Adelante Peel-Away Introducer Set AP07016B5Recalls & reported problems
Oscor Inc. · Peel-Away Introducer Set, 7F · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10885672100022), received between Feb 13, 2018 and Jun 10, 2024 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Defective Device” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K073100); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Defective Device1
- Detachment of Device or Device Component1
- Material Frayed1
- Material Split, Cut or Torn1
- Material Too Rigid or Stiff1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Blood Loss1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 10, 2024Material Too Rigid or Stiff; Material Too Rigid or Stiff
SUMMARY: DIFFICULTY IN PEELING AWAY THE SHEATH. THE CUSTOMER REPORTED FOR A NEW PACEMAKER IMPLANTATION, A LEAD WAS INSERTED AFTER THE SHEATH IN QUESTION WAS INSERTED. AFTER THE LEAD REACHED THE RIGHT ATRIUM, AN ATTEMPT WAS MADE TO PEEL THE SHEATH AWAY, BUT THE SHEATH SHAFT PART WAS TORN HALFWAY THROUGH THE ATTEMPT. THE SHEATH WAS THEN REMOVED BY HOLDING THE TORN PART WITH FORCEPS. THE LEAD REMAINED IN THE RIGHT ATRIUM, AND THE PROCEDURE WAS CONTINUED AND COMPLETED WITHOUT ANY PROBLEMS. NO HEALTH DAMAGE WAS REPORTED. THIS EVENT IS NOT REPORTABLE TO PMDA. (B)(6) 2024: THE CUSTOMER STATED THERE WAS NO HARM TO THE PATIENT. PATIENT INFORMATION PROVIDED; INITIAL (B)(6), 72-YEAR-OLD MALE. THERE WAS NO MEDICAL OR SURGICAL INTERVENTION AND NO PROCEDURE DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19505155 ↗MalfunctionJan 26, 2023Material Frayed; Material Split, Cut or Torn; Material Frayed; Material Split, Cut or Torn
DURING IMPLANTATION, BEFORE INSERTING THE SHEATH, IT IS NOTICED THAT THE TIP OF THE INTRODUCER IS FRAYED. NO ADVERSE PATIENT EFFECTS WERE REPORTED. CUSTOMER DID NOT PROVIDE THE EVENT DATE, HOWEVER, ON (B)(6)2023, THEY SUBMITTED THEIR EVENT REPORT FORM. BASED UPON THIS INFORMATION, THIS IS THE DATE WE ARE SUBMITTING AS THE EVENT DATE. SHOULD THE ACTUAL EVENT DATE BE PROVIDED, WE WILL UPDATE OUR REPORT ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16248954 ↗MalfunctionJan 13, 2020Defective Device; Defective Device
THE CUSTOMER REPORTED THE LENGTH OF THE DILATOR WAS FOUND UNUSUALLY SHORT, IT WAS FOUND BEFORE THE DEVICE WAS USED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9581946 ↗InjuryFeb 13, 2018Detachment of Device or Device Component; Detachment of Device or Device Component
SUMMARY: THE SHEATH HUB OF ADELANTE AP07016B5 WAS DETACHED FROM THE SHEATH TUBE WHEN A DILATOR AND GUIDEWIRE WERE EXTRACTED. DESCRIPTION / SEQUENCE OF EVENTS: (B)(6) 2016 04:00 P.M. DURING A PACING LEAD IMPLANTATION PROCEDURE, WHEN A DILATOR AND GUIDEWIRE WERE EXTRACTED FROM THE LEFT SUBCLAVIAN VEIN THROUGH THE AP07016B5 IN QUESTION, THE SHEATH HUB ATTACHED TO THE SHEATH TUBE WAS UNEXPECTEDLY DETACHED AND ONLY THE SHEATH TUBE WAS LEFT INSIDE THE VEIN. AS A SLIGHT PART OF THE SHEATH TUBE WAS STICKING OUT FROM THE VEIN, THE SHEATH TUBE WAS SUCCESSFULLY REMOVED FROM THE VEIN USING FORCEPS. ·DURING THE REMOVAL OF THE SHEATH TUBE, A SIGNIFICANT AMOUNT OF BLOOD LEAKED OUT, WHICH DECREASED THE VISIBILITY INSIDE THE DEVICE POCKET.
Patient problems: Blood Loss
MAUDE report 7267659 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K073100. One submission often covers many models, so check the product and lot details in each record.
- Z-0055-2012 · Aug 1, 2011 · Terminated8F Safe Sheath Introducer, Catalog number: V18; AngioDynamics, INC. The Valved Introducer sheath is utilized to surgically place the catheter during the port implantation procedu
- Z-0130-2009 · Jul 3, 2008 · TerminatedOscor Adelante Luer-Lock Peel Away Introducer Set, for introduction of diagnostic or therapeutic devices into the body.
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page