Skip to content

Shiley 125039 (UDI 10884521831575)Recalls & reported problems

Covidien LP · Endobronchial Tube Left;Endobronchial Suction Catheter;Endobronchial Tube Accessories · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884521831575), received between Apr 6, 2023 and Jul 7, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Use of Device Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Use of Device Problem2
  • Gas/Air Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Low Oxygen Saturation1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 7, 2023Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, THE ATTEMPT TO ISOLATE THE LUNG USING THE DOUBLE LUMEN TUBE WAS UNSUCCESSFUL AND THE OXYGEN SATURATION DROPPED BRIEFLY TO 94% WITH TWO LUNGS. IT WAS REALIZED THAT THE CUFF OF THE ENDOBRONCHIAL TUBE HAD AN ISSUE AND WAS NOT HOLDING PRESSURE. THE DOUBLE LUMEN TUBE WAS CHANGED TO SMALLER SIZE AND NO DESATURATION WAS NOTED. THE CUFF WAS EXAMINED AND WAS FOUND LEAKING AT THE GLUE SEAL. A SMALL HOLE IN THE CUFF WAS ALSO NOTED.

    Patient problems: Low Oxygen Saturation

    MAUDE report 17280745 ↗
  • MalfunctionApr 6, 2023Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE DEVICES CUFF HAD INFLATION OR DEFLATION ISSUE PRIOR TO USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16691376 ↗
  • MalfunctionApr 6, 2023Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE DEVICE'S CUFF HAD INFLATION OR DEFLATION ISSUE PRIOR TO USE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16691347 ↗

Other TUBE, BRONCHIAL (W/WO CONNECTOR) devices

All TUBE, BRONCHIAL (W/WO CONNECTOR) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page