QUICKANCHOR ETHIBOND 212133Recalls & reported problems
Depuy Mitek, LLC · Mini QUICKANCHOR Plus (Number 0 Suture) Size 0 (3.5 metric) green ETHIBOND braided polyester suture, 36 Inches (91cm), double-armed with OS-2 needles, with disposable inserter. · Class 2 · Implant
All models of QUICKANCHOR ETHIBOND (5)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10886705001972), received between May 9, 2018 and Feb 19, 2025 — classified as 1 injury, 16 malfunctions.
The most frequently reported specific device problem is “Break” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break8
- Component Missing2
- Failure to Advance2
- Bent1
- Device-Device Incompatibility1
- Entrapment of Device1
- Failure to Clean Adequately1
- Material Twisted/Bent1
Patient problems
- No Consequences Or Impact To Patient11
- No Clinical Signs, Symptoms or Conditions4
- Not Applicable3
- Injury1
- No Patient Involvement1
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 19, 2025Unintended Movement; Unintended Movement
IT WAS REPORTED THAT DURING A PERONEAL TENDON REPAIR PROCEDURE ON THE RIGHT LEG FOR A PERONEAL TENDON DISLOCATION, IT WAS DISCOVERED THAT THE ANCHOR ON THE MINI QA+ #0 ETH OS-2 HAD BECOME LOOSE. ACCORDING TO THE REPORTER, AFTER INSERTING THE DEVICE INTO THE BONY HOLE, THE SURGEON WAITED TEN SECONDS AND THEN WITHDREW IT ACCORDING TO THE PROCEDURE MANUAL. HOWEVER, THE ARC DID NOT RETURN TO ITS ARCHED SHAPE, AND THE ANCHOR BECAME LOOSE. IT WAS REPORTED THAT THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE AS A SPARE DEVICE WAS AVAILABLE TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21412943 ↗MalfunctionSep 7, 2023Migration; Migration
IT WAS REPORTED BY THE AFFILIATE IN JAPAN THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE MINI QUICKANCHOR® PLUS #0, SINGLE PACK DEVICE CAME OFF AFTER INSERTION. IT WAS UNKNOWN IF AND HOW THE PROCEDURE WAS COMPLETED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17701256 ↗MalfunctionJul 14, 2023Break; Break
IT WAS REPORTED BY THE SALES REP IN SOUTH KOREA THAT PREOPERATIVELY TO AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE MINI QUICKANCHOR® PLUS #0, SINGLE PACK DEVICE WAS BROKEN. IT WAS UNKNOWN IF AND HOW THE PROCEDURE WAS COMPLETED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17325670 ↗MalfunctionOct 29, 2021Break; Device-Device Incompatibility; Failure to Clean Adequately; Break; Device-Device Incompatibility; Failure to Clean Adequately
IT WAS REPORTED BY SALES REP THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2021, IT WAS OBSERVED THAT THE A MINI QUICKANCHOR® PLUS #0, SINGLE PACK DID NOT DEPLOY PROPERLY. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12723505 ↗MalfunctionNov 10, 2020Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED BY THE SALES REP VIA PERSONAL INTERACTION THAT DURING AN UNKNOWN PROCEDURE AND AFTER DEPLOYING TWO ANCHORS, THE SURGEON INSERTED THE 3RD ANCHOR AND WHEN THE SUTURE WAS PULLED THE MINI QA+ #0 ETH OS-2 *EA CAME OFF. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE USING THE SAME BURR HOLE. NO PATIENT CONSEQUENCES OR SURGICAL DELAY REPORTED. THE DEVICE WAS BRAND NEW AND THE FIRST USE WHEN THE ISSUE OCCURRED. IT IS UNKNOWN IF THE DEVICE IS AVAILABLE FOR EVALUATION. ADDITIONAL INFORMATION PROVIDED BY THE AFFILIATE REPORTED THE DEVICE WILL NOT RETURNED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10820338 ↗MalfunctionNov 5, 2020Failure to Advance; Failure to Advance
