Shiley 125039 (UDI 40884522008755)Recalls & reported problems

Covidien LP · Endobronchial Suction Catheters with Color Coded Connectors; Endobronchial Tube, Left; Endobronchial Tube Accessories; (2) Right Angle Swivel Connectors; Carlens Y Adapter · Class 2

Quick summary

This Shiley double-lumen bronchial tube kit is used for single-lung ventilation and suction. Reports mostly describe cuff or balloon air leaks and inflation/deflation problems, along with some mechanical or component issues and, in a few cases, a foreign object or patient injury. There are 41 adverse event reports, no deaths, and 7 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
40
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 40 adverse event reports that list this exact device (DI 40884522008755), received between Aug 8, 2017 and Apr 29, 2026 — classified as 7 injuries, 33 malfunctions.

The most frequently reported specific device problem is “Use of Device Problem” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 3 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 4 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem10
  • Gas/Air Leak7
  • Defective Component5
  • Contamination /Decontamination Problem3
  • Deflation Problem3
  • Mechanical Problem3
  • Partial Blockage3
  • Human-Device Interface Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions13
  • No Patient Involvement9
  • Unintended Extubation9
  • Extubate4
  • Foreign Body In Patient3
  • Injury3
  • No Consequences Or Impact To Patient2
  • Adult Respiratory Distress Syndrome1

Recent adverse event reports

The 25 most recent of 40 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2026Use of Device Problem

    IT WAS REPORTED THAT THE BRONCO CATHETER HAD A CUFF INFLATION/DEFLATION ISSUE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25029506 ↗
  • MalfunctionApr 28, 2026Use of Device Problem

    IT WAS REPORTED THAT THE BRONCO CATHETER HAD A CUFF INFLATION/DEFLATION ISSUE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25023485 ↗
  • MalfunctionFeb 9, 2026Use of Device Problem

    IT WAS REPORTED THAT DURING PRE-TEST, THE DOUBLE LUMEN ENDOTRACHEAL TUBE'S TRACHEAL CUFF HAD AN AIR LEAKAGE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24292560 ↗
  • MalfunctionJan 27, 2026Contamination /Decontamination Problem

    IT WAS REPORTED THAT DURING USE OF A DOUBLE-LUMEN TUBE, A FOREIGN OBJECT (BLUE FIBER) WAS DETECTED IN THE BRONCHIAL LUMEN AFTER INTUBATION AND DURING FIBEROPTIC POSITION CONTROL. AN ATTEMPT WAS MADE TO RETRIEVE THE FIBER USING SUCTION, BUT IT COULD NOT BE DETECTED IN THE SUCTION SYSTEM, THE TUBE, OR THE PATIENT AFTER THE RETRIEVAL ATTEMPT. A FIBEROPTIC INSPECTION OF THE PLASTIC COMPONENTS AND TUBES SHOWED NO EVIDENCE OF THE FOREIGN OBJECT, AND A VISUAL INSPECTION REVEALED THAT BOTH CUFFS APPEARED TO BE INTACT AND LEAK-FREE. THE DOUBLE-LUMEN TUBE WAS REPLACED BEFORE THE SURGICAL PROCEDURE BEGAN THROUGH A TROUBLE-FREE REINTUBATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unintended Extubation

    MAUDE report 24193133 ↗
  • MalfunctionJan 8, 2026Gas/Air Leak

    IT WAS REPORTED THAT THE THREE BRONCO CATHETERS HAD SLOW LEAKS DETECTED IN THE TRACHEAL CUFFS DURING PATIENT INTUBATION ONCE THE TUBE WAS IN SITU, IN A HOSPITAL SETTING. THE TUBES WERE REPLACED EVERY TIME IT FAILED AND LASTLY, A 41L TUBE WAS USED, AND SUCCESSFUL INTUBATION WAS ACHIEVED WITH NO FURTHER CUFF ISSUES.

    Patient problems: Unintended Extubation

    MAUDE report 24012520 ↗
  • MalfunctionJan 8, 2026Gas/Air Leak

    IT WAS REPORTED THAT THE THREE BRONCO CATHETERS HAD SLOW LEAKS DETECTED IN THE TRACHEAL CUFFS DURING PATIENT INTUBATION ONCE THE TUBE WAS IN SITU, IN A HOSPITAL SETTING. THE TUBES WERE REPLACED EVERY TIME IT FAILED AND LASTLY, A 41L TUBE WAS USED, AND SUCCESSFUL INTUBATION WAS ACHIEVED WITH NO FURTHER CUFF ISSUES.

