Alaris Pump Infusion Set: SmartSite Bag Access Non-Vented… recallZ-1970-2025 · Bd Switzerland Sarl · initiated May 12, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Bd Switzerland Sarl or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

FDA root cause category: Process control

What the company is doing

On 5/12/2025, removal notices were mailed or emailed to Medical Director, Risk Manager, Medical Device Safety Officer, Nurse Managers, and Distributors who were asked to do the following: 1. Immediately review your inventory for affected product. Destroy all unused affected products subject to the recall following your institution s process for destruction. 2. Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. 3. Complete the Customer Response Form via email to [email protected] 4. In addition, distributors were asked to identify all customers that purchased any affected product and to provide a copy of the customer letter to all customers and to advise them of the removal notification. Report any complaints experienced with the use of this product to the firm's North American Regional Complaint Center: Phone: 1-844-823-5433 Mon-Fri 8:00am and 5:00pm CT, Email: [email protected]

Recall details

Recall number
Z-1970-2025
Status
Open, Classified
Initiated by the company
May 12, 2025
Posted by the FDA
Jun 12, 2025
Recalling firm
Bd Switzerland Sarl
Quantity
2,024,780
Product code
FPA
FDA submissions
K053049
Distribution
Worldwide - US Nationwide distribution including in the states of MA, CA, IL, AK, AL, MO, FL, VA, ME, NC, MD, IA, WI, AZ, CO, KY, IN, TX, AR, NJ, NY, MT, NE, TN, PA, GA, WA, KS, MN, ID, SC, MI, SD, OH, WV, LA, ND, VT, GU, NV, UT, OK, OR, DE, NM, MS, WY and the countries of CA, AU, NZ, CN.

Product description

Alaris Pump Infusion Set: SmartSite Bag Access Non-Vented Bonded Texium With Priming Cap Back Check Valve 2 SmartSite Y-Sites, REF: 10013361T; SmartSite Bag Access Non-Vented Bonded Texium Closed Male Luer With Priming Cap, REF: 10321213T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium Closed Male Luer With Priming Cap, REF: 22000-B007T; Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, REF: 22600-0007T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium With Priming Cap, REF: 22602-B007T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium With Priming Cap, REF: 22602-B007T; SmartSite Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded Texium With Priming Cap, REF: 22603-B007T; Bonded Texium Closed Male Luer with priming cap Back Check Valve 2 SmartSite Y-Sites, REF: 24010-0007T; SmartSite Bag Access Non-Vented 0.2 MicronFilter Bonded Texium Closed Low Sorbing Tubing (PE Lined) Male Luer with priming cap SmartSite Y-site, REF: 24301-0007T; SmartSite Bag Access Non-Vented Low Sorbing Texium Back Check Valve 2 SmartSite Y-Sites, REF: 24601-B007T. BD 31-IN SEC SET 20DP W/BAG ACCESS PORT TEXIUMTM AND HANGER LOW SORBING, REF: 70001B-07T; BD 20-IN EXTENSION SET W/0.2 MICRON FILTER ONE SmartSite VALVE AND Texium, REF: 20350ET; BD Secondary Set 20 Drop Vented Non Vented With Texium & Hanger, REF: 40000-07T; BD 20 DROP SECONDARY SET SmartSite VALVE BAG ACCESS W/TEXIUM CLOSED MALE LUER HANGER, REF: 10013364T; BD 32-IN GRAV SET MICRON FILTER BAG ACCESS PORT SS VALVE TEXIUMTM LOW ABSORPTION, REF: 4030B-07T; BD Texium Smartsite MICROBORE EXTENSION SET 60 IN, REF: 22005E-07T

Code information (models, lots, serial numbers)

Show the full code information (5k characters)

REF/UDI-DI/Lot(Expiration): 10013361T/10885403223204/23025303(February 10, 2026), 23035235(March 11, 2026), 23045223(April 6, 2026), 23075395(July 17, 2026), 23085227(August 8, 2026), 23115150(November 8, 2026), 23115151(November 8, 2026), 23125167(December 5, 2026), 24025684(February 6, 2027), 24045377(April 23, 2027), 24055576(May 27, 2027), 24075224(July 26, 2027), 24105079(October 11, 2027), 24115159(November 7, 2027); 10321213T/10885403228971/23035169(March 15, 2026), 23045174(April 4, 2026), 23085140(August 3, 2026), 23095142(September 11, 2026), 23105297(October 12, 2026), 24025686(February 6, 2027); 22000-B007T/10885403245022/23025320(February 13, 2026), 23045224(April 11, 2026), 23065209(June 10, 2026), 23075394(July 21, 2026), 23095143(September 11, 2026), 23105133(October 6, 2026), 24015278(January 13, 2027), 24025685(February 6, 2027), 24036209(March 25, 2027), 24075260(July 26, 2027), 24066653(June 24, 2027); 22600-0007T/10885403239595/23095151(September 11, 2026), 24036205(March 25, 2027), 24055601(May 27, 2027); 22602-B007T/10885403240935/23045501(April 20, 2026), 23095152(September 11, 2026), 24015186(January 9, 2027), 24026152(February 27, 2027), 24036206(March 25, 2027), 24045384(April 23, 2027), 24115032(November 2, 2027); 22603-B007T/10885403240942/23035482(March 23, 2026), 23035504(March 25, 2026), 23055379(May 23, 2026), 23065210(June 12, 2026), 23075056(July 13, 2026), 23085141(August 3, 2026), 23095808(September 14, 2026), 23105296(October 12, 2026), 23125107(December 4, 2026), 24015185(January 9, 2027), 24025681(February 6, 2027), 24026153(February 27, 2027), 24026154(February 27, 2027), 24036207(March 25, 2027), 24045381(April 23, 2027), 24055577(May 27, 2027), 24066654(June 24, 2027), 24085243(August 26, 2027), 24095223(September 23, 2027), 24115312(December 2, 2027); 24010-0007T/10885403271021/23025108(February 8, 2026), 23035249(March 14, 2026), 23075392(July 18, 2026), 23085223(August 8, 2026), 23095763(September 19, 2026), 23105136(October 6, 2026), 23115054(November 3, 2026), 24015279(January 13, 2027), 24015386(January 17, 2027), 24025683(February 6, 2027), 24026149(February 27, 2027), 24045382(April 23, 2027), 24055598(May 27, 2027), 24075225(July 26, 2027), 24085240(August 26, 2027), 24095220(September 23, 2027), 24115311(November 14, 2027), 24066656(June 24, 2027); 24301-0007T/10885403223198/23025183(February 9, 2026), 23025238(February 10, 2026), 23035338(March 16, 2026), 23045222(April 12, 2026), 23075393(July 20, 2026), 23085226(August 8, 2026), 23095764(September 19, 2026), 23105284(October 11, 2026), 23115055(November 3, 2026), 24026089(February 21, 2027), 24026150(February 27, 2027), 24036178(March 25, 2027), 24045383(April 23, 2027), 24055599(May 27, 2027), 24075212(July 26, 2027), 24085241(August 26, 2027), 24095221(September 23, 2027), 24115279(November 12, 2027), 24015474(January 20, 2027); 24601-B007T/10885403240959/23035170(March 9, 2026), 23055380(May 24, 2026), 23065280(June 13, 2026), 23085142(August 3, 2026), 23095153(September 11, 2026), 23105285(October 11, 2026), 23115052(November 3, 2026), 24025682(February 6, 2027), 24026151(February 27, 2027), 24036208(March 25, 2027), 24055600(May 27, 2027), 24075261(July 26, 2027), 24085242(August 26, 2027), 24105106(October 22, 2027), 24066655(June 24, 2027), 24015475(January 20, 2027); 70001B-07T/10885403240997/24109019(October 2, 2027), 24109018(October 2, 2027), 24089215(August 9, 2027), 24089214(August 9, 2027), 24019266(January 16, 2027), 23109293(October 23, 2026), 24099179(September 5, 2027), 24049266(April 15, 2027), 24039248(March 14, 2027); 20350ET/10885403229756/24129026(November 13, 2027), 24079037(July 2, 2027), 24059361(May 15, 2027), 25029088(February 4, 2028), 24049028(April 2, 2027); 40000-07T/10885403225628/24119116(November 6, 2027), 24089019(August 1, 2027), 24079173(July 3, 2027), 24059480(May 28, 2027), 24059152(May 7, 2027), 24019267(January 16, 2027), 23119354(November 29, 2026), 24049267(April 15.2027), 24039212(March 12, 2027), 23109275(October 19, 2026), 23079215(July 20, 2026); 10013364T/10885403198915/24109021(October 2, 2027), 24109020(October 2, 2027), 24099172(September 5, 2027), 24089059(August 5, 2027), 24089058(August 4, 2027), 24079175(July 3, 2027), 24079174(July 3, 2027), 24059506(May 30, 2027), 24059469(May 28, 2027), 24059151(May 7, 2027), 24039059(March 4, 2027), 24019435(January 31, 2027), 24019268(January 16, 2027), 24019433(January 30, 2027), 24019269(January 16, 2027), 23119087(November 6, 2026), 23129229(December 13, 2026), 23129230(December 13, 2026), 23059178(May 24, 2026), 23059177(May 23, 2026), 23059156(May 16, 2026), 23049061(April 26, 2026), 24059150(May 7, 2027), 24049265(April 15, 2027), 23109175(October 13, 2026), 23109006(October 2, 2026), 23099219(September 12, 2026), 23099217(September 11, 2026), 23059179(May 29, 2026); 4030B-07T/10885403240867/24089216(August 9, 2027), 24059484(May 28, 2027), 24039060(March 4, 2027), 25029393(February 27, 2028), 23109296(October 23, 2026); 22005E-07T/10885403250781/23029335(February 24, 2026)

Devices named in this recall

15 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 96737).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page