FMS 284504Recalls & reported problems
Depuy Mitek, LLC · FMS Fluid Management System Inflow Tubing (FMS DUO Plus or FMS Solo) · Class 2
Quick summary
This DePuy Mitek FMS inflow tubing is used with fluid management systems during arthroscopic surgery. Many of the 81 reports describe the tubing's chamber overfilling, the pump reading high pressure, and fluid leaking. The reports include 1 injury and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 81
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 81 adverse event reports that list this exact device (DI 10886705017393), received between Apr 6, 2018 and May 20, 2024 — classified as 1 injury, 80 malfunctions.
The most frequently reported specific device problem is “Overfill” (30 reports).
3 other recalls cover devices cleared under the same FDA submission (K951843); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Overfill30
- Filling Problem17
- Fluid/Blood Leak16
- Leak/Splash8
- Increase in Pressure6
- Crack3
- Break2
- Output Problem2
Patient problems
- No Consequences Or Impact To Patient49
- No Clinical Signs, Symptoms or Conditions25
- Not Applicable5
- No Known Impact Or Consequence To Patient3
- Swelling1
Recent adverse event reports
The 25 most recent of 81 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 20, 2024Increase in Pressure; Increase in Pressure
IT WAS REPORