Skip to content

TruClear 7209808SRecalls & reported problems

Covidien LP · Control Unit, Refurbished · Class 2

All models of TruClear (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884521749917), received between Jan 10, 2018 and Sep 5, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Intermittent Energy Output1
  • Mechanics Altered1

Patient problems

  • No Patient Involvement2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 5, 2024Intermittent Energy Output; Intermittent Energy Output

    ACCORDING TO THE REPORTER, INTRA-OPERATIVELY, THE UNIT ITSELF SHORTED OUT AND DOES NOT WORK. THE HANDPIECES WOULD NOT WORK WHEN PLUGGED INTO THIS UNIT. THERE WAS ANOTHER CONTROL UNIT AT THE FACILITY AND THAT UNIT WORKED AND WERE ABLE TO RULE OUT A HANDPIECE OR FOOT PEDAL ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20147402 ↗
  • MalfunctionMay 2, 2018Break; Mechanics Altered; Break; Mechanics Altered

    ACCORDING TO THE REPORTER, DURING TESTING, THE DEVICE WOULD NOT WORK AND HAND PIECE WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 7479941 ↗
  • MalfunctionJan 10, 2018Break; Break

    ACCORDING TO THE REPORTER, PRE-OPERATIVELY, THE PART OF THE UNIT THAT CONNECTS THE HAND PIECE WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 7180171 ↗

Other HYSTEROSCOPE (AND ACCESSORIES) devices

All HYSTEROSCOPE (AND ACCESSORIES) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page