Skip to content

NA 219962Recalls & reported problems

Medos International Sàrl · Slider Unit · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
17
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 10886705007042), received between Jul 24, 2020 and Jun 25, 2024 — classified as 17 malfunctions.

The most frequently reported specific device problem is “Degraded” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded8
  • Corroded6
  • Unintended Movement2
  • Mechanical Jam1
  • Use of Device Problem1

Patient problems

  • No Consequences Or Impact To Patient11
  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 25, 2024Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT DURING INSPECTION, IT WAS OBSERVED THAT THE GRAFT PREP SYS SLIDERUNIT DEVICE WAS MISSING A SCREW TO KEEP IT TOGETHER. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19606736 ↗
  • MalfunctionAug 7, 2023Mechanical Jam; Mechanical Jam

    THIS IS REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED BY SALES REP VIA EMAIL THAT DURING ANTERIOR CRUCIATE LIGAMENT REPAIR GRAFT PREP BOARD AND SLIDER UNIT DID NOT SLIDE ONTO THE TRACK WHILE IN USE TOGETHER. ACCORDING TO THE REPORT, BOTH SLIDER UNITS FROM BOTH ENDS WERE TRIED. IT WAS FURTHER REPORTED THAT THEY USED THE GRAFT GRASPERS IN THE STATIONARY HOLES ON THE BLUE FRAME OF THE BOARD TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17471341 ↗
  • MalfunctionAug 3, 2022Unintended Movement; Unintended Movement

    THIS IS REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED BY THE SALES REP THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT TWO OF THE SLIDER UNIT DEVICES WERE NOT HOLDING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15156760 ↗
  • MalfunctionAug 3, 2022Unintended Movement; Unintended Movement

    THIS IS REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED BY THE SALES REP THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT TWO OF THE SLIDER UNIT DEVICES WERE NOT HOLDING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15156757 ↗
  • MalfunctionDec 29, 2020Degraded; Degraded

    IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT DURING INSPECTION ON (B)(6) 2020, IT WAS OBSERVED THAT THE GRAFT PREP SYSTEM SLIDER UNIT DEVICE WAS CORRODED. THERE WAS NO PROCEDURE INVOLVED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11089221 ↗
  • MalfunctionDec 29, 2020Degraded; Degraded

    IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT DURING INSPECTION ON 12/15/2020, IT WAS OBSERVED THAT THE GRAFT PREP SYSTEM SLIDER UNIT DEVICE WAS CORRODED. THERE WAS NO PROCEDURE INVOLVED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11089211 ↗
  • MalfunctionSep 14, 2020Degraded; Degraded

    IT WAS REPORTED BY THE AFFILIATE VIA CST THAT DURING INSPECTION OF THE ORTHOKIT THE FOLLOWING DEVICES WERE RUSTED: TIBIAL GUIDE RAIL, TIBIAL GUIDE ARM-TIP AIMER, TIBIAL GUIDE BULLET, CRUCIATE + FEMORAL CAPTURE, GRAFT PREP SYS SLIDERUNIT (2 REQ) *EA. ALL DEVICES ARE AVAILABLE TO BE RETURNED FOR EVALUATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10527874 ↗
  • MalfunctionSep 14, 2020Degraded; Degraded

    IT WAS REPORTED BY THE AFFILIATE VIA CST THAT DURING INSPECTION OF THE ORTHOKIT THE FOLLOWING DEVICES WERE RUSTED: TIBIAL GUIDE RAIL, TIBIAL GUIDE ARM-TIP AIMER, TIBIAL GUIDE BULLET, CRUCIATE + FEMORAL CAPTURE, GRAFT PREP SYS SLIDERUNIT (2 REQ) *EA. ALL DEVICES ARE AVAILABLE TO BE RETURNED FOR EVALUATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10527873 ↗
  • MalfunctionSep 14, 2020Degraded; Degraded

    IT WAS REPORTED BY THE AFFILIATE VIA CST THAT DURING INSPECTION OF THE ORTHOKIT THE FOLLOWING DEVICES WERE RUSTED: TIBIAL GUIDE RAIL, TIBIAL GUIDE ARM-TIP AIMER, TIBIAL GUIDE BULLET, CRUCIATE + FEMORAL CAPTURE, GRAFT PREP SYS SLIDERUNIT (2 REQ) *EA. ALL DEVICES ARE AVAILABLE TO BE RETURNED FOR EVALUATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10527872 ↗
  • MalfunctionSep 14, 2020Degraded; Degraded

