BD SmartSite 70001B-07TRecalls & reported problems

BD Switzerland Sàrl · Low Sorbing Secondary Set Needle-Free Valve Bag Access Port, Closed Male Luer, Nonvented, Priming Cap (orange colored) Hanger 79 cm (31 in) · Class 2

All models of BD SmartSite (29)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This BD SmartSite secondary infusion set is used to deliver fluids or drugs into a vein. The 34 reports that name its identifier mostly describe liquid leaking at the connector during infusion, sometimes delaying treatment. There were no deaths and no injuries. It has 2 open recalls, one for leaking at the luer connection and one because sets labeled DEHP-free actually contain DEHP.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
34
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 10885403240997), received between May 10, 2023 and Jun 15, 2026 — classified as 34 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (26 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 10 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 4 reportsMar 2026: 4 reportsApr 2026: 5 reportsMay 2026: 2 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 12, 2025Open, Classified

    Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

    Z-1970-2025 · Root cause: Process control · Recall details · FDA record ↗

  2. Oct 11, 2023Open, Classified

    Affected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP).

    Z-0351-2024 · Root cause: Error in labeling · Recall details · FDA record ↗

Most reported problems

Device problems

  • Leak/Splash26
  • Fluid/Blood Leak5
  • Complete Blockage1
  • Component Missing1
  • Defective Component1
  • Infusion or Flow Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions32
  • Chemical Exposure1
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 15, 2026Leak/Splash

    IT WAS REPORTED THAT THE BD TEXIUM SECONDARY SET SMARTSITE NEEDLE-FREE VALVE BAG ACCESS PORT HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: LEAKING AT GREEN TEXIUM BONDED SITE DURING ADMINISTRATION. ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER: WHAT WAS THE IMPACT TO THE PATIENT? INFUSION WAS SLIGHTLY DELAYED. A PHARMACY STAFF HAD TO COME IN OVERNIGHT TO REPLACE THE ADMINISTRATION LINE IN ORDER FOR PROPER INFUSION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25510773 ↗
  • MalfunctionMay 29, 2026Leak/Splash

    IT WAS REPORTED THAT THE BD TEXIUM SECONDARY SET SMARTSITE NEEDLE-FREE VALVE BAG ACCESS PORT HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MANY REPORTS RECEIVED IN REGARDS TO BD LOW SORBING SECONDARY SET SMARTSITE NEEDLE-FREE VALVE BAG ACCESS PORT REF # 70001B-07T. 5 REPORTS TOTAL REPORTED IN TERMS OF LEAKING. ONE REPORT INVOLVED STARTING A PIGGYBACK INFUSION OF FLUOROURACIL 900MG, A LEAK OCCURRED AT THE CONNECTOR SITE TO THE PRIMARY IV LINE. APPROXIMATELY 5 ML OF FLUID LEAKED, BUT NONE CAME INTO CONTACT WITH THE PATIENT. THE LEAK WAS CONTAINED WITH A DISPOSABLE PAD. PHARMACY WAS NOTIFIED. SIMILAR EVENTS COULD OCCUR AT OTHER FACILITIES USING THE SAME BD LOW SORBING SECONDARY SET SMARTSITE NEEDLE-FREE VALVE BAG ACCESS PORT IF THERE ARE PRODUCT DEFECTS, COMPATIBILITY ISSUES, OR ASSEMBLY ERRORS. IF LEAKAGE OCCURS WITH HAZARDOUS DRUGS LIKE CHEMOTHERAPY, STAFF COULD BE EXPOSED TO CYTOTOXIC FLUIDS, PRESENTING HEALTH RISKS SUCH AS SKIN IRRITATION OR LONG-TERM EFFECTS. ADDITIONALLY, MEDICATION LOSS COULD LEAD TO UNDERDOSING, IMPACTING PATIENT CARE. DESCRIBE ANY PATIENT HARM, INJURY, COMPLICATION OR NEGATIVE OUTCOME THAT OCCURRED AS A RESULT OF THE EVENT. NEGA

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25337618 ↗
  • MalfunctionMay 29, 2026Leak/Splash