IT WAS REPORTED THAT THIS WAS AN UNKNOWN PROCEDURE PERFORMED ON (B)(6) 2020. ALTHOUGH THE SURGEON FOLLOWED THE SURGICAL TECHNIQUE ON INSERTING THE ANCHOR INTO THE BURR HOLE, THE ANCHOR CAME OFF. THERE WAS NO SURGICAL DELAY AND NO HARM TO THE PATIENT. THE DEVICE WAS BRAND NEW AND THE FIRST USE WHEN THE ISSUE OCCURRED. ADDITIONAL INFORMATION RECEIVED FROM THE AFFILIATE REPORTED THE DEVICE WILL NOT BE RETURNING FOR EVALUATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10795912 ↗MalfunctionJun 29, 2020Component Missing; Component Missing
IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT THE NEEDLE WAS MISSING INSIDE THE BOX WHICH PREVENTED THE PROCEDURE. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE, OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WERE NO REPORTS OF ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10207268 ↗MalfunctionJun 29, 2020Component Missing; Component Missing
IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT THE NEEDLE WAS MISSING INSIDE THE BOX WHICH PREVENTED THE PROCEDURE. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE, OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WERE NO REPORTS OF ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10207247 ↗InjuryJun 29, 2020Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED BY THE AFFILIATE IN (B)(4) THAT DURING A HAND SURGERY, IT WAS OBSERVED THAT TWO ANCHORS FAILED. ON THE FIRST ANCHOR: THE INSERTER GOT BENT WHEN INSERTED AND COULDN'T BE USED SINCE THE ANCHOR LOOSENED FROM INSERTER. ON THE SECOND ANCHOR: THE SUTURE BROKE AND THE ANCHOR WAS LEFT INSIDE THE PATIENT. SURGEON MENTIONED THAT THE SUTURE SLIDED POORLY IN ANCHOR. DRILLING WAS USED IN NORMAL WAY. SURGEON VERY EXPERIENCED IN USING THESE ANCHORS. SURGERY WAS DELAYED OF FIVE MINUTES. A THIRD ANCHOR WAS USED TO COMPLETE THE SURGERY SUCCESSFULLY. THERE WERE NO PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: Injury; Not Applicable
MAUDE report 10207128 ↗MalfunctionJun 29, 2020Failure to Advance; Material Twisted/Bent; Failure to Advance; Material Twisted/Bent
IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT DURING A HAND SURGERY, IT WAS OBSERVED THAT TWO ANCHORS FAILED. ON THE FIRST ANCHOR: THE INSERTER GOT BENT WHEN INSERTED AND COULDN'T BE USED SINCE THE ANCHOR LOOSENED FROM INSERTER. ON THE SECOND ANCHOR: THE SUTURE BROKE AND THE ANCHOR WAS LEFT INSIDE THE PATIENT. SURGEON MENTIONED THAT THE SUTURE SLIDED POORLY IN ANCHOR. DRILLING WAS USED IN NORMAL WAY. SURGEON VERY EXPERIENCED IN USING THESE ANCHORS. SURGERY WAS DELAYED OF FIVE MINUTES. A THIRD ANCHOR WAS USED TO COMPLETE THE SURGERY SUCCESSFULLY. THERE WERE NO PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: Not Applicable
MAUDE report 10207113 ↗MalfunctionDec 10, 2019Break; Break
IT WAS REPORTED THAT DURING THE LATERAL MALLEOLUS LIGAMENT REPAIR SURGERY, WHEN TIED THE KNOT, THE SUTURE WAS BROKEN,REMOVED THE ANCHOR WITH 3-0 ABSORBABLE SUTURE TO REPLACE THE BROKEN SUTURE TO COMPLETE THE SURGERY. THE SURGERY DELAYED ABOUT 30 MINUTES. THE PATIENT IS STABLE AND IN HOSPITAL NOW.
Patient problems: No Consequences Or Impact To Patient; No Patient Involvement; Not Applicable
MAUDE report 9446801 ↗MalfunctionJan 24, 2019Break; Break
THIS IS REPORT 2 OF 2 FOR THE SAME EVENT. IT WAS REPORTED BY THE SALES REP THAT THE SUTURE BROKE ON TWO OF THEIR MINI QUICKANCHOR PLUS #0 ETHIB WHEN TYING THE TISSUE DOWN DURING AN UNSPECIFIED SURGICAL PROCEDURE OF THE ANKLE. THERE WERE NO PATIENT CONSEQUENCES OR DELAYS, THE CASE WAS COMPLETED WITH THE DEVICES, CONTINUING WITH THE SMALLER PORTION OF SUTURE THAT WAS LEFT. THE DEVICES ARE NOT BEING RETURNED AS THEY WERE USED TO COMPLETE THE CASE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8275971 ↗MalfunctionJan 24, 2019Break; Break
THIS IS REPORT 1 OF 2 FOR THE SAME EVENT. IT WAS REPORTED BY THE SALES REP THAT THE SUTURE BROKE ON TWO OF THEIR MINI QUICKANCHOR PLUS #0 ETHIB WHEN TYING THE TISSUE DOWN DURING AN UNSPECIFIED SURGICAL PROCEDURE OF THE ANKLE. THERE WERE NO PATIENT CONSEQUENCES OR DELAYS, THE CASE WAS COMPLETED WITH THE DEVICES, CONTINUING WITH THE SMALLER PORTION OF SUTURE THAT WAS LEFT. THE DEVICES ARE NOT BEING RETURNED AS THEY WERE USED TO COMPLETE THE CASE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8275969 ↗MalfunctionOct 19, 2018Break; Break
ADDITIONAL EVENT INFORMATION RECEIVED ON 10/21/2018. NO FURTHER INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7984765 ↗MalfunctionAug 16, 2018Break; Break
OUR AFFILIATE REPORTED THIS PATIENT HAD A FRACTURE OF THE DISTAL TIBIA WHICH WAS FIXATED WITH A 3.5 X 50 MM CANNULA. AN ANCHOR WAS INSTALLED AFTER OPENING THE CORTICAL WITH A DRILL BIT. THE SUTURE BROKE WHILE THE KNOT WAS BEING TIED, AND THE ANCHOR WAS PULLED OUT. AN ATTEMPT WAS MADE TO PASS THE SUTURE ALONE WITH THE NEEDLE, BUT IT BROKE LOSE AGAIN FROM THE NEEDLE ASSEMBLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7788856 ↗MalfunctionJul 9, 2018Positioning Problem; Positioning Problem
IT WAS REPORTED BY A SALES REPRESENTATIVE THAT DURING A TEMPOROMANDIBULAR JOINT SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE SUTURE ON HIS MINI QUICK ANCHOR PLUS W/ETHIBOND WAS UNRAVELING AFTER THE SURGEON TIED THE KNOT. THERE WAS NO DELAY IN THE PROCEDURE. IT WAS REPORTED THE PROCEDURE WAS COMPLETED USING LEFT OVER SUTURE FROM THE KNOT, WRAPPING IT AROUND THE ANCHOR AND TYING ANOTHER KNOT TO REINFORCE IT. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7670782 ↗MalfunctionMay 9, 2018Bent; Bent
ANCHOR TIP BEND. IT WAS NOTED THAT TIP OF ANCHOR BENT WHEN OUT OF PACKAGE. CHANGED THE NEW ONE TO COMPLETE. THERE WAS NO ADVERSE EVENT ON PATIENT REPORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7499845 ↗
FDA clearance and classification
The FDA regulates product code HWC (screw, fixation, bone) under 21 CFR 888.3040 as a Class II device. 21 CFR 888.3040 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the QUICKANCHOR ETHIBOND 212133
Has the QUICKANCHOR ETHIBOND 212133 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886705001972), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the QUICKANCHOR ETHIBOND 212133?
The FDA's MAUDE database lists 17 adverse event reports for this device, received between May 9, 2018 and Feb 19, 2025. In those reports, the most common specific device problems are “Break” (8), “Component Missing” (2) and “Failure to Advance” (2); the most common patient problems are “Not Applicable” (3) and “Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the QUICKANCHOR ETHIBOND 212133?
None of the reports are classified as deaths, 1 as an injury and 16 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the QUICKANCHOR ETHIBOND 212133 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the QUICKANCHOR ETHIBOND 212133, and is it still sold?
The FDA's device database lists Depuy Mitek, LLC as the company responsible for it (brand QUICKANCHOR ETHIBOND, model 212133). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone498 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone248 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-200Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-325Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-550Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-425Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page