICAL TIME WAS EXTENDED 30 MINUTES OR MORE DUE TO THE PRODUCT PROBLEM. HAND TO SWITCH OUT HANDPIECES AND BLADES AND CASE WAS ABORTED AFTER TWO BLADES AND TWO HANDPIECES WERE ATTEMPTED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17356541 ↗MalfunctionMar 27, 2019Connection Problem; Connection Problem
ACCORDING TO THE REPORTER, THE PORT WHERE TUBING ATTACHES HAS DETACHED.
Patient problems: No Information
MAUDE report 8457215 ↗MalfunctionMar 27, 2019Break; Break
ACCORDING TO THE REPORTER, DURING REPROCESSING, DEVICE HAD BROKEN CONNECTOR. NO PATIENT INVOLVED.
Patient problems: No Patient Involvement
MAUDE report 8457193 ↗InjuryMar 21, 2019Break; Material Deformation; Break; Material Deformation
ACCORDING TO THE REPORTER, THERE WAS A BROKEN PIECE MISSING OFF OF UNIT. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Information
MAUDE report 8440346 ↗MalfunctionJan 28, 2019Material Protrusion/Extrusion; Material Protrusion/Extrusion
ACCORDING TO THE REPORTER, INTRA-OPERATIVELY, THE HAND-PIECE WAS CAUSING THE BLADE TO SKIP WHEN OPERATING. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8280781 ↗MalfunctionDec 3, 2018Material Fragmentation; Material Fragmentation
ACCORDING TO THE REPORTER, PRE-OPERATIVELY, THE UNIT'S COATING/INSIDE WAS CHIPPING OFF AND WAS MISSING. NO PATIENT INVOLVED.
Patient problems: No Patient Involvement
MAUDE report 8127537 ↗MalfunctionOct 12, 2018Break; Device Damaged Prior to Use; Break; Device Damaged Prior to Use
ACCORDING TO THE REPORTER, DURING TESTING, THE METAL PIECE SNAPPED OUT ON THE SHAFT OF THE HAND PIECE. THERE WAS NO PATIENT INVOLVEMENT. NO FURTHER INFORMATION AVAILABLE.
Patient problems: No Patient Involvement
MAUDE report 7960437 ↗MalfunctionFeb 28, 2018Break; Break
ACCORDING TO THE REPORTER, THE DEVICE WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Patient Involvement
MAUDE report 7304458 ↗MalfunctionFeb 26, 2018Break; Break
ACCORDING TO THE REPORTER, THE UNIT ADAPTER BROKE OFF. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 7297855 ↗
FDA clearance and classification
The FDA regulates product code HIH (hysteroscope (and accessories)) under 21 CFR 884.1690 as a Class II device. 21 CFR 884.1690 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the TruClear 7209807S
Has the TruClear 7209807S been recalled?
No FDA recall record in our data names this exact device identifier (DI 10884521749900), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the TruClear 7209807S?
The FDA's MAUDE database lists 10 adverse event reports for this device, received between Feb 26, 2018 and May 5, 2025. In those reports, the most common specific device problems are “Break” (5), “Connection Problem” (1) and “Device Damaged Prior to Use” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the TruClear 7209807S?
None of the reports are classified as deaths, 2 as injuries and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the TruClear 7209807S, and is it still sold?
The FDA's device database lists Covidien LP as the company responsible for it (brand TruClear, model 7209807S). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other HYSTEROSCOPE (AND ACCESSORIES) devices
- Olympus WA2T430AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)783 reports
- Olympus WA2T412AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)452 reports
- Olympus A22001AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)366 reports
- Olympus WA2T470AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)323 reports
- Olympus A22002AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)297 reports
- Olympus A22003AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)198 reports
- n.a. 27040XAKarl Storz GmbH & Co. KG · HYSTEROSCOPE (AND ACCESSORIES)102 reports
- VISERA HYSTEROVIDEOSCOPE OLYMPUS HYF TYPE VOlympus Medical Systems Corp. · HYSTEROSCOPE (AND ACCESSORIES)1 recall48 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page