    Patient problems: Unintended Extubation

    MAUDE report 24012515 ↗
  • MalfunctionJan 8, 2026Gas/Air Leak

    IT WAS REPORTED THAT THE THREE BRONCO CATHETERS HAD SLOW LEAKS DETECTED IN THE TRACHEAL CUFFS DURING PATIENT INTUBATION ONCE THE TUBE WAS IN SITU, IN A HOSPITAL SETTING. THE TUBES WERE REPLACED EVERY TIME IT FAILED AND LASTLY, A 41L TUBE WAS USED, AND SUCCESSFUL INTUBATION WAS ACHIEVED WITH NO FURTHER CUFF ISSUES.

    Patient problems: Unintended Extubation

    MAUDE report 24012487 ↗
  • MalfunctionNov 3, 2025Defective Component; Defective Component

    IT WAS REPORTED THAT PRIOR TO USAGE, THE DOUBLE LUMEN TUBE'S TIP HAD COLLOID DEFORMATION. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23450570 ↗
  • MalfunctionSep 18, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT DURING TESTING OF THE TWO BRONCO CATHETERS, A CUFF LEAK WAS IDENTIFIED. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23089918 ↗
  • MalfunctionSep 18, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT DURING TESTING OF THE TWO BRONCO CATHETERS, A CUFF LEAK WAS IDENTIFIED. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23089900 ↗
  • MalfunctionSep 12, 2025Mechanical Problem; Material Split, Cut or Torn; Mechanical Problem; Material Split, Cut or Torn

    IT WAS REPORTED THAT PRIOR TO USAGE, IT WAS FOUND THAT THE DOUBLE LUMEN TUBE HAD A DEFECTIVE CAP AND DIFFICULT TO SEAL, AS SHOWN IN THE ATTACHED IMAGE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23038357 ↗
  • MalfunctionMar 29, 2025Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT THE 39FR BRONCO CATHETER'S CUFF WAS TESTED PRIOR TO USE AT THE HOSPITAL, BUT NO PRESSURE INCREASE WAS OBSERVED, INDICATING A MALFUNCTION. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21727297 ↗
  • MalfunctionJan 15, 2025Defective Component; Defective Component

    IT WAS REPORTED THAT THE BRONCO CATHETER'S TUBE BODY WAS BENT. THE ISSUE OCCURRED IN A HOSPITAL SETTING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21164647 ↗
  • MalfunctionDec 10, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem

    ACCORDING TO THE REPORTER, INTRA-OPERATIVELY, THERE WAS A BLUE PLASTIC SEEN IN THE BRONCHIAL LUMEN FOLLOWING INTUBATION. IT HAD TO BE REMOVED AND REPLACED CAUSING ADDITIONAL AIRWAY INSTRUMENTATION.

    Patient problems: Unintended Extubation

    MAUDE report 20888928 ↗
  • InjuryDec 2, 2024Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT DURING A PROCEDURE INVOLVING A BRONCO CATHETER, A PATIENT REQUIRED REMOVAL OF A CORRECTLY SITED DOUBLE LUMEN TUBE AND REINTUBATION DUE TO INCREASED INSTRUMENTATION, LIP LACERATION AND SORE THROAT AT THE AIRWAY. A PIECE OF BLUE MATERIAL WAS EVIDENT IN THE BRONCHIAL LUMEN OF THE DOUBLE LUMEN TUBE FOLLOWING A DIFFICULT INTUBATION, WHICH RESULTED IN NO FURTHER VENTILATION THROUGH THE LUMEN. THERE WAS A CONCERN THAT AN UNRECOGNIZED FOREIGN BODY COULD BE INTRODUCED INTO THE AIRWAY OR LUNG WITH VENTILATION. THE TUBE WAS SUBSEQUENTLY REMOVED AND REPLACED.

    Patient problems: Adult Respiratory Distress Syndrome; Foreign Body In Patient; Laceration(s); Unintended Extubation; Unspecified Tissue Injury

    MAUDE report 20818226 ↗
  • MalfunctionJun 21, 2024Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE CUFFS DID NOT EXPAND AND DEFLATE NORMALLY DURING THE PRE-USAGE INSPECTION. WHEN AIR WAS INJECTED INTO THE WHITE CUFF, BOTH THE WHITE AND BLUE CUFFS INFLATED. THERE WERE NO CHANGES EVEN AFTER AIR INJECTION INTO THE BLUE CUFF. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19586693 ↗
  • MalfunctionJul 7, 2022Gas/Air Leak; Manufacturing, Packaging or Shipping Problem; Gas/Air Leak; Manufacturing, Packaging or Shipping Problem

    ACCORDING TO THE REPORTER, DURING THE OPENING TEST, IT WAS FOUND THAT THE CUFF LEAKED AIR. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14946010 ↗
  • MalfunctionMar 18, 2022Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE CUFF OF THE DEVICE HAD A LEAK WHEN INFLATED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13815444 ↗
  • MalfunctionMar 8, 2022Defective Component; Defective Component

    ACCORDING TO THE REPORTER, DURING PLACING THE TUBE ON THE PATIENT'S LEFT SIDED BRONCHUS, THE PEDIATRIC BRONCHOSCOPE COULD EASILY PASS VIA BRONCHIAL LUMEN BUT WHEN THE DEVICE WAS INSERTED, IT COULD NOT ADVANCE IN THE TRACHEAL LUMEN AND MORE EFFORT NEEDED TO APPLIED. THE DEVICE WAS REPLACED WITH A SINGLE LUMEN TUBE THEN A FRENCH 41 DOUBLE LUMEN TUBE VIA BOUGIE IN ORDER TO HAVE A DOUBLE LUMEN TUBE THAT WOULD ALLOW FOR EASE OF PASSAGE OF PEDIATRIC BRONCHOSCOPE. THE PATIENT HAS PARTIAL VIEW OF GLOTTIS. THE FAILURE RESULTED TO DELAY OF TREATMENT. THE PHYSICIAN TRIED THE SCOPE TO ANOTHER NEW FR 39 TUBE TO TEST IF IT ALSO HAD SAME ISSUE AND FOUND THAT THERE WAS ALSO RESISTANCE IN TRACHEAL LUMEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13700911 ↗
  • MalfunctionMar 8, 2022Defective Component; Defective Component

    ACCORDING TO THE REPORTER, DURING PLACING THE TUBE ON THE PATIENT'S LEFT SIDED BRONCHUS, THE PEDIATRIC BRONCHOSCOPE COULD EASILY PASS VIA BRONCHIAL LUMEN BUT WHEN THE DEVICE WAS INSERTED, IT COULD NOT ADVANCE IN THE TRACHEAL LUMEN AND MORE EFFORT NEEDED TO APPLIED. THE DEVICE WAS REPLACED WITH A SINGLE LUMEN TUBE THEN A FRENCH 41 DOUBLE LUMEN TUBE VIA BOUGIE IN ORDER TO HAVE A DOUBLE LUMEN TUBE THAT WOULD ALLOW FOR EASE OF PASSAGE OF PEDIATRIC BRONCHOSCOPE. THE PATIENT HAS PARTIAL VIEW OF GLOTTIS. THE FAILURE RESULTED TO DELAY OF TREATMENT. THE PHYSICIAN TRIED THE SCOPE TO ANOTHER NEW FR 39 TUBE TO TEST IF IT ALSO HAD SAME ISSUE AND FOUND THAT THERE WAS ALSO RESISTANCE IN TRACHEAL LUMEN.

    Patient problems: Unintended Extubation

    MAUDE report 13700679 ↗
  • MalfunctionJun 11, 2021Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, THE DOUBLE-LUMEN'S WHITE BALLOON WAS LEAKING. RE-INTUBATION WAS REQUIRED.

    Patient problems: Unintended Extubation

    MAUDE report 11986088 ↗
  • InjuryJul 2, 2020Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, DURING USE, THEY NOTICED A VISIBLE HOLE IN THE WALL OF THE 39FR TUBE THAT RESULTED IN A MASSIVE AIR LEAK. RESULTED IN A DELAY OF THE PROCEDURE WAS INDICATED. RE-INTUBATION WAS REQUIRED.

    Patient problems: Extubate; Unintended Extubation

    MAUDE report 10227544 ↗
  • MalfunctionJun 18, 2020Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, THE CUFF WAS INFLATABLE, BUT NOT FULL BECAUSE OF LEAKAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 10167980 ↗
  • MalfunctionJun 2, 2020Defective Component; Defective Component

    ACCORDING TO THE REPORTER, THE DEVICE¿S SIZE WAS FOUND NOT GOOD WHEREIN SIZE CH 39 WAS USED INSTEAD OF CH 41. IT WAS ALSO STATED THAT AN ERROR MESSAGE APPEARED IN THE RESPIRATOR STATING ¿VACUUM BELLOWS SYSTEM LEAK, CANNOT OPERATE BELLOWS. THERE WAS NO REPORTED PATIENT INVOLVEMENT OR OUTCOME.

    Patient problems: No Information

    MAUDE report 10108425 ↗
  • InjuryFeb 25, 2020Mechanical Problem; Mechanical Problem

    ACCORDING TO THE REPORTER, THE DEVICE'S TUBE WAS UTILIZED FOR THE PATIENT AND THE GLUE ON THE TUBE STARTED COMING LOOSE IN LESS THAN A WEEK. THE PATIENT REQUIRED INTERVENTION.