BIOINTRAFIX 254625Recalls & reported problems

Medos International Sàrl · BIO-INTRAFIX Tapered Screw 8-10mm x 30mm · Class 2 · Implant

All models of BIOINTRAFIX (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10886705013227), received between Jun 29, 2018 and Jan 5, 2024 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Crack3
  • Failure to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Consequences Or Impact To Patient3
  • Foreign Body In Patient1
  • No Code Available1
  • Not Applicable1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 5, 2024Break; Break

    IT WAS REPORTED BY THE SALES REP IN CHINA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT REPAIR SURGERY ON (B)(6) 2023 THE BIOINTRAFIX 8-10MMX30MM TPRSCR DEVICE ANCHOR WAS BROKEN OFF. CHANGED ANOTHER BIOINTRAFIX 8-10MMX30MM TPRSCR DEVICE AND TWO MILAGRO ADVANCE SCREW 10X30MM DEVICES TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 1 OF 4 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18452209 ↗
  • MalfunctionJan 5, 2024Break; Break

    IT WAS REPORTED BY THE SALES REP IN CHINA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT REPAIR SURGERY ON (B)(6) 2023 THE BIOINTRAFIX 8-10MMX30MM TPRSCR DEVICE ANCHOR WAS BROKEN OFF. CHANGED ANOTHER BIOINTRAFIX 8-10MMX30MM TPRSCR DEVICE AND TWO MILAGRO ADVANCE SCREW 10X30MM DEVICES TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 2 OF 4 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18451830 ↗
  • MalfunctionJun 15, 2023Break; Break

    IT WAS REPORTED BY THE SALES REP THAT THE BIO-INTRAFIX TAPERED SCREW, 8-10 MM X 30 MM DEVICE BROKE. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17139540 ↗
  • MalfunctionJun 16, 2022Crack; Crack

    IT WAS REPORTED BY THE CUSTOMER IN CHINA THAT DURING AN ACL (ANT.CRUCIATE LIGAMENT) RECONSTRUCTION PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE BIO-INTRAFIX TAPERED SCREW, 8-10 MM X 30 MM ANCHOR DEVICE HAD A CRACK ON IT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14711210 ↗
  • InjuryOct 8, 2019Failure to Advance; Failure to Advance

    IT WAS REPORTED BY THE AFFILIATE VIA COMPLAINT SUBMISSION TOOL THAT DURING AN UNKNOWN PROCEDURE THE BIOINTRAFIX 8-10MMX30MM TPRSCR WHEN IT WAS BEEN INSERTED INTO THE SHEATH IN THE TUNNEL. THE INTERNAL HEX THREAD WAS LOST BEFORE THE SCREW WAS SUNK TO THE CORTICES (ABOUT 3MM PROTRUDING BEYOND BONE SURFACE). SCREW COULD THEREFORE NOT BE ADVANCED OR REMOVED. EXPOSED TIP OF SCREW WAS OSTEOTOMED OFF AT THE LEVEL OF THE BONE. A STAPLE WAS ALSO USED IN THE TIBIA TO ENSURE A SECURE FIXATION OF THE GRAFT. BEFORE IT WAS SUNK INTO THE CORTICES. THEREFORE IT COULDN¿T GO FURTHER OR BE REMOVED AND IT HAD TO BE LOWERED OFF AT THE LEVEL OF THE BONE. A STAPLE HAD TO BE USED IN THE TIBIA TO ENSURE AND SECURE THE FIXATION OF THE GRAFT. A SURGICAL DELAYED OF 2 MINUTES WAS REPORTED. NO ADDITIONAL INFORMATION WAS REPORTED.

    Patient problems: Foreign Body In Patient; No Code Available; Not Applicable

    MAUDE report 9169156 ↗
  • MalfunctionJan 28, 2019Break; Break

    IT WAS REPORTED BY THE SALES REP THAT DURING AN ANTERIOR CRUCIATE LIGAMENT ACL REPAIR SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE CUSTOMER'S BIOINTRAFIX TAPERED SCREW, 8-10 X 30MM BROKE UPON INSERTION INTO VERY HARD BONE. THE SALES REP REPORTED THAT THE IMPLANT CRACKED BUT DID NOT BREAK INTO SEPARATE PIECES AND WAS FULLY REMOVED FROM THE JOINT SPACE. THE SURGEON COMPLETED THE PROCEDURE WITH ANOTHER LIKE DEVICE IN THE SAME BONE HOLE WITH NO PATIENT CONSEQUENCES OR DELAYS. THE SALES REP REPORTED THAT THE SURGEON WAS OFF AXIS. THE CUSTOMER DISCARDED THE COMPLAINT DEVICE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8282651 ↗
  • MalfunctionJun 29, 2018Crack; Crack

    THE SALES REP REPORTED THAT DURING AN ACL PROCEDURE THAT TWO OF HIS BIOINTRAFIX TAPERED SCREWS AND LARGE SHEATHS BROKE UPON INSERTION. THE SALES REP STATED THAT THE SURGEON HAD PREDRILLED THE HOLE AND DID NOT TAP THE DEVICES IN. THE SALES REP REPORTED THAT THE BONE QUALITY WAS VERY HARD AND HE BELIEVED THE SURGEON MAY HAVE BEEN OFF AXIS. THE SURGEON COMPLETELY REMOVED BOTH OF THE SCREWS AND SHEATHS AND COMPLETED THE PROCEDURE WITH A THIRD SIMILAR SET OF DEVICES WITH NO PATIENT CONSEQUENCES OR DELAY. THE DEVICES WERE ALREADY DISCARDED. PLEASE SEE ALSO MEDWATCH REPORT 1221934-2018-51030, 1221934-2018-51031, AND 1221934-2018-51034

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7652973 ↗
  • MalfunctionJun 29, 2018Crack; Crack

    THE SALES REP REPORTED THAT DURING AN ACL PROCEDURE THAT TWO OF HIS BIOINTRAFIX TAPERED SCREWS AND LARGE SHEATHS BROKE UPON INSERTION. THE SALES REP STATED THAT THE SURGEON HAD PREDRILLED THE HOLE AND DID NOT TAP THE DEVICES IN. THE SALES REP REPORTED THAT THE BONE QUALITY WAS VERY HARD AND HE BELIEVED THE SURGEON MAY HAVE BEEN OFF AXIS. THE SURGEON COMPLETELY REMOVED BOTH OF THE SCREWS AND SHEATHS AND COMPLETED THE PROCEDURE WITH A THIRD SIMILAR SET OF DEVICES WITH NO PATIENT CONSEQUENCES OR DELAY. THE DEVICES WERE ALREADY DISCARDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7652935 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →