Mallinckrodt 125037Recalls & reported problems

Covidien LP · Endobronchial Suction Catheters with Color Coded Connectors; Endobronchial Tube; Endobronchial Tube Accessories,Left,(2) Right Angle Swivel Connectors; Carlens Y Adapter · Class 2

All models of Mallinckrodt (73)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is a Covidien bronchial tube used to keep a patient's airway open during lung surgery and ventilation. The 49 reports mostly describe the tube's inflatable cuff not deflating or leaking air, so air stayed trapped after the tube was removed; a few mention partial blockage and one fiberscope getting stuck inside the tube during surgery, briefly interrupting the operation. Two reports describe injuries and no deaths. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
49
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 49 adverse event reports that list this exact device (DI 30884522008741), received between Mar 17, 2017 and Oct 4, 2024 — classified as 2 injuries, 47 malfunctions.

The most frequently reported specific device problem is “Deflation Problem” (33 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deflation Problem33
  • Inflation Problem7
  • Use of Device Problem4
  • Partial Blockage3
  • Gas/Air Leak2
  • Appropriate Device Problem Term/Code Not Available1
  • Material Integrity Problem1

Patient problems

  • No Patient Involvement41
  • No Clinical Signs, Symptoms or Conditions3
  • No Known Impact Or Consequence To Patient3
  • Unintended Extubation2
  • Low Oxygen Saturation1

Recent adverse event reports

The 25 most recent of 49 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 4, 2024Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, THE BRONCO CATHETER'S TRACHEAL CUFF HAD AN AIR LEAK DURING THE PROCEDURE. THE TUBE WAS REPLACED.

    Patient problems: Unintended Extubation

    MAUDE report 20376817 ↗
  • InjuryDec 14, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    ACCORDING TO THE REPORTER, DURING A PROCEDURE TO REMOVE THE PATIENT¿S LEFT LUNG, THE FIBERSCOPE STUCK IN THE LEFT BRONCO CATHETER LUMEN. THE PATIENT WAS EXTUBATED AND REINTUBATED WITH A NORMAL LOW-PRESSURE 7F PROBE, WHICH RESULTED IN A MOMENTARY SUSPENSION OF SURGERY. THE PATIENT DESATURATED DESPITE ECMO (EXTRACORPOREAL MEMBRANE OXYGENATION) IN PLACE.

    Patient problems: Low Oxygen Saturation; Unintended Extubation

    MAUDE report 18329001 ↗
  • MalfunctionFeb 22, 2022Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE DEVICE AIR REMAINED AFTER REMOVAL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13582541 ↗
  • MalfunctionFeb 22, 2022Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE DEVICE AIR REMAINED AFTER REMOVAL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13582451 ↗
  • MalfunctionFeb 22, 2022Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, THE DEVICE WOULD NOT INFLATE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13582401 ↗
  • MalfunctionJul 13, 2020Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, THE DEVICE HAD A CUFF LEAKAGE WHEN INFLATED. THERE WAS NO REPORTED PATIENT OUTCOME.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10265545 ↗
  • MalfunctionOct 14, 2019Use of Device Problem; Use of Device Problem

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE BALLOON DID INFLATE BUT WOULD NOT DEFLATE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 9186241 ↗
  • MalfunctionApr 3, 2019Partial Blockage; Partial Blockage

    ACCORDING TO THE REPORTER, WHEN AIR WAS ATTEMPTED TO BE REMOVED AFTER CUFF CHECK WAS PERFORMED, DEGASSING WAS UNABLE TO BE PERFORMED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8478866 ↗
  • MalfunctionDec 27, 2018Inflation Problem; Inflation Problem

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE DEVICE HAD AIR LEAK FROM TRACHEAL CUFF UPON CUFF CHECK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 8200068 ↗
  • MalfunctionJul 6, 2018Partial Blockage; Partial Blockage

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE AIR FROM THE DEVICE WAS NOT DEFLATED COMPLETELY DURING CUFF TEST. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 7666677 ↗
  • InjuryMay 21, 2018Inflation Problem; Inflation Problem

    ACCORDING TO THE REPORTER, THE DEVICE'S CUFF COULD NOT BE INFLATED. THE CUSTOMER REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 7530289 ↗
  • MalfunctionApr 4, 2018Partial Blockage; Partial Blockage

    ACCORDING TO THE REPORTER, DURING CUFF TEST, IT WAS FOUND THAT THE TRACHEAL CUFF HAD A LEAK. IT WAS INDICATED THERE WAS NO PATIENT I INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 7397919 ↗
  • MalfunctionMar 2, 2018Deflation Problem; Deflation Problem

    ACCORDING TO THE REPORTER, DURING TESTING, THE INFLATED TRACHEAL CUFF DID NOT DEFLATE. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 7309423 ↗
  • MalfunctionMar 2, 2018Deflation Problem; Deflation Problem

    ACCORDING TO THE REPORTER, DURING TESTING, THE INFLATED TRACHEAL CUFF DID NOT DEFLATE. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 7309414 ↗
  • MalfunctionNov 30, 2017Deflation Problem; Deflation Problem

    MEDTRONIC RECEIVED A COMMUNICATION REGARDING A CUFF THAT WOULDN'T DEFLATE. ACCORDING TO THE REPORTER, DURING THE TEST, TUBE DID NOT DEFLATE AFTER BEING INFLATED. THE CUSTOMER REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 7071757 ↗
  • MalfunctionNov 20, 2017Deflation Problem; Deflation Problem

    MEDTRONIC RECEIVED A COMMUNICATION REGARDING A CUSTOMER THAT COULD TAKE OUT THE AIR FROM THE BALLOON. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 7046982 ↗
  • MalfunctionNov 20, 2017Deflation Problem; Deflation Problem

    MEDTRONIC RECEIVED A COMMUNICATION REGARDING A CUSTOMER THAT COULD TAKE OUT THE AIR FROM THE BALLOON. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 7046971 ↗
  • MalfunctionNov 16, 2017Deflation Problem; Deflation Problem

    MEDTRONIC RECEIVED A COMMUNICATION REGARDING A CUFF THAT COULD NOT BE DEFLATED. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT AS THE ISSUE WAS FOUND WHILE CHECKING PRIOR TO USE.

    Patient problems: No Patient Involvement

    MAUDE report 7039334 ↗
  • MalfunctionNov 14, 2017Deflation Problem; Deflation Problem

    MEDTRONIC RECEIVED A COMMUNICATION REGARDING DIFFICULTY DEFLATING THE CUFF. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 7029624 ↗
  • MalfunctionNov 14, 2017Deflation Problem; Deflation Problem

    MEDTRONIC RECEIVED A COMMUNICATION REGARDING DIFFICULTY DEFLATING THE CUFF. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 7029618 ↗
  • MalfunctionNov 14, 2017Deflation Problem; Deflation Problem

    MEDTRONIC RECEIVED A COMMUNICATION REGARDING DIFFICULTY DEFLATING THE CUFF. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 7029601 ↗
  • MalfunctionNov 7, 2017Inflation Problem; Inflation Problem

    MEDTRONIC RECEIVED A REPORT THAT PRIOR TO USE ON A PATIENT, THE CUFF DID NOT INFLATE SYMMETRICALLY. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 7006450 ↗
  • MalfunctionNov 6, 2017Deflation Problem; Deflation Problem

    MEDTRONIC RECEIVED A REPORT THAT PRIOR TO USE ON A PATIENT, THE AIRWAY CUFF DID NOT DEFLATE. IT WAS REPORTED THAT PART OF THE POLYURETHANE CUFF WAS COVERING THE DEFLATION HOLE OF THE CUFF. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 7003532 ↗
  • MalfunctionOct 4, 2017Deflation Problem; Deflation Problem

    MEDTRONIC RECEIVED A REPORT THAT DURING THE PRE-TEST THE CUFF WOULD NOT DEFLATE. THE CUSTOMER REPORTED THAT THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Patient Involvement

    MAUDE report 6914505 ↗
  • MalfunctionOct 4, 2017Deflation Problem; Deflation Problem

    MEDTRONIC RECEIVED A REPORT THAT DURING THE PRE-TEST THE TUBE'S CUFF WOULD NOT DEFLATE. THE CUSTOMER REPORTED THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Patient Involvement

    MAUDE report 6914409 ↗

Other TUBE, BRONCHIAL (W/WO CONNECTOR) devices

All TUBE, BRONCHIAL (W/WO CONNECTOR) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page