Medos International Sàrl · BIO-INTRAFIX Tibial Sheath, Large · Class 2 · Implant

All models of BIOINTRAFIX (6)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This BIOINTRAFIX tibial sheath is a bone fixation screw used in ACL knee surgery. Reports mostly describe the sheath breaking or cracking during the procedure, with broken parts removed and surgery completed using another device. There are 91 adverse event reports naming it, with no deaths and 5 injuries. No recalls name this device. Reports fell from 5 in the prior 12 months to none in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
91
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 91 adverse event reports that list this exact device (DI 10886705013241), received between Jun 18, 2018 and Sep 19, 2025 — classified as 5 injuries, 86 malfunctions.

The most frequently reported specific device problem is “Break” (69 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break69
  • Crack9
  • Fracture6
  • Adverse Event Without Identified Device or Use Problem5
  • Device Contaminated During Manufacture or Shipping1
  • Loose or Intermittent Connection1
  • Material Split, Cut or Torn1

Patient problems

  • No Consequences Or Impact To Patient53
  • No Clinical Signs, Symptoms or Conditions27
  • No Known Impact Or Consequence To Patient4
  • Unspecified Infection3
  • Inflammation2
  • Not Applicable2
  • Foreign Body In Patient1
  • Foreign Body Reaction1

Recent adverse event reports

The 25 most recent of 91 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 19, 2025Break; Break

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) SURGICAL PROCEDURE IT WAS OBSERVED THAT THE BIOINTRAFIX TBIAL SH LRG 30MM SHEATH DEVICE WAS BROKEN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23094351 ↗
  • MalfunctionJul 7, 2025Break; Break

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT REPAIR (ACL) SURGICAL PROCEDURE IT WAS OBSERVED THAT WHILE USING THE BIOINTRAFIX TBIAL SH LRG 30MM DEVICE AND THE MILAGRO INSCR SMALL SIZE 6X23 DEVICE ANCHOR WAS BROKEN OFF. ALL BROKEN PARTS WERE REMOVED. THE USER CHANGED TO ANOTHER DEVICE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 1 OF 2 OF (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22416241 ↗
  • MalfunctionMar 26, 2025Break; Break

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) REPAIR PROCEDURE, THE BIOINTRAFIX TBIAL SH LRG 30MM SHEATH WAS BROKEN OFF. IT WAS REPORTED THAT ALL BROKEN PARTS WERE REMOVED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21700108 ↗
  • MalfunctionFeb 20, 2025Break; Break

    IT WAS REPORTED THAT DURING A CRUCIATE LIGAMENT RECONSTRUCTION SURGICAL PROCEDURE THE BIOINTRAFIX TBIAL DEVICE ANCHOR WAS BROKEN OFF. AFTER IMPLANTATION THE ANCHOR SHEATH DEHISCENCE. ALL BROKEN PARTS WERE REMOVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21425849 ↗
  • InjuryJan 9, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING ANTERIOR CRUCIATE LIGAMENT REPAIR SURGERY, THE PATIENT EXPERIENCED POST-OPERATIVE INFLAMMATION. IT WAS REPORTED THAT THE PATIENT WAS ADMITTED TO XI'AN RED CROSS HOSPITAL 216 DAYS AGO DUE TO "LEFT KNEE JOINT PAIN AND DISCOMFORT CAUSED BY SPRAIN FOR ONE MONTH". ACCORDING TO THE REPORTER, UNDER GENERAL ANESTHESIA COMBINED WITH NERVE BLOCK, THE PATIENT UNDERWENT "LEFT KNEE ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION SURGERY AND ARTHROSCOPIC MENISCUS SHAPING SURGERY AND ARTHROSCOPIC KNEE SYNOVECTOMY AND TENDON EXCISION SURGERY AND JOINT THERAPEUTIC SUBSTANCE INJECTION". IT WAS REPORTED THAT THE ANTERIOR CRUCIATE LIGAMENT REPAIR SYSTEM WAS USED DURING THE SURGERY, AND THE SURGERY WENT SMOOTHLY HOWEVER POSTOPERATIVE CT SCAN SHOWED FIRM INTERNAL FIXATION. ACCORDING TO THE REPORTER, AFTER SIX WEEKS OF SURGERY, THE PATIENT EXPERIENCED POOR WOUND HEALING AND WAS ADMITTED TO THE HOSPITAL 166 DAYS AGO WITH THE COMPLAINT OF "POOR WOUND HEALING AFTER 6 WEEKS OF LEFT ANTERIOR CRUCIATE LIGAMENT REVISION SURGERY AND ONE WEEK OF POOR WOUND HEALING". IT WAS REPORTED THAT UNDER GENERAL ANESTHESIA, THE PATIENT UNDERWENT "DEBRIDEMENT AND VSD DRAINAGE SURGERY FOR POOR WOUND HE

    Patient problems: Inflammation; Unspecified Infection

    MAUDE report 21110370 ↗
  • MalfunctionFeb 8, 2024Break; Break

    IT WAS REPORTED BY THE SALES REP IN CHINA THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE ON (B)(6) 2024 THE BIOINTRAFIX TBIAL DEVICE WAS BROKEN. ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18672394 ↗
  • MalfunctionFeb 8, 2024Break; Break

    IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT REPAIR PROCEDURE ON (B)(6)2024, IT WAS OBSERVED THAT THE BIO-INTRAFIX TIBIAL SHEATH - LARGE DEVICE SHEATH WAS BROKEN OFF. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18666327 ↗
  • MalfunctionJan 31, 2024Break; Break

    IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE ON (B)(6) 2024, IT WAS OBSERVED THAT THE SHEATH ON THE BIOINTRAFIX TBIAL SH LRG 30MM DEVICE WAS BROKEN OFF, THEN REMOVED ALL THE BROKEN PARTS. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18618911 ↗
  • MalfunctionDec 29, 2023Break; Break

    IT WAS REPORTED BY THE SALES REP IN CHINA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) REPAIR PROCEDURE PERFORMED ON (B)(6) 2023, WHEN THE PHYSICIAN IMPLANTED THE BIOINTRAFIX "TIBIAL" SH LRG 30MM SHEATH INTO THE BODY, THE SHEATH BROKE INTO SEVERAL PIECES. THE SURGEON CHANGED TO ANOTHER BIOINTRAFIX "TIBIAL" SH LRG 30MM SHEATH AND THE SAME PROBLEM OCCURRED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 2 OF 2 FOR THE SAME EVENT IN (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18418415 ↗
  • MalfunctionDec 29, 2023Break; Break

    IT WAS REPORTED BY THE SALES REP IN CHINA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) REPAIR PROCEDURE PERFORMED ON (B)(6) 2023, WHEN THE PHYSICIAN IMPLANTED THE BIOINTRAFIX "TIBIAL" SH LRG 30MM SHEATH INTO THE BODY, THE SHEATH BROKE INTO SEVERAL PIECES. THE SURGEON CHANGED TO ANOTHER BIOINTRAFIX "TIBIAL" SH LRG 30MM SHEATH AND THE SAME PROBLEM OCCURRED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 1 OF 2 FOR THE SAME EVENT IN (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18418383 ↗
  • MalfunctionNov 15, 2023Break; Break

    IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE BIOINTRAFIX TBIAL SH LRG 30MM DEVICE WAS BROKEN OFF, THEN REMOVED ALL THE BROKEN PARTS. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18143077 ↗
  • MalfunctionNov 15, 2023Break; Break

    IT WAS REPORTED BY THE SALES REP IN CHINA THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE ON (B)(6) 2023 THE BIOINTRAFIX TBIAL SH LRG 30MM DEVICE SHEATH WAS BROKEN OFF. ALL THE BROKEN PARTS WERE REMOVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18142725 ↗
  • MalfunctionNov 15, 2023Break; Break

    THIS IS REPORT 2 OF 3 FOR (B)(4). IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE BIOINTRAFIX TBIAL SH LRG 30MM DEVICE WAS BROKEN OFF, THEN REMOVED ALL THE BROKEN PARTS. CHANGED TO ANOTHER TWO TO CONTINUE THE SURGERY BUT THE SAME PROBLEM HAPPENED AGAIN. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18140870 ↗
  • MalfunctionNov 15, 2023Break; Break

    THIS IS REPORT 1 OF 3 FOR (B)(4). IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE BIOINTRAFIX TBIAL SH LRG 30MM DEVICE WAS BROKEN OFF, THEN REMOVED ALL THE BROKEN PARTS. CHANGED TO ANOTHER TWO TO CONTINUE THE SURGERY BUT THE SAME PROBLEM HAPPENED AGAIN. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18140864 ↗
  • MalfunctionNov 15, 2023Break; Break

    THIS IS REPORT 3 OF 3 FOR PC-(B)(4). IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE ON 10/1/2023, IT WAS OBSERVED THAT THE BIOINTRAFIX TBIAL SH LRG 30MM DEVICE WAS BROKEN OFF, THEN REMOVED ALL THE BROKEN PARTS. CHANGED TO ANOTHER TWO TO CONTINUE THE SURGERY BUT THE SAME PROBLEM HAPPENED AGAIN. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18140846 ↗
  • MalfunctionAug 31, 2023Break; Break

    IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE SHEATH ON THE BIO-INTRAFIX TIBIAL SHEATH - LARGE DEVICE WAS BROKEN OFF UPON INSERTION OF THE ANCHOR, THEN REMOVED ALL THE BROKEN PARTS. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17664474 ↗
  • MalfunctionAug 30, 2023Break; Break

    IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE SHEATH ON THE BIO-INTRAFIX TIBIAL SHEATH - LARGE DEVICE WAS BROKEN OFF, THEN REMOVED ALL THE BROKEN PARTS. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17651391 ↗
  • MalfunctionAug 28, 2023Break; Break

    IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE BIO-INTRAFIX TIBIAL SHEATH - LARGE DEVICE WAS BROKEN OFF, THEN REMOVED ALL THE BROKEN PARTS. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17631958 ↗
  • MalfunctionAug 11, 2023Break; Break

    IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING A ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE BIO-INTRAFIX TIBIAL SHEATH - LARGE DEVICE WAS BROKEN OFF, THEN REMOVED ALL THE BROKEN PARTS. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17510117 ↗
  • MalfunctionJul 25, 2023Break; Break

    IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE BIO-INTRAFIX TIBIAL SHEATH - LARGE DEVICE BROKE OFF. REMOVED ALL THE BROKEN PARTS AND ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17390998 ↗
  • MalfunctionApr 19, 2023Break; Break

    IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE BIO-INTRAFIX TIBIAL SHEATH - LARGE DEVICE WAS BROKEN. REMOVED ALL THE BROKEN PARTS. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16770874 ↗
  • InjuryFeb 14, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS IS REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED BY THE SALES REP IN CHINA THAT A PATIENT UNDERWENT AN ARTHROSCOPIC RECONSTRUCTION PROCEDURE FOR ANTERIOR CRUCIATE LIGAMENT INJURY ON (B)(6) 2022; AND WAS IMPLANTED WITH TWO BIOINTRAFIX IMPLANTS; BIO-INTRAFIX TIBIAL SHEATH - LARGE DEVICE AND BIO-INTRAFIX TAPERED SCREW, 7-9 MM X 30 MM DEVICE. ACCORDING TO THE REPORT, THE PATIENT DEVELOPED REJECTION WITH PERSISTENT LOW-GRADE FEVER THREE MONTHS AFTER THE INITIAL PROCEDURE. IT WAS FURTHER REPORTED THAT A SECOND SURGERY WAS UNDERGONE TO REMOVE THE IMPLANTS ON (B)(6) 2023. AND THE PEEK IMPLANTS WERE REPLACED FOR IMPLANTING. IT WAS REPORTED THAT THE PATIENT WAS IN STABLE CONDITION NOW. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Foreign Body Reaction; Inflammation

    MAUDE report 16373532 ↗
  • MalfunctionNov 22, 2022Break; Break

    IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION SURGERY ON 10/27/2022, IT WAS OBSERVED THAT THE BIO-INTRAFIX TIBIAL SHEATH - LARGE DEVICE WAS BROKEN OFF. REMOVED ALL THE BROKEN PARTS. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15848248 ↗
  • MalfunctionNov 22, 2022Break; Break

    IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION SURGERY ON (B)(6) 2022, IT WAS OBSERVED THAT THE BIO-INTRAFIX TIBIAL SHEATH - LARGE DEVICE WAS BROKEN OFF. REMOVED ALL THE BROKEN PARTS. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15848105 ↗
  • MalfunctionNov 22, 2022Break; Break

    IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION SURGERY ON (B)(6) 2022, IT WAS OBSERVED THAT THE BIO-INTRAFIX TIBIAL SHEATH - LARGE WAS BROKEN OFF. REMOVED ALL THE BROKEN PARTS. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15847995 ↗

FDA clearance and classification

Cleared through 510(k) K032167 on Oct 15, 2003 (decision: Substantially Equivalent). Submitted by Mitek Worldwide as “Bio-Intrafix Tibial Screw And Sheath”. The FDA review took 91 days. FDA record ↗

The FDA regulates product code HWC (screw, fixation, bone) under 21 CFR 888.3040 as a Class II device. 21 CFR 888.3040 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the BIOINTRAFIX 254628

Has the BIOINTRAFIX 254628 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10886705013241), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the BIOINTRAFIX 254628?

The FDA's MAUDE database lists 91 adverse event reports for this device, received between Jun 18, 2018 and Sep 19, 2025. In those reports, the most common specific device problems are “Break” (69), “Crack” (9) and “Fracture” (6); the most common patient problems are “Unspecified Infection” (3), “Inflammation” (2) and “Not Applicable” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the BIOINTRAFIX 254628?

None of the reports are classified as deaths, 5 as injuries and 86 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the BIOINTRAFIX 254628 FDA approved?

It was cleared through 510(k) K032167 on Oct 15, 2003. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the BIOINTRAFIX 254628 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the BIOINTRAFIX 254628, and is it still sold?

The FDA's device database lists Medos International Sàrl as the company responsible for it (brand BIOINTRAFIX, model 254628). Its record shows it as in commercial distribution.