    IT WAS REPORTED THAT THE BD TEXIUM SECONDARY SET SMARTSITE NEEDLE-FREE VALVE BAG ACCESS PORT HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MANY REPORTS RECEIVED IN REGARDS TO BD LOW SORBING SECONDARY SET SMARTSITE NEEDLE-FREE VALVE BAG ACCESS PORT REF # (B)(4). 5 REPORTS TOTAL REPORTED IN TERMS OF LEAKING. ONE REPORT INVOLVED STARTING A PIGGYBACK INFUSION OF FLUOROURACIL 900MG, A LEAK OCCURRED AT THE CONNECTOR SITE TO THE PRIMARY IV LINE. APPROXIMATELY 5 ML OF FLUID LEAKED, BUT NONE CAME INTO CONTACT WITH THE PATIENT. THE LEAK WAS CONTAINED WITH A DISPOSABLE PAD. PHARMACY WAS NOTIFIED. SIMILAR EVENTS COULD OCCUR AT OTHER FACILITIES USING THE SAME BD LOW SORBING SECONDARY SET SMARTSITE NEEDLE-FREE VALVE BAG ACCESS PORT IF THERE ARE PRODUCT DEFECTS, COMPATIBILITY ISSUES, OR ASSEMBLY ERRORS. IF LEAKAGE OCCURS WITH HAZARDOUS DRUGS LIKE CHEMOTHERAPY, STAFF COULD BE EXPOSED TO CYTOTOXIC FLUIDS, PRESENTING HEALTH RISKS SUCH AS SKIN IRRITATION OR LONG-TERM EFFECTS. ADDITIONALLY, MEDICATION LOSS COULD LEAD TO UNDERDOSING, IMPACTING PATIENT CARE. DESCRIBE ANY PATIENT HARM, INJURY, COMPLICATION OR NEGATIVE OUTCOME THAT OCCURRED AS A RESULT OF THE EVENT. NEGATIVE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25334942 ↗
  • MalfunctionApr 29, 2026Leak/Splash

    IT IS REPORTED, BD LOW SORBING SECONDARY SET SMARTSITE NEEDLE-FREE VALVE BAG ACCESS PORT, LEAKING AROUND CONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25034188 ↗
  • MalfunctionApr 22, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT BD ALARIS TEXIUM INFUSION SET WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY CUSTOMER THAT "LEAKING OCCURRED MULTIPLE TIMES IN THE PAST".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24957684 ↗
  • MalfunctionApr 22, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT BD TEXIUM EXTENSION SET WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY CUSTOMER THAT "LEAKING OCCURRED MULTIPLE TIMES IN THE PAST".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24957019 ↗
  • MalfunctionApr 10, 2026Leak/Splash

    IT WAS REPORTED BY CUSTOMER THAT UPON HOOKING THE MEDICATION UP, RN NOTICED LEAKING AT THE HOOK-UP SITE. ATTEMPTED TO RECONNECT AND STILL NOTICED LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24844414 ↗
  • MalfunctionApr 6, 2026Leak/Splash

    IT WAS REPORTED THAT THE BD TEXIUM CLOSED MALE LUER HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE HAVE RECEIVED ANOTHER REPORT OF A LEAKING BD LOW SORBING SECONDARY SET SMARTSITE NEEDLE-FREE VALVE BAG ACCESS PORT REF # 70001B-07T. UNFORTUNATELY, THE LOT# WAS NOT ABLE TO BE OBTAINED. THE EVENT HAPPENED ON 3/19/2026. I DO NOT HAVE ANY PICTURES OF THE EVENT. IT WAS AGAIN NOTED ON A DOXORUBICIN INFUSION FOR CHEMOTHERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24793631 ↗
  • MalfunctionMar 24, 2026Fluid/Blood Leak

    IT WAS REPORTED THAT BD EXTENSION TEXIUM SET WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM: IT WAS REPORTED BY CUSTOMER THAT MULTIPLE LEAKAGE INCIDENTS HAVE BEEN REPORTED WITH THIS SPECIFIC INFUSION SET.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24675645 ↗
  • MalfunctionMar 24, 2026Fluid/Blood Leak

    NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24675621 ↗
  • MalfunctionMar 24, 2026Fluid/Blood Leak; Leak/Splash

    NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24675620 ↗
  • MalfunctionMar 23, 2026Leak/Splash

    IT IS REPORTED BD LOW SORBING SECONDARY SET SMARTSITE NEEDLE FREE VALVE BAG ACCESS PORT LEAKED. THE LEAK OCCURRED FROM THE BOTTOM END WHERE IT ATTACHED TO THE YSITE, BUT WAS UNABLE TO BE TIGHTENED ANY FURTHER.

    Patient problems: Insufficient Information

    MAUDE report 24665321 ↗
  • MalfunctionFeb 18, 2026Leak/Splash

    IT IS REPORTED LEAKAGE. PRIMARY OR SECONDARY TUBING WITH TEXIUM ENDS WOULD START LEAKING FROM THE CONNECTION POINT. LEAKS WOULD REGARDLESS OF HOW TIGHT THE CONNECTION WAS SECURED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24379978 ↗
  • MalfunctionFeb 18, 2026Leak/Splash

    IT IS REPORTED LEAKAGE. PRIMARY OR SECONDARY TUBING WITH TEXIUM ENDS WOULD START LEAKING FROM THE CONNECTION POINT. LEAKS WOULD REGARDLESS OF HOW TIGHT THE CONNECTION WAS SECURED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24379435 ↗
  • MalfunctionFeb 18, 2026Leak/Splash

    IT IS REPORTED LEAKAGE. PRIMARY OR SECONDARY TUBING WITH TEXIUM ENDS WOULD START LEAKING FROM THE CONNECTION POINT. LEAKS WOULD REGARDLESS OF HOW TIGHT THE CONNECTION WAS SECURED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24379434 ↗
  • MalfunctionFeb 18, 2026Leak/Splash

    IT IS REPORTED LEAKAGE. PRIMARY OR SECONDARY TUBING WITH TEXIUM ENDS WOULD START LEAKING FROM THE CONNECTION POINT. LEAKS WOULD REGARDLESS OF HOW TIGHT THE CONNECTION WAS SECURED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24379433 ↗
  • MalfunctionJan 20, 2026Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24120708 ↗
  • MalfunctionJan 20, 2026Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24120691 ↗
  • MalfunctionSep 12, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS SMARTSITE CLOSED MALE LUER WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY THE CUSTOMER THAT CHEMO DRUGS HAVE BEEN LEAKING FROM THEM CAUSING SPILLS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23038832 ↗
  • MalfunctionSep 12, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS SMARTSITE CLOSED MALE LUER WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY THE CUSTOMER THAT CHEMO DRUGS HAVE BEEN LEAKING FROM THEM CAUSING SPILLS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23038830 ↗
  • MalfunctionMay 23, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS, SMARTSITE TEXIUM LOW SORBING SECONDARY SET WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY THE CUSTOMER THAT THEY ARE CHECKING THE INTEGRITY OF THE TUBING AS INSTRUCTED BEFORE AND WHEN PRIMING. THE LEAK IS HAPPENING WELL INTO HAZARDOUS DRUG INFUSIONS.

    Patient problems: Chemical Exposure

    MAUDE report 22081712 ↗
  • MalfunctionMay 23, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE BD ALARIS TEXIUM SMARTSITE LOW SORBING SECONDARY SET LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CHEMO RN CAME TO THE PHARMACY STATING KEYTRUDA IV PIGGYBACK LEAKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22079336 ↗
  • MalfunctionMay 19, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS SMARTSITE TEXIUM WAS LEAKING THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY THE CUSTOMER THAT LEAKING SECONDARY AT THE TEXIUM CAP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22049349 ↗
  • MalfunctionMay 15, 2025Leak/Splash; Leak/Splash

    CODES UPDATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22028179 ↗
  • MalfunctionMay 15, 2025Leak/Splash; Leak/Splash

    CODES UPDATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22028175 ↗

Other Set, administration, intravascular devices