QUICKANCHOR ETHIBOND 212867Recalls & reported problems
Depuy Mitek, LLC · MICRO QUICKANCHOR Plus (Number 4/0) Suture 4/0 (1.5 metric) white ETHIBOND braided polyester suture, with P-3 precision point needles and a 1.3 x 5.0mm drill bit · Class 2 · Implant
All models of QUICKANCHOR ETHIBOND (5)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10886705002542), received between Dec 19, 2018 and Jul 22, 2021 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break5
- Difficult to Advance1
- Difficult to Insert1
- Difficult to Open or Close1
- Entrapment of Device1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Consequences Or Impact To Patient4
- Injury1
- Not Applicable1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 22, 2021Difficult to Advance; Difficult to Advance
IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL THAT DURING AN UNKNOWN PROCEDURE ON (B)(4) 2021, IT WAS OBSERVED THAT THE MICRO QUICKANCHOR PLUS PRELOADED WITH 4/0 ETHIBOND SUTURE AND P-3 NEEDLES WOULD NOT STAY IN THE PATIENT. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12213076 ↗MalfunctionJun 18, 2021Break; Break
IT WAS REPORTED BY THE SALES REP THAT DURING A LATERAL CATHOPLASTY PROCEDURE ON (B)(6)2021, IT WAS OBSERVED THAT THE MICRO QUICKANCHOR PLUS PRELOADED WITH 4/0 ETHIBOND SUTURE AND P-3 NEEDLE DEVICE BROKE OFF. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12026773 ↗MalfunctionJun 18, 2021Break; Break
IT WAS REPORTED BY THE SALES REP THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2021 , IT WAS OBSERVED THAT THE MICRO QUICKANCHOR PLUS PRELOADED WITH 4/0 ETHIBOND SUTURE AND P-3 NEEDLE DEVICE BROKE OFF. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12026764 ↗MalfunctionFeb 19, 2021Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT DURING AN UNKNOWN SURGERY ON (B)(6) 2020, IT WAS OBSERVED THAT THE MICRO QA+ #4/0 ETH P-3 DEVICE WAS BLOCKED. ACCORDING TO THE REPORT, THE SURGEON COULD NOT OPEN THE DEVICE WITH THE WHITE CORD EVEN WITH A HARD FORCE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE SHAFT INSERTER ON THE DEVICE WAS BROKEN. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11347945 ↗InjuryNov 9, 2020Break; Entrapment of Device; Break; Entrapment of Device
IT WAS REPORTED BY THE AFFILIATE VIA EMAIL THAT WHEN THE MICRO QA+ #4/0 ETH P-3 ANCHOR WAS INSERTED (DURING A FOOT & ANKLE CASE), THE INSERTER CUT THROUGH THE SUTURES AND SO THE ANCHOR HAD TO BE LEFT IN THE BONE AND ANOTHER ONE WAS USED. IT WAS REPORTED IT DIDN¿T CAUSE ANY HARM TO THE PATIENT, BUT IT ADDED 10 MINUTES ONTO THE CASE. THE INSERTER CAN BE RETURNED, BUT THE ANCHOR IS IN THE PATIENT.
Patient problems: Injury; Not Applicable
MAUDE report 10814339 ↗MalfunctionDec 3, 2019Break; Break
IT WAS REPORTED BY THE SALES REP VIA EMAIL THAT THE SURGEON WAS PUTTING IN A MICRO QUICKANCHOR PLUS FOR A NASAL VALVE SUSPENSION AND THE SUTURE ON THE ANCHOR BROKE. SHE HAD TO PLACE A SECOND ANCHOR AND THAT WAS SUCCESSFUL. NO PATIENT CONSEQUENCES OR SURGICAL DELAY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9406736 ↗MalfunctionOct 23, 2019Use of Device Problem; Use of Device Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2019, IT WAS REPORTED BY THE AFFILIATE VIA CST THAT THE MICRO QA+ #4/0 ETH P-3 WAS MISSING A TIP AND COULD THEREFORE NOT BE USED. THE FACILITY REQUIRES A REPLACEMENT UNIT. NO PATIENT CONSEQUENCE. NO SURGICAL DELAY. THIS REPORT IS FOR ONE (1) MICRO QA+ #4/0 ETH P-3. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9226387 ↗MalfunctionJan 17, 2019Difficult to Insert; Difficult to Insert
IT WAS REPORTED BY THE SALES REP THAT PRIOR TO AN UNSPECIFIED SURGICAL PROCEDURE OF THE WRIST, IT WAS OBSERVED THAT THE MICRO Q/A DEVICE FELL OFF OF THE INSERTER. THEY OPENED ANOTHER LIKE DEVICE TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES OR DELAYS. THE CUSTOMER DISCARDED THE DEVICE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8258616 ↗MalfunctionDec 19, 2018Break; Break
IT WAS REPORTED BY THE CUSTOMER THAT DURING A THUMB REPAIR PROCEDURE, IT WAS OBSERVED THAT THE CUSTOMER'S MICRO QUICK ANCHOR+ WITH ETHIBOND SUTURE DEVICE BROKE WHILE INSERTING. THE CUSTOMER COULD NOT PROVIDE WHERE ON THE ANCHOR IT BROKE. THE CUSTOMER STATED THAT ALL DEBRIS WAS REMOVED FROM THE PATIENT. THE CASE WAS COMPLETED WITH ANOTHER LIKE-ANCHOR IN THE SAME BONE HOLE WITH NO PATIENT HARM OR DELAY. THE CUSTOMER STATED THAT NO SURGICAL INTERVENTION IS PLANNED. THE CUSTOMER COULD NOT PROVIDE ANY ADDITIONAL INFORMATION. THE ANCHOR IS BEING RETURNED FOR EVALUATION. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8179233 ↗
FDA clearance and classification
The FDA regulates product code HWC (screw, fixation, bone) under 21 CFR 888.3040 as a Class II device. 21 CFR 888.3040 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the QUICKANCHOR ETHIBOND 212867
Has the QUICKANCHOR ETHIBOND 212867 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886705002542), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the QUICKANCHOR ETHIBOND 212867?
The FDA's MAUDE database lists 9 adverse event reports for this device, received between Dec 19, 2018 and Jul 22, 2021. In those reports, the most common specific device problems are “Break” (5), “Difficult to Advance” (1) and “Difficult to Insert” (1); the most common patient problems are “Injury” (1) and “Not Applicable” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the QUICKANCHOR ETHIBOND 212867?
None of the reports are classified as deaths, 1 as an injury and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the QUICKANCHOR ETHIBOND 212867 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the QUICKANCHOR ETHIBOND 212867, and is it still sold?
The FDA's device database lists Depuy Mitek, LLC as the company responsible for it (brand QUICKANCHOR ETHIBOND, model 212867). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone498 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone248 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-200Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-325Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-550Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-425Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.