GRYPHON ORTHOCORD 210811Recalls & reported problems

Medos International Sàrl · GRYPHON T BR DS Anchor w/ORTHOCORD TCP/PLGA Absorbable Anchor (1) Violet (1) Blue Strand, Size 2 (5 metric) ORTHOCORD braided composite suture, 36 Inches (91cm) · Class 2 · Implant

All models of GRYPHON ORTHOCORD (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medos GRYPHON ORTHOCORD anchor is a soft-tissue fixation anchor with a braided suture, used in hip, shoulder, and foot repair surgery. The 30 reports mostly describe the anchor breaking during surgery, sometimes in pieces that had to be removed, or pulling loose so it could not be repositioned; a few mention the anchor shifting or being hard to place. No deaths and 6 injuries were reported, along with 24 malfunctions. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
30
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 10886705001255), received between Jul 2, 2018 and Jan 22, 2026 — classified as 6 injuries, 24 malfunctions.

The most frequently reported specific device problem is “Break” (23 reports).

6 other recalls cover devices cleared under the same FDA submission (K073412, K150209); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 4 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break23
  • Detachment of Device or Device Component5
  • Migration2
  • Difficult to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions21
  • No Code Available6
  • No Consequences Or Impact To Patient3
  • Not Applicable3
  • Foreign Body In Patient2

Recent adverse event reports

The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 22, 2026Difficult to Advance

    IT WAS REPORTED THAT DURING LABRAL REPAIR SURGERY, THE GRYPHON T BR DS ANCH W/OC ANCHOR DEVICE SLIPPED AND COULDN'T BE FIXED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24149610 ↗
  • MalfunctionJan 9, 2026Migration

    IT WAS REPORTED THAT DURING A HIP LABRAL REPAIR PROCEDURE, THE ANCHOR ON THE GRYPHON T BR DS ANCH W/OC DEVICE PULLED OUT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24021235 ↗
  • MalfunctionJun 23, 2025Break; Migration; Break; Migration

    REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED THAT DURING A ARTHROSCOPIC SHOULDER STABILIZING SURGERY, IT WAS OBSERVED THAT A GRYPHON T BR DS ANCH W/OC DEVICE ANCHOR WAS IN A GRYPHON P BR DS ANCHOR W/OC ANCHOR PACKAGING. ACCORDING TO THE REPORTER, AS NO ONE EXPECTED IT, THE ANCHOR WAS USED LIKE A GRYPHON P DEVICE, AND THE ANCHOR BROKE AND PULLED OUT. THERE WAS A FIFTEEN MINUTE DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22305942 ↗
  • MalfunctionAug 5, 2024Break; Break

    IT WAS REPORTED FROM CHINA THAT DURING AN OPEN REDUCTION AND INTERNAL FIXATION OF THE FIFTH METATARSAL BASE FRACTURE IN THE LEFT FOOT SURGERY, IT WAS DISCOVERED THAT THE ANCHOR ON THE GRYPHON T BR DS ANCH W/OC DEVICE BROKE OFF. ALL THE BROKEN PARTS WERE REMOVED. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19905744 ↗
  • MalfunctionJul 15, 2024Break; Break

    IT WAS REPORTED, BY THE SALES REP IN CHINA. THAT DURING, A HIP GLENOID LIGAMENT SURGICAL PROCEDURE ON (B)(6) 2024. THE 2X GRYPHON T BR DS ANCH W/OC DEVICE ANCHOR WAS BROKEN OFF. ALL THE BROKEN PARTS WERE REMOVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 2 OF 2 OF THE SAME EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19742775 ↗
  • MalfunctionJul 15, 2024Break; Break

    IT WAS REPORTED BY THE SALES REP IN CHINA THAT DURING A HIP GLENOID LIGAMENT SURGICAL PROCEDURE ON (B)(6) 2024 THE 2X GRYPHON T BR DS ANCH W/OC DEVICE ANCHOR WAS BROKEN OFF. ALL THE BROKEN PARTS WERE REMOVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 1 OF 2 OF THE SAME EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19742772 ↗
  • MalfunctionJul 11, 2024Break; Break

    THIS IS REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED FROM CHINA THAT DURING A HIP GLENOID LIGAMENT REPAIR PROCEDURE, IT WAS DISCOVERED THAT THE ANCHOR ON THE GRYPHON T BR DS ANCH W/OC DEVICE WAS BROKEN OFF. ALL THE BROKEN PARTS WERE REMOVED. CHANGED TO ANOTHER DEVICE TO CONTINUE THE SURGERY, BUT THE SAME PROBLEM OCCURRED AGAIN. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19719887 ↗
  • MalfunctionJul 11, 2024Break; Break

    THIS IS REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED FROM CHINA THAT DURING A HIP GLENOID LIGAMENT REPAIR PROCEDURE, IT WAS DISCOVERED THAT THE ANCHOR ON THE GRYPHON T BR DS ANCH W/OC DEVICE WAS BROKEN OFF. ALL THE BROKEN PARTS WERE REMOVED. CHANGED TO ANOTHER DEVICE TO CONTINUE THE SURGERY, BUT THE SAME PROBLEM OCCURRED AGAIN. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19719756 ↗
  • MalfunctionJun 19, 2024Break; Break

    IT WAS REPORTED FROM CHINA THAT DURING A LIP REPAIR PROCEDURE, IT WAS DISCOVERED THAT WHEN THE SUTURE ON THE GRYPHON T BR DS ANCH W/OC WAS TIGHTENED, IT BROKE OFF. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19566503 ↗
  • MalfunctionJun 11, 2024Break; Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE GRYPHON T BR DS ANCH W/OC DEVICE ANCHOR WAS BROKEN OFF, REMOVED ALL THE BROKEN PARTS. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19512725 ↗
  • MalfunctionSep 25, 2023Break; Break

    THIS IS REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN UNKNOWN PROCEDURE OF THE HIP ON (B)(6) 2023, IT WAS OBSERVED THAT THE ANCHOR ON THE GRYPHON T BR DS ANCHORS W/ORTHOCORD DEVICE WAS BROKEN OFF, THEN REMOVED ALL THE BROKEN PARTS. CHANGED TO ANOTHER ONE TO CONTINUE THE SURGERY BUT THE SAME PROBLEM HAPPENED AGAIN. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17809837 ↗
  • MalfunctionSep 25, 2023Break; Break

    THIS IS REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN UNKNOWN PROCEDURE OF THE HIP ON (B)(6) 2023, IT WAS OBSERVED THAT THE ANCHOR ON THE GRYPHON T BR DS ANCHORS W/ORTHOCORD DEVICE WAS BROKEN OFF, THEN REMOVED ALL THE BROKEN PARTS. CHANGED TO ANOTHER ONE TO CONTINUE THE SURGERY BUT THE SAME PROBLEM HAPPENED AGAIN. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17809818 ↗
  • MalfunctionSep 25, 2023Break; Break

    THIS IS REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN UNKNOWN PROCEDURE OF THE HIP ON (B)(6) 2023, IT WAS OBSERVED THAT THE ANCHOR ON THE GRYPHON T BR DS ANCHORS W/ORTHOCORD DEVICE WAS BROKEN OFF, THEN REMOVED ALL THE BROKEN PARTS. CHANGED TO ANOTHER ONE TO CONTINUE THE SURGERY BUT THE SAME PROBLEM HAPPENED AGAIN. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17809694 ↗
  • MalfunctionSep 25, 2023Break; Break

    THIS IS REPORT 1 OF 2 FOR PC-(B)(4). IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN UNKNOWN PROCEDURE OF THE HIP ON (B)(6) 2023, IT WAS OBSERVED THAT THE ANCHOR ON THE GRYPHON T BR DS ANCHORS W/ORTHOCORD DEVICE WAS BROKEN OFF, THEN REMOVED ALL THE BROKEN PARTS. CHANGED TO ANOTHER ONE TO CONTINUE THE SURGERY BUT THE SAME PROBLEM HAPPENED AGAIN. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17809677 ↗
  • MalfunctionSep 25, 2023Break; Break

    IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN UNKNOWN PROCEDURE OF THE HIP ON (B)(6) 2023, IT WAS OBSERVED THAT THE ANCHOR ON THE GRYPHON T BR DS ANCHORS W/ORTHOCORD DEVICE WAS BROKEN OFF, THEN REMOVED ALL THE BROKEN PARTS. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17809498 ↗
  • MalfunctionApr 6, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS IS REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED BY THE AFFILIATE IN ITALY THAT DURING AN ARTHROSCOPIC SHOULDER STABILIZATION PROCEDURE ON (B)(6) 2023, TWO GRYPHON T BR DS ANCHORS W/ORTHOCORD DEVICES WERE USED. ACCORDING TO THE REPORT, AFTER POSITIONING THE ANCHOR AND PASSING THE SUTURE THREAD THROUGH THE LABRUM, THERE WAS DISASSEMBLY OF THE THREAD FROM THE ANCHOR OCCURRED AT THE TIME OF SUTURE RETRIEVAL TO TIE THE KNOT. IT WAS UNKNOWN IF AND HOW THE PROCEDURE WAS COMPLETED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16691796 ↗
  • MalfunctionApr 6, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THIS IS REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED BY THE AFFILIATE IN ITALY THAT DURING AN ARTHROSCOPIC SHOULDER STABILIZATION PROCEDURE ON (B)(6) 2023, TWO GRYPHON T BR DS ANCHORS W/ ORTHOCORD DEVICES WERE USED. ACCORDING TO THE REPORT, AFTER POSITIONING THE ANCHOR AND PASSING THE SUTURE THREAD THROUGH THE LABRUM, THERE WAS DISASSEMBLY OF THE THREAD FROM THE ANCHOR OCCURRED AT THE TIME OF SUTURE RETRIEVAL TO TIE THE KNOT. IT WAS UNKNOWN IF AND HOW THE PROCEDURE WAS COMPLETED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16691777 ↗
  • MalfunctionMar 3, 2022Break; Break

    IT WAS REPORTED BY THE CUSTOMER IN (B)(6) THAT DURING AN ARTHROSCOPIC SHOULDER STABILIZATION PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE ANCHOR ON THE GRYPHON T BR DS ANCH W/OC DEVICE BROKE. ALL THE BROKEN PARTS WERE REMOVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13660088 ↗
  • MalfunctionFeb 28, 2022Break; Break

    IT WAS REPORTED BY THE CUSTOMER IN (B)(6) THAT DURING AN ARTHROSCOPIC HIP REPAIR OF GLENOID LIP PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE ANCHOR ON THE GRYPHON T BR DS ANCH W/OC DEVICE BROKE. ALL THE BROKEN PARTS WERE REMOVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13623861 ↗
  • MalfunctionJan 10, 2022Break; Break

    IT WAS REPORTED BY THE CUSTOMER IN (B)(6) THAT DURING A KNEE SURGERY FOR PATELLAR DISLOCATION OF THE RIGHT KNEE ON (B)(6) 2021, IT WAS OBSERVED THAT BOTH THE SUTURE AND ANCHOR ON THE GRYPHON T BR DS ANCH W/OC DEVICE WERE BROKEN OFF. ALL THE BROKEN PARTS WERE REMOVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13206918 ↗
  • MalfunctionNov 23, 2021Break; Break

    IT WAS REPORTED BY THE CUSTOMER IN (B)(6) THAT DURING AN ANKLE LIGAMENT REPAIR SURGERY ON (B)(6) 2021, IT WAS OBSERVED THAT THE SUTURE ON THE GRYPHON T BR DS ANCH W/OC DEVICE BROKE WHILE TIGHTENING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12862881 ↗
  • InjuryAug 22, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED BY THE AFFILIATE VIA EMAIL THAT DURING SURGERY THREE GRYPHON T BR DS ANCHOR TRIED TO BE IMPLANTED, WHICH DETACHED AND COULD NOT BE IMPLANTED. ADDITIONAL INFORMATION RECEIVED FROM THE AFFILIATE REPORTED 3 DEVICES; PRODUCT CODE 210811, OF LOT L630469 WERE USED DURING THE PROCEDURE. IT WAS ALSO REPORTED THERE WAS AN HOUR SURGICAL DELAY WITH NO PATIENT CONSEQUENCES AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER ANCHOR; PRODUCT CODE 222245. THE AFFILIATE ALSO REPORTED THE FAILURE OF THE DEVICE OCCURRED WHEN THE SURGEON IMPLANTED THE ANCHOR AND THEN PULL THE SUTURES THE ANCHOR DETACHED. IT WAS STATED THE DEVICES ARE AVAILABLE FOR RETURN.

