INVOS PMPAMP71 (UDI 10884521780101)Recalls & reported problems
Covidien LP · Preamplifier. For use with PM7100 · Class 2
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10884521780101), received between Nov 9, 2020 and Nov 6, 2024 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Low Readings” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K182868, K211561); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Low Readings2
- Failure to Analyze Signal1
- High Readings1
- Incorrect Measurement1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Insufficient Information1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 6, 2024Low Readings; Low Readings
ACCORDING TO THE REPORTER, WHILE CARDIOPULMONARY RESUSCITATION WAS ONGOING, THE MONITOR BOOTED AND, MEANWHILE, APPLIED THE SENSOR TO THE FOREHEAD OF THE PATIENT. AFTER BEING ABLE TO SEE THE READINGS ON THE MONITOR, THE HCP'S WONDERED WHY THE READING WAS SO LOW (15% RSO2). IMMEDIATELY USED ANOTHER MONITOR, WHICH SHOWED A READING OF 48%, WHEREAS THE OTHER MONITOR STILL SHOWED 15%. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20621569 ↗MalfunctionJan 31, 2024Low Readings; Low Readings
ACCORDING TO THE REPORTER, INTRA-OPERATIVELY, ONE OF THE TWO VALUES GIVEN BY THE DEVICE WAS TOO LOW WITH 36% WHILE THE TARGET WAS 50%. THE ISSUE WAS ISOLATED TO THE RSC CABLE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18619707 ↗MalfunctionJan 17, 2024Incorrect Measurement; Incorrect Measurement
ACCORDING TO THE REPORTER, THE DEVICE HAD INCORRECT READINGS. THERE WAS NO REPORTED PATIENT INVOLVEMENT OR OUTCOME.
Patient problems: Insufficient Information
MAUDE report 18531681 ↗MalfunctionJul 28, 2021High Readings; High Readings
ACCORDING TO THE REPORTER, THE DEVICE HAD INACCURATE READINGS. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12236473 ↗MalfunctionNov 9, 2020Failure to Analyze Signal; Failure to Analyze Signal
ACCORDING TO THE REPORTER, THE DEVICE WAS INTERMITTENTLY STOPPED DETECTING RSO2 VALUES. THERE WAS NO ANY ALLEGATION OF PATIENT DEATH OR SERIOUS INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10814189 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K182868, K211561. One submission often covers many models, so check the product and lot details in each record.
- Z-1732-2024 · Jan 9, 2024 · Open, ClassifiedMedtronic Invos, Reusable Sensor Cable for PM7100, REF: PMAC71RSC
- Z-1731-2024 · Jan 9, 2024 · Open, ClassifiedMedtronic Invos, Reusable Infant Sensor Adapter Cable for PM7100, REF: PMAC71RIC
Other Oximeter, tissue saturation devices
- INVOS PMAC71RSCCovidien LP · Oximeter, tissue saturation1 recall67 reports
- HEMOSPHERE FORESIGHT MODULE HEMFSM10Edwards Lifesciences LLC · Oximeter, Tissue Saturation1 recall12 reports
- INVOS PMAC71RICCovidien LP · Oximeter, tissue saturation1 recall4 reports
- FORE-SIGHT ELITE® 01-06-3000Cas Medical Systems, Inc. · Oximeter, Tissue Saturation1 recall1 death2 reports
- Invos 5100C-EUCovidien LP · Oximeter, tissue saturation22 reports
- INVOS PM7100Covidien LP · Oximeter, tissue saturation21 reports
- INVOS PM7100Covidien LP · Oximeter, tissue saturation18 reports
- INVOS 5100C-PACovidien LP · Oximeter, tissue saturation15 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.