FMS 281103Recalls & reported problems
Medos International Sàrl · FMS Fluid Management System Intermediary Tubing Without One-Way valve · Class 2
All models of FMS (39)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10886705015764), received between Apr 8, 2020 and Jul 25, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
3 other recalls cover devices cleared under the same FDA submission (K951843); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Overfill2
- Crack1
- Device Contaminated During Manufacture or Shipping1
- Leak/Splash1
- Output Problem1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 25, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT PRIOR TO AN ANTERIOR CRUCIATE LIGAMENT SURGICAL PROCEDURE, THE INTERMEDIARY TUBESET DEVICE TIP WAS ABOUT TO COME OFF. THE TUBING WAS NOT USED IN THE SURGERY AT THE DISCRETION OF THE SURGEON, AND A SUBSTITUTE WAS USED IN THE SURGERY. THE SURGERY WAS COMPLETED SUCCESSFULLY. THERE WAS NO HARM TO THE PATIENT AND NO SURGICAL DELAY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22607758 ↗MalfunctionApr 23, 2025Overfill; Output Problem; Overfill; Output Problem
THIS IS REPORT 4 OF 4 FOR (B)(4). IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR PROCEDURE ON THE SHOULDER JOINT, THE FMS VUE II PUMP-SHAVER BOX DEVICE AND THE INTERMEDIARY TUBESET 24PK DEVICE ACCUMULATED AN EXCESSIVE AMOUNT OF WATER IN THE FILL CHAMBER, CAUSING THE PUMP TO STOP. ALTHOUGH THE INFLOW TUBE WAS REPLACED, THE SAME ISSUE PERSISTED, RENDERING THE DEVICES UNUSABLE. IT WAS REPORTED THAT THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE AS A SPARE DEVICE WAS AVAILABLE TO COMPLETE THE SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21891492 ↗MalfunctionMar 19, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping
IT WAS REPORTED THAT PRIOR TO AN UNSPECIFIED SURGICAL PROCEDURE, HAIR WAS DISCOVERED INSIDE THE SEALED TRAY OF THE INTERMEDIARY TUBESET 24PK DEVICE. IT WAS REPORTED THAT THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE AS A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21645156 ↗MalfunctionMay 31, 2022Use of Device Problem; Detachment of Device or Device Component; Use of Device Problem; Detachment of Device or Device Component
IT WAS REPORTED BY THE SALES REP THAT THE BLUE CAP OF THE FMS FLUID MANAGEMENT SYSTEM INTERMEDIARY TUBING WITHOUT ONE-WAY VALVE WAS OFF. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE ¿FEMALE VENTED CAP BLUE¿ WAS DETACHED FROM THE TIP ON THE DEVICE. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14547549 ↗MalfunctionDec 29, 2020Crack; Leak/Splash; Crack; Leak/Splash
IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT DURING AN ARCR (ARTHROSCOPIC ROTATOR CUFF REPAIR) TREATING ROTATOR CUFF TEAR ON (B)(6) 2020, IT WAS OBSERVED THAT THE CONNECTING PART OF THE TUBING WAS BROKEN THAT THERE WAS FLUID LEAKING FROM IT. THE BROKEN AREA WAS CUT TO COMPLETE THE PROCEDURE WITH LESS THAN 30 MINUTES OF DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THE DEVICE WAS BAND NEW AND THE FIRST USE WHEN THE ISSUE OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11089788 ↗MalfunctionApr 8, 2020Overfill; Overfill
IT WAS REPORTED BY AFFILIATE VIA PERSONAL INTERACTION THAT DURING ROTATOR CUFF REPAIR THE INFLOW LIQUID THROUGH THE INFLOW TUBING PASSED OVER THE FILL CHAMBER MARK. AS LOW PRESSURE LAMP BLINKED, THEY TRIED SOME TROUBLESHOOTING SUCH AS DRAINING AND RECONNECTING THE DEVICES CONCERNED, WHICH DID NOT CORRECT THE ISSUE. FINALLY THE INFLOW TUBING AND THE INTERMEDIARY TUBING (281103) WERE REPLACED, WHICH LEAD TO THE COMPLETION OF THE PROCEDURE WITHOUT SURGICAL DELAY. THERE WAS NO HARM TO THE PATIENT. THE DEVICES ARE AVAILABLE TO BE RETURNED FOR EVALUATION. ADDITIONAL INFORMATION RECEIVED FROM THE AFFILIATE REPORTED THE DEVICES ARE NO LONGER AVAILABLE TO BE RETURNED FOR EVALUATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9940994 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K951843. One submission often covers many models, so check the product and lot details in each record.
- Z-0191-2014 · Aug 29, 2013 · TerminatedDePuy Mitek FMS Intermediary Tubing with One-Way valve Product Code: 281142 Product Usage: The Intermediary Tubing for FMS Fluid Management Systems connects the Inflow Tubing to
- Z-1948-2011 · Dec 3, 2010 · TerminatedDePuy Mitek FMS Irrigation Intermediary Tube Without One-way Valve Arthroscopy tubing (100 units/box) (International Distribution Only) Product Code: 281102 Fluid Tissue Management
- Z-1946-2011 · Dec 3, 2010 · TerminatedDePuy Mitek FMS Irrigation Intermediary Tube With One-way Valve Arthroscopy tubing Product Code: 281142 Fluid Tissue Management
Other ARTHROSCOPE devices
- FMS 283512Medos International Sàrl · ARTHROSCOPE1,800 reports
- Arthrex® AR-6480Arthrex, Inc. · ARTHROSCOPE949 reports
- FMS VUE 284002Depuy Mitek, LLC · ARTHROSCOPE829 reports
- MITEK 242018Medos International Sàrl · Arthroscope731 reports
- METRx® System 9560100Medtronic Sofamor Danek, Inc. · Arthroscope667 reports
- FMS VUE II 284004Medos International Sàrl · Arthroscope562 reports
- VAPR COOLPULSE 228146Depuy Mitek, LLC · ARTHROSCOPE474 reports
- FMS DUO 284580Depuy Mitek, LLC · ARTHROSCOPE458 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page