Medos International Sàrl · RESTORE FEMORAL AIMER, 8mm OFFSET Percent ACCURACY=Plus or Minus 5 Percent · Class 1
All models of RESTORE (33)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886705005987), received between Dec 7, 2018 and Jul 3, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fracture2
- Device-Device Incompatibility1
- Mechanical Jam1
Patient problems
- No Consequences Or Impact To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 3, 2023Device-Device Incompatibility; Mechanical Jam; Device-Device Incompatibility; Mechanical Jam
IT WAS REPORTED BY SALES REP THAT DURING AN ANTERIOR CRUCIATE LIGAMENT REPAIR PROCEDURE ON (B)(6)2023, IT WAS OBSERVED THAT THE RESTORE FEMORAL AIMER 8MM OFFSET PIN GOT STUCK WHILE PASSING GUIDE. ACCORDING TO THE REPORT, THE PHYSICIAN WAS NOT ABLE TO GET THE NEEDLE OUT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH A MINIMUM SURGICAL DELAY 2-3 MINUTES. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17247304 ↗MalfunctionDec 7, 2018Fracture; Fracture
THIS IS REPORT 2 OF 2 FOR THE SAME EVENT. IT WAS REPORTED BY THE SALES REP REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGICAL PROCEDURE, IT WAS OBSERVED THAT WHEN DRILLING THE DISPOSABLE ACL BAYPINS W/EYE .094X18", ONE OF THE PINS BROKE OFF IN THEIR FEMORAL AIMER 8MM OFFSET IN THE PATIENT. THE SALES REP STATED THAT THEY ARE UNABLE TO REMOVE THE BROKEN PIN FROM THE DEVICE. THE SALES REP STATED THAT THERE WAS NO DEBRIS LEFT IN THE PATIENT. THE SALES REP REPORTED THAT THE SURGEON COMPLETED THE PROCEDURE BY DOWNSIZING TO A DIFFERENT SIZE AND DRILL. THE SALES REP REPORTED THAT THERE WERE NO PATIENT CONSEQUENCES BUT THERE WAS A FIVE MINUTE DELAY IN THE CASE. THE SALES REP COULD NOT PROVIDE A LOT NUMBER FOR THE DEVICE BUT STATED THAT THE DEVICE IS AN OLDER DEVICE. THE DEVICE WILL BE RETURNING FOR EVALUATION. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8142492 ↗MalfunctionDec 7, 2018Fracture; Fracture
THIS IS REPORT 1 OF 2 FOR THE SAME EVENT. IT WAS REPORTED BY THE SALES REP THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGICAL PROCEDURE, IT WAS OBSERVED THAT WHEN DRILLING THE DISPOSABLE ACL BAYPINS W/EYE .094X18", ONE OF THE PINS BROKE OFF IN THEIR FEMORAL AIMER 8MM OFFSET IN THE PATIENT. THE SALES REP STATED THAT THEY ARE UNABLE TO REMOVE THE BROKEN PIN FROM THE DEVICE. THE SALES REP STATED THAT THERE WAS NO DEBRIS LEFT IN THE PATIENT. THE SALES REP REPORTED THAT THE SURGEON COMPLETED THE PROCEDURE BY DOWNSIZING TO A DIFFERENT SIZE AND DRILL. THE SALES REP REPORTED THAT THERE WERE NO PATIENT CONSEQUENCES BUT THERE WAS A FIVE MINUTE DELAY IN THE CASE. THE SALES REP COULD NOT PROVIDE A LOT NUMBER FOR THE DEVICE BUT STATED THAT THE DEVICE IS AN OLDER DEVICE. THE DEVICE WILL BE RETURNING FOR EVALUATION. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8142038 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.