MILAGRO 231825Recalls & reported problems
Depuy Mitek, LLC · MILAGRO INTERFERENCE SCREW Absorbable - TCP/PLGA 9 x 30mm · Class 2 · Implant
All models of MILAGRO (10)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10886705010806), received between Jul 27, 2018 and Sep 3, 2024 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
- Unspecified Infection1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjurySep 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY THE SALES REP IN INDIA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) SURGICAL PROCEDURE ON (B)(6) 2024 WHILE USING THE CORTICAL FXATION FXD LOOP 20MM DEVICE AND THE MILAGRO INTSCR 9X30MM *EA ON FEMUR SIDE THE SURGERY WENT WELL AND THE PATIENT WAS DISCHARGED. AFTER TWO TO THREE MONTHS THE PATIENT SHOWED SIGN ON INFECTION AT THE OPERATED SITE AND TOOK ADVICE FROM OTHER SURGEON, WHO AFTER CHECKING THE PATIENT REPORTED AND IMPLANTS COMMENTED TO THE PATIENT ABOUT SUB STANDARD IMPLANTS USED. THE PATIENT HAS FILED COMPLIANT AGAINST OPERATIVE SURGEON DOCTOR OF NORTH STAR MULTISPECIALTY HOSPITAL KOLHAPUR. HOSPITAL HAS REQUESTED FROM COMPANY A QUALITY CERTIFICATE TO FURTHER ENDORSED CLINICALLY PROVEN IMPLANTS WERE USED IN THE SURGERY. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 2 OF 2 OF THE SAME EVENTS.
Patient problems: Unspecified Infection
MAUDE report 20124690 ↗MalfunctionJun 25, 2024Break; Break
IT WAS REPORTED FROM CHINA THAT DURING A POSTERIOR CRUCIATE LIGAMENT (PCL) RECONSTRUCTION PROCEDURE, IT WAS DISCOVERED THAT THE ANCHOR ON THE MILAGRO INTSCR 9X30MM DEVICE WAS BROKEN OFF. ALL THE BROKEN PARTS WERE REMOVED. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19605601 ↗MalfunctionMar 6, 2024Break; Break
IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN ACL (ANT.CRUCIATE LIGAMENT) REPAIR SURGERY ON (B)(6) 2024, IT WAS OBSERVED THAT THE MILAGRO INTERFERENCE SCREW, 9 X 30MM DEVICE ANCHOR WAS BROKEN OFF, REMOVED ALL THE BROKEN PARTS. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18844604 ↗MalfunctionAug 27, 2018Break; Break
IT WAS REPORTED THAT AS THE SURGEON WAS SCREWING THE SCREW INTO THE TIBIA, THE TIP OF THE IMPLANT BROKE. REPORT ONLY, DEVICE WAS DISCARDED BY THE CUSTOMER. THE PROCEDURE WAS COMPLETED WITH SAME LIKE DEVICE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7819950 ↗MalfunctionJul 27, 2018Break; Break
IT WAS REPORTED BY THE AFFILIATE IN VIETNAM THAT DURING A KIDNEY SURGERY, IT WAS OBSERVED THAT THE MILAGRO INTERFERENCE SCREW, 9 X 30MM WAS BROKEN WHEN THE SURGEON SCREWED IT INTO THE TIBIAL TUNNEL. THE SURGEON HAD TO USE ANOTHER SCREW TO KEEP THE GAFT STABILITY INTO TUNNEL. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE PROCEDURE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7725807 ↗
FDA clearance and classification
The FDA regulates product code HWC (screw, fixation, bone) under 21 CFR 888.3040 as a Class II device. 21 CFR 888.3040 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the MILAGRO 231825
Has the MILAGRO 231825 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886705010806), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the MILAGRO 231825?
The FDA's MAUDE database lists 5 adverse event reports for this device, received between Jul 27, 2018 and Sep 3, 2024. In those reports, the most common specific device problem is “Break” (4); the most common patient problem is “Unspecified Infection” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the MILAGRO 231825?
None of the reports are classified as deaths, 1 as an injury and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the MILAGRO 231825 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the MILAGRO 231825, and is it still sold?
The FDA's device database lists Depuy Mitek, LLC as the company responsible for it (brand MILAGRO, model 231825). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Screw, fixation, bone devices
- NA 03.043.029Synthes GmbH · Screw, fixation, bone498 reports
- OSTEORAPTOR 72201995Smith & Nephew, Inc. · Screw, fixation, bone248 reports
- Control 360 SK12Treace Medical Concepts, Inc. · Screw, Fixation, Bone195 reports
- HEALIX PEEK ORTHOCORD 222205Medos International Sàrl · Screw, fixation, bone186 reports
- PRECICE® Bone Transport System STP3.5-200Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-325Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-550Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
- PRECICE® Bone Transport System STP3.5-425Nuvasive Specialized Orthopedics, Inc. · Screw, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page