INVOS CNN/SNN (UDI 10884521810372)Recalls & reported problems

Covidien LP · OxyAlert NIRSensor Cerebral/Somatic - Neonatal · Class 2

Adverse event reports
11
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10884521810372), received between Jun 7, 2023 and Feb 6, 2025 — classified as 2 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem10
  • Insufficient Device Problem Information2
  • Erratic or Intermittent Display1
  • High Readings1
  • Low Readings1

Patient problems

  • Pressure Sore/Ulcer5
  • Burn(s)2
  • Insufficient Information2
  • Localized Skin Lesion1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 6, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT A WOUND WAS FORMED ON A CHILD'S FOREHEAD ASSOCIATED WITH THE USE OF THE SENSOR. IT WAS IN USED BETWEEN 12 TO 24 HOURS. THE PRODUCT WAS REPLACED AND WOUND WAS TREATED.

    Patient problems: Pressure Sore/Ulcer

    MAUDE report 21325535 ↗
  • MalfunctionFeb 6, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT A WOUND WAS FORMED ON A CHILD'S FOREHEAD ASSOCIATED WITH THE USE OF THE SENSOR. IT WAS IN USED BETWEEN 12 TO 24 HOURS. THE PRODUCT WAS REPLACED AND WOUND WAS TREATED.

    Patient problems: Pressure Sore/Ulcer

    MAUDE report 21324799 ↗
  • MalfunctionJan 23, 2025Insufficient Device Problem Information; Patient Device Interaction Problem; Insufficient Device Problem Information; Patient Device Interaction Problem

    IT WAS REPORTED THAT A WOUND WAS FORMED ON A CHILD'S FOREHEAD ASSOCIATED WITH THE USE OF THE SENSOR. IT WAS NOTED THAT THREE SENSORS FROM THE BOX WERE USED, ALL OF WHICH CAUSED SIMILAR WOUNDS. THE PRODUCT WAS REPLACED.

    Patient problems: Pressure Sore/Ulcer

    MAUDE report 21223523 ↗
  • MalfunctionJan 23, 2025Insufficient Device Problem Information; Patient Device Interaction Problem; Insufficient Device Problem Information; Patient Device Interaction Problem

    IT WAS REPORTED THAT A WOUND WAS FORMED ON A CHILD'S FOREHEAD ASSOCIATED WITH THE USE OF THE SENSOR. IT WAS NOTED THAT THREE SENSORS FROM THE BOX WERE USED, ALL OF WHICH CAUSED SIMILAR WOUNDS. THE PRODUCT WAS REPLACED.

    Patient problems: Pressure Sore/Ulcer

    MAUDE report 21223511 ↗
  • MalfunctionJan 23, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT A WOUND WAS FORMED ON A CHILD'S FOREHEAD ASSOCIATED WITH THE USE OF THE SENSOR. IT WAS IN USED BETWEEN 12 TO 24 HOURS. THE PRODUCT WAS REPLACED AND WOUND WAS TREATED.

    Patient problems: Pressure Sore/Ulcer

    MAUDE report 21223484 ↗
  • MalfunctionNov 15, 2024Erratic or Intermittent Display; High Readings; Low Readings; Erratic or Intermittent Display; High Readings; Low Readings

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE UNIT HAD AN ERRATIC SPO2 (PERIPHERAL OXYGEN SATURATION) READING, RSO2 (REGIONAL OXYGEN SATURATION) VALUE WERE LOW 0 AND HIGH 60,70 AND 80. 2 CHANNELS WERE IN USED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20694936 ↗
  • MalfunctionOct 31, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    ACCORDING TO THE REPORTER, DURING USE, THE PATIENT HAD A BURN-LIKE LESION ON THE FOREHEAD DUE TO SENSOR.

    Patient problems: Localized Skin Lesion

    MAUDE report 20584455 ↗
  • InjuryJul 9, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    ACCORDING TO THE REPORTER, DURING USE, SKIN BURN WAS FOUND IN THE AREAS WHERE THE SENSORS WERE USED, THE LEFT FRONTAL REGION AND THE RIGHT ILIAC. THE SENSOR REMAINED AT THE SITE OF THE SKIN INTEGRITY PROBLEM FOR 10 DAYS, AND IT WAS VISIBLE IMMEDIATELY AFTER THE SENSOR REMOVAL. THE PATIENT PASSED AWAY, BUT IT WAS NOT RELATED TO THE SENSOR.

    Patient problems: Insufficient Information

    MAUDE report 19706366 ↗
  • InjuryJul 9, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    ACCORDING TO THE REPORTER, DURING USE, SKIN BURNS WERE FOUND IN THE AREAS WHERE THE SENSORS WERE USED, SPECIFICALLY THE LEFT FRONTAL REGION AND THE RIGHT ILIAC. THE SENSOR REMAINED AT THE SITE OF THE SKIN INTEGRITY PROBLEM FOR AT MOST EIGHT DAYS, AND THE BURN WAS VISIBLE IMMEDIATELY AFTER THE SENSOR WAS REMOVED. THE PATIENT PASSED AWAY, BUT IT WAS NOT RELATED TO THE SENSOR.

    Patient problems: Insufficient Information

    MAUDE report 19704772 ↗
  • MalfunctionJun 29, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE PATIENT HAS BEEN CONNECTED AND MONITORED WITH THE DEVICE. ONE SENSOR ON THE FOREHEAD AND ONE ON THE LOWER BACK TO MONITOR THE KIDNEYS ONCE ARRIVED TO THE NICU. THE SKIN WAS CLEANED WITH DISINFECTANT SOLUTION BEFORE APPLICATION. THE PATIENT HAD SHOWN A REACTION TO THE SENSORS. NOT ESTABLISHED YET WHETHER THIS WAS PRESSURE SORES OR KIND OF BURNS, BUT IT WAS A RED/PURPLE DISCOLORATION OF THE SKIN, THE DEMARCATION WAS EXACTLY LIKE THE SHAPE OF THE SENSOR. THE SENSOR SITE WAS CHECKED TO ASSURE THAT IT'S STILL IN THE RIGHT POSITION, THE SKIN AROUND THE SENSOR WAS NORMAL, BUT THE PROBLEM WAS UNDERNEATH THE SENSOR. IT WAS PLACED FOR 8 DAYS BEFORE THE STAFF NOTICED THE DISCOLORATION. THE PATIENT WAS SUPINE MOST OF THE TIME, BUT PATIENT WAS POSITIONED ON HIS SIDE FOR A SHORT PERIOD OF TIME. THE TREATMENT WAS FREQUENT REPOSITION OF THE PATIENT (MORE PRONE TIME) UNTIL THE DISCOLORATION IMPROVED. HOWEVER, THE PATIENT HAD IMPROVED DRAMATICALLY. THE PATIENT HAD NOT BEEN RECONNECTED WITH SENSORS AND AS MENTIONED THE FOREHEAD AND LOWER BACK HAVE IMPROVED.

    Patient problems: Burn(s)

    MAUDE report 17228004 ↗
  • MalfunctionJun 7, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE PATIENT HAS BEEN CONNECTED AND MONITORED WITH THE DEVICE. ONE SENSOR ON THE FOREHEAD AND ONE ON THE LOWER BACK TO MONITOR THE KIDNEYS ONCE ARRIVED TO THE NICU. THE SKIN WAS CLEANED WITH DISINFECTANT SOLUTION BEFORE APPLICATION. THE PATIENT HAD SHOWN A REACTION TO THE SENSORS. NOT ESTABLISHED YET WHETHER THIS WAS PRESSURE SORES OR KIND OF BURNS, BUT IT WAS A RED/PURPLE DISCOLORATION OF THE SKIN, THE DEMARCATION WAS EXACTLY LIKE THE SHAPE OF THE SENSOR. THE SENSOR SITE WAS CHECKED TO ASSURE THAT IT'S STILL IN THE RIGHT POSITION, THE SKIN AROUND THE SENSOR WAS NORMAL, BUT THE PROBLEM WAS UNDERNEATH THE SENSOR. IT WAS PLACED FOR 8 DAYS BEFORE THE STAFF NOTICED THE DISCOLORATION. THE PATIENT WAS SUPINE MOST OF THE TIME, BUT PATIENT WAS POSITIONED ON HIS SIDE FOR A SHORT PERIOD OF TIME. THE TREATMENT WAS FREQUENT REPOSITION OF THE PATIENT (MORE PRONE TIME) UNTIL THE DISCOLORATION IMPROVED. HOWEVER, THE PATIENT HAD IMPROVED DRAMATICALLY. THE PATIENT HAD NOT BEEN RECONNECTED WITH SENSORS AND AS MENTIONED THE FOREHEAD AND LOWER BACK HAVE IMPROVED.

    Patient problems: Burn(s)

    MAUDE report 17080303 ↗

Other Oximeter, tissue saturation devices