    IT WAS REPORTED BY THE AFFILIATE VIA CST THAT DURING INSPECTION OF THE ORTHOKIT THE FOLLOWING DEVICES WERE RUSTED: TIBIAL GUIDE RAIL, TIBIAL GUIDE ARM-TIP AIMER, TIBIAL GUIDE BULLET, CRUCIATE + FEMORAL CAPTURE, GRAFT PREP SYS SLIDER UNIT (2 REQ) EA. ALL DEVICES ARE AVAILABLE TO BE RETURNED FOR EVALUATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10527868 ↗
  • MalfunctionSep 14, 2020Degraded; Degraded

    IT WAS REPORTED BY THE AFFILIATE VIA CST THAT DURING INSPECTION OF THE ORTHOKIT THE FOLLOWING DEVICES WERE RUSTED: TIBIAL GUIDE RAIL, TIBIAL GUIDE ARM-TIP AIMER, TIBIAL GUIDE BULLET, CRUCIATE + FEMORAL CAPTURE, GRAFT PREP SYS SLIDER UNIT (2 REQ) EA. ALL DEVICES ARE AVAILABLE TO BE RETURNED FOR EVALUATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10527866 ↗
  • MalfunctionSep 11, 2020Corroded; Corroded

    IT WAS REPORTED BY THE AFFILIATE VIA CST THAT DURING INSPECTION OF THE ORTHOKIT THE FOLLOWING DEVICES WERE RUSTED: TIBIAL GUIDE RAIL, TIBIAL GUIDE ARM-TIP AIMER, TIBIAL GUIDE BULLET, CRUCIATE + FEMORAL CAPTURE, GRAFT PREP SYS SLIDERUNIT (2 REQ) EA. ALL DEVICES ARE AVAILABLE TO BE RETURNED FOR EVALUATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10523626 ↗
  • MalfunctionSep 11, 2020Corroded; Corroded

    IT WAS REPORTED BY THE AFFILIATE VIA COMPLAINT SUBMISSION TOOL THAT WHILE INSPECTING THE ORTHOKIT, IT WAS FOUND THAT THE ARTICLES IN THE LENDING SET WERE RUSTING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10523379 ↗
  • MalfunctionSep 11, 2020Corroded; Corroded

    IT WAS REPORTED BY THE AFFILIATE VIA COMPLAINT SUBMISSION TOOL THAT WHILE INSPECTING THE ORTHOKIT, IT WAS FOUND THAT THE ARTICLES IN THE LENDING SET WERE RUSTING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10523369 ↗
  • MalfunctionSep 11, 2020Corroded; Corroded

    IT WAS REPORTED BY THE AFFILIATE VIA COMPLAINT SUBMISSION TOOL THAT WHILE INSPECTING THE ORTHOKIT, IT WAS FOUND THAT THE ARTICLES IN THE LENDING SET WERE RUSTING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10522692 ↗
  • MalfunctionSep 11, 2020Corroded; Corroded

    IT WAS REPORTED BY THE AFFILIATE VIA COMPLAINT SUBMISSION TOOL THAT WHILE INSPECTING THE ORTHOKIT, IT WAS FOUND THAT THE ARTICLES IN THE LENDING SET WERE RUSTING.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10522480 ↗
  • MalfunctionJul 24, 2020Corroded; Degraded; Corroded; Degraded

    THIS IS REPORT 2 OF 5 FOR THE SAME EVENT. IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT DURING STERILIZATION PRIOR TO AN UNKNOWN SURGERY ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE GRAFT PREP SYS SLIDERUNIT (2 REQ) EA DEVICE HAD SOME RUST SPOTS PARTICULARLY ON THE WRITINGS (LETTERS). IT WAS NOT REPORTED IF THERE WAS A DELAY IN THE SURGICAL PROCEDURE. IT WAS NOT REPORTED IF A SPARE DEVICE WAS AVAILABLE FOR USE. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10319585 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →