SILS SILSCLINCH36Recalls & reported problems
Covidien LP · Clinch · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10884523000580), received between Sep 15, 2016 and Jun 12, 2017 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break1
- Grounding Malfunction1
Patient problems
- Burn(s)1
- No Patient Involvement1
- Tissue Damage1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 12, 2017Break; Break
ACCORDING TO REPORTER DURING A LAP HYSTERECTOMY PROCEDURE THE HANDPIECE WAS VISIBLY BROKEN WHEN OPENED FROM PACKAGING AND UNABLE TO BE USED. NO ADDITIONAL OPERATING TIME. NO ADDITIONAL BLOOD LOSS. NO ADVERSE EFFECT TO PATIENT. NO TISSUE DAMAGE OR UNANTICIPATED TISSUE LOSS, OPENED UP ANOTHER HAND PIECE AND THERE WAS NO ISSUE.
Patient problems: No Patient Involvement
MAUDE report 6633547 ↗MalfunctionSep 15, 2016Grounding Malfunction; Grounding Malfunction
ACCORDING TO THE REPORTER, DURING THE SILS APPENDECTOMY, THE DEVICE GROUNDED ON THE PRE-UMBILICAL SKIN AND CAUSED A SMALL BURN AT THE SITE WHERE THEY ALL TOUCH. THE SILS DEVICE INSULATION WAS NOT DURABLE ENOUGH TO SUPPORT OTHER INSTRUMENTS RUBBING AGAINST IT AT THE SINGLE INSERTION ACCESS POINT. THERE WAS NO MEDICAL INTERVENTION DONE TO THE PATIENT. A WOUND DRESSING (GAUZE) WAS PLACED ON THE PATIENT BUT THAT WAS PER NORMAL PROCEDURE. THERE WAS MINOR TISSUE DAMAGE DUE TO THE BURN.
Patient problems: Burn(s); Tissue Damage
MAUDE report 5952405 ↗
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page