MILAGRO 231809Recalls & reported problems
Medos International Sàrl · MILAGRO INTERFERENCE SCREW Absorbable - TCP/PLGA 6 x 30mm · Class 2 · Implant
All models of MILAGRO (26)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10886705010769), received between Jul 25, 2018 and May 13, 2026 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
5 other recalls cover devices cleared under the same FDA submission (K103831, K150209); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break6
- Difficult to Advance1
- Fracture1
- Migration or Expulsion of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient3
- No Code Available1
- Tissue Damage1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 13, 2026Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, WHEN USING THE GRYPHON ORTHOCORD, THE ANCHOR BROKE OFF. ALL THE BROKEN PARTS WERE REMOVED. CHANGED ANOTHER THE SAME PRODUCT CODE OF THE GRYPHON ORTHOCORD ANCHOR TO CONTINUE THE SURGERY. AFTERWARDS, WHEN THE MILAGRO SCREW WAS INSERTED, THE SCREW BROKE. ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER THE SAME PRODUCT CODE OF THE MILAGRO SCREW WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25162688 ↗MalfunctionAug 1, 2025Difficult to Advance; Difficult to Advance
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE MILAGRO INSCR SMALL SIZE 6X30 SCREW DEVICE SCREW SLIPPED OFF AND COULD NOT BE INSERTED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22689958 ↗MalfunctionJan 3, 2025Break; Break
IT WAS REPORTED THAT DURING A TENDON REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE MILAGRO INSCR SMALL SIZE 6X30 ANCHOR DEVICE BROKE OFF. IT WAS REPORTED THAT ALL BROKEN PARTS WERE REMOVED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21072630 ↗MalfunctionAug 9, 2023Break; Break
IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING A PATELLAR LIGAMENT RECONSTRUCTION PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE 6MM X 30MM MILAGRO SCREW DEVICE WAS BROKEN OFF THEN REMOVED ALL THE BROKEN PARTS. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17490634 ↗MalfunctionAug 3, 2021Break; Break
IT WAS REPORTED BY THE CUSTOMER IN (B)(4) THAT DURING AN ARTHROSCOPIC PATELLAR EXPLORATION + INCISION ANCHOR RECONSTRUCTION PROCEDURE ON (B)(6) 2021, IT WAS OBSERVED THAT THE MILAGRO INSCR SMALL SIZE 6X30 DEVICE BROKE. ALL BROKEN PARTS WERE REMOVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12269374 ↗InjuryMar 27, 2020Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED BY AFFILIATE VIA COMPLAINT SUBMISSION TOOL THAT A MONTH AFTER PROCEDURE THE MILAGRO INSCR SMALL SIZE 6X30 SCREW MOVED UPWARDS INTO THE JOINT SPACE, AS A CONSEQUENCE THE PATIENT HAD STRANGE FEELING IN THE POPLITEUS, A SURGERY WAS PERFORMED AND THE SURGEON WAS NOT ABLE TO REMOVE THE SCREW. THE DEVICE IS NOT AVAILABLE TO BE RETURNED FOR EVALUATION. ADDITIONAL INFORMATION PROVIDED BY THE AFFILIATE REPORTED THE SURGEON REPORTED THE PATIENT FEELS SOMETHING DISTURBING TO HIM NEAR HIS POPLITEUS. IT WAS REPORTED AN X-RAY IS SCHEDULED TO OCCUR SOME TIME SOON BUT THE EXACT TIME IS UNKNOWN. THE AFFILIATE REPORTED IT WAS UNKNOWN IF THE SCREW HAS CAUSED ANY ADDITION HARM OR WHAT THE PATIENT'S PROGNOSIS MAYBE. IT WAS ALSO REPORTED THE LOT NUMBER IS UNKNOWN.
Patient problems: No Code Available; Tissue Damage
MAUDE report 9890985 ↗MalfunctionFeb 7, 2019Break; Break
PHYSICIAN REPORTED THAT SCREW IS BROKEN. NO ADDITIONAL DETAILS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8317418 ↗MalfunctionJan 30, 2019Fracture; Fracture
IT WAS REPORTED THAT DURING MPFL RECONSTRUCTION THE SURGEON USED THE DEVICE FOR GRAFT FIXATION OF THE FEMUR. IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE WHEN HE INSERTED THE DEVICE IN A 6MM BONE HOLE AFTER PASSING A GRAFT THROUGH THE BONE HOLE AND INSERTING A GUIDE PIN. HE STOPPED USING THE DEVICE AND SUCCESSFULLY INSERTED A SHORTER SIZE OF THE DEVICE (6X23MM). IT WAS ALSO REPORTED THAT NO BROKEN PIECE WAS LEFT IN THE PATIENT¿S BODY. THERE WAS NO SURGICAL DELAY OR HARM TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8288434 ↗MalfunctionJul 25, 2018Break; Break
IT WAS REPORTED THAT THE SCREWS IN QUESTION WERE USED FOR OUTSIDE-IN ACL RECONSTRUCTION USING BTB GRAFT. DURING INSERTION OF MILAGRO BR 6 X 30 MM INTO THE FEMORAL SIDE, IT COULD NOT BE INSERTED FURTHER IN THE MIDDLE OF THE BONE HOLE. WHEN THE SURGEON REVERSED AND TRIED PULLED IT OUT, THE TIP OF THE SCREW (3-5 MM) WAS BROKEN AND THE BROKEN PIECE GOT STUCK BETWEEN THE BONE HOLE AND THE GRAFT. THE SURGEON REMOVED THE BROKEN PIECE USING AN OSTEOTOME AND RASPATORY, AND FIXED THE GRAFT WITH THE COMPETITOR¿S SCREW (6 X 30 MM) INSTEAD. FOR THE TIBIAL SIDE, MILAGRO ADVANCE 8 X 23 MM WAS USED, HOWEVER, THE DESIRED TENSION WAS NOT OBTAINED BY THE SCREW. THE SURGEON REMOVED THE SCREW AND RE-FIXED WITH THE COMPETITOR¿S DSP. THE PROCEDURE WAS COMPLETED WITHOUT ANY SURGICAL DELAY OR HARM TO THE PATIENT BECAUSE OF THESE PROBLEMS. THE BACKUP DEVICE WAS USED TO COMPLETE THE CASE. PLEASE SEE ALSO MEDWATCH REPORT 1221934-2018-52390.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7720195 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K103831, K150209. One submission often covers many models, so check the product and lot details in each record.
- Z-1680-2023 · Apr 13, 2023 · Open, ClassifiedDePuy Mitek BIOKNOTLESS Rapid W/ OC-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212722
- Z-1679-2023 · Apr 13, 2023 · Open, ClassifiedDePuy Mitek BIOKNOTLESS Plus w/ OC-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212726
- Z-1681-2023 · Apr 13, 2023 · Open, ClassifiedDePuy Mitek BIOKNOTLESS Rapid w/ Pan-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212723
- Z-1677-2023 · Apr 13, 2023 · Open, ClassifiedDePuy Mitek BIOKNOTLESS Plus w/ Ethibond-Iindicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212724
- Z-1678-2023 · Apr 13, 2023 · Open, ClassifiedDePuy Mitek BIOKNOTLESS Plus w/ Panacryl-Indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization Model Number: 212725
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone503 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone253 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-400Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-225Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-350Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-300Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page