GRYPHON 211039Recalls & reported problems
Medos International Sàrl · GRYPHON Sawtooth Guide · Class 1
All models of GRYPHON (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10886705001514), received between Sep 26, 2019 and Nov 21, 2022 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Deformation4
- Break2
- Detachment of Device or Device Component2
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 21, 2022Material Deformation; Material Deformation
IT WAS REPORTED BY SALES REP VIA EMAIL THAT PRE-OP TO A ROTATOR CUFF REPAIR PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT A GRYPHON SAWTOOTH GUIDE WAS BENT. ACCORDING TO THE REPORT, THE DEVICE COULD DRIVE BUT WOULD NOT FIT DOWN GUIDE. THEY HAD A SLOTTED GRYPHON GUIDE IN TRAY TO COMPLETE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15838803 ↗MalfunctionNov 4, 2022Material Deformation; Material Deformation
IT WAS REPORTED BY THE SALES REP THAT BEFORE STERILIZATION ON 10/21/2022, IT WAS OBSERVED THAT THE GRYPHON SAWTOOTH GUIDE DEVICE WAS BENT. NO PROCEDURE OR PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15727009 ↗MalfunctionJun 15, 2022Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER IN (B)(6) THAT IT WAS OBSERVED THAT THE GRYPHON SAWTOOTH GUIDE WAS BENT AT THE TIP WHILE CHECKING THE SETS. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14702811 ↗MalfunctionJun 7, 2022Material Deformation; Material Deformation
IT WAS REPORTED BY THE CUSTOMER THAT PREOPERATIVELY TO AN UNKNOWN SURGERY ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE GRYPHON SAWTOOTH GUIDE DEVICE WAS NOT STRAIGHT WHEN TESTED OUT WITH TROCAR INSERT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14626216 ↗MalfunctionMay 9, 2022Break; Break
IT WAS REPORTED BY THE CUSTOMER THAT DURING A SHOULDER REPAIR SURGERY ON (B)(6) 2022, IT WAS OBSERVED THAT THE GRYPHON SLOTTED SAWTOOTH GDE DEVICE BROKE INTO TWO PIECES. THERE WERE NO FRAGMENTS GENERATED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14337802 ↗MalfunctionAug 17, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY SALES REP VIA COMPLAINT SUBMISSION TOOL THAT PRE-OPERATIVELY TO A ROTATOR CUFF REPAIR WHILE USING A GRYPHON SAWTOOTH GUIDE, DRILL GUIDE OF HANDLE CAME OFF. NO SURGICAL DELAY OR PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10417511 ↗MalfunctionJun 10, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED BY THE SALES REP VIA PHONE THAT DURING A ROTATOR CUFF REPAIR THE GRYPHON SAWTOOTH GUIDE DETACHED FROM THE SHAFT. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT CONSEQUENCES OR SURGICAL DELAY REPORTED. THE DEVICE IS AVAILABLE FOR EVALUATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10140244 ↗MalfunctionSep 26, 2019Break; Break
IT WAS REPORTED BY THE SALES REP VIA PHONE THAT DURING A LABRUM REPAIR THE CUSTOMER'S GRYPHON SAWTOOTH GUIDE GOT STUCK IN THE CANNULA AND THE HANDLE SEPARATED FROM THE REST OF THE GUIDE. THE SALES REP STATED THAT FRAGMENTS WERE NOT CREATED WHEN THE DEVICE BROKE AND THE DEVICE WAS FULLY REMOVED FROM THE PATIENT. THE CASE WAS COMPLETED WITH ANOTHER LIKE-DEVICE WITH NO PATIENT HARM OR DELAY. THE SALES REP STATED THERE WAS NO NEED FOR SURGICAL INTERVENTION. THE SALES REP WAS NOT PRESENT AND COULD NOT PROVIDE A LOT NUMBER OR ANY ADDITIONAL INFORMATION. THERE WAS NO DELAY IN THE SURGICAL PROCEDURE. THERE WAS A SPARE DEVICE AVAILABLE FOR USE TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9122735 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.