    Patient problems: No Code Available; Not Applicable

    MAUDE report 8920153 ↗
  • InjuryAug 22, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED BY THE AFFILIATE VIA EMAIL THAT DURING SURGERY THREE GRYPHON T BR DS ANCHOR TRIED TO BE IMPLANTED, WHICH DETACHED AND COULD NOT BE IMPLANTED. ADDITIONAL INFORMATION RECEIVED FROM THE AFFILIATE REPORTED 3 DEVICES; PRODUCT CODE 210811, OF LOT L630469 WERE USED DURING THE PROCEDURE. IT WAS ALSO REPORTED THERE WAS AN HOUR SURGICAL DELAY WITH NO PATIENT CONSEQUENCES AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER ANCHOR; PRODUCT CODE 222245. THE AFFILIATE ALSO REPORTED THE FAILURE OF THE DEVICE OCCURRED WHEN THE SURGEON IMPLANTED THE ANCHOR AND THEN PULL THE SUTURES THE ANCHOR DETACHED. IT WAS STATED THE DEVICES ARE AVAILABLE FOR RETURN.

    Patient problems: No Code Available; Not Applicable

    MAUDE report 8920105 ↗
  • InjuryAug 22, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED BY THE AFFILIATE VIA EMAIL THAT DURING SURGERY THREE GRYPHON T BR DS ANCHOR TRIED TO BE IMPLANTED, WHICH DETACHED AND COULD NOT BE IMPLANTED. ADDITIONAL INFORMATION RECEIVED FROM THE AFFILIATE REPORTED 3 DEVICES; PRODUCT CODE 210811, OF LOT L630469 WERE USED DURING THE PROCEDURE. IT WAS ALSO REPORTED THERE WAS AN HOUR SURGICAL DELAY WITH NO PATIENT CONSEQUENCES AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER ANCHOR; PRODUCT CODE 222245. THE AFFILIATE ALSO REPORTED THE FAILURE OF THE DEVICE OCCURRED WHEN THE SURGEON IMPLANTED THE ANCHOR AND THEN PULL THE SUTURES THE ANCHOR DETACHED. IT WAS STATED THE DEVICES ARE AVAILABLE FOR RETURN.

    Patient problems: No Code Available; Not Applicable

    MAUDE report 8920100 ↗
  • InjuryAug 20, 2019Break; Break

    IT WAS REPORTED BY THE AFFILIATE VIA PHONE THAT DURING THE ARTHROSCOPIC LABRAL REPAIR OF SHOULDER JOINT PROCEDURE, WHEN INSERT THE GRYPHON T BR DS ANCH W/OC, THE DEVICES WERE BROKEN OFF. TWO OF THEM REMAINED IN THE PATIENT, ONE BROKEN PART WAS REMOVED FROM PATIENT AS THE PHOTO SHOWS. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. AS THE MATERIAL IS BIO,THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THE SURGERY TIME WAS NOT DELAYED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Code Available

    MAUDE report 8910227 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K073412, K150209. One submission often covers many models, so check the product and lot details in each record.

  • Z-1679-2023 · Apr 13, 2023 · Open, Classified
    DePuy Mitek BIOKNOTLESS Plus w/ OC-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212726
  • Z-1678-2023 · Apr 13, 2023 · Open, Classified
    DePuy Mitek BIOKNOTLESS Plus w/ Panacryl-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212725
  • Z-1680-2023 · Apr 13, 2023 · Open, Classified
    DePuy Mitek BIOKNOTLESS Rapid W/ OC-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212722
  • Z-1677-2023 · Apr 13, 2023 · Open, Classified
    DePuy Mitek BIOKNOTLESS Plus w/ Ethibond-Iindicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212724
  • Z-1681-2023 · Apr 13, 2023 · Open, Classified
    DePuy Mitek BIOKNOTLESS Rapid w/ Pan-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212723
  • Z-2208-2011 · Feb 24, 2011 · Terminated
    DePuy Mitek GRYPHON T BR DS Anchor w/ORTHOCORD Product Code: 210811 Gryphon BR Suture Anchors are intended for: Shoulder: Rotor Cuff Repair, Bankart Repair, SLAP Lesion Repair, Bic

Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices

All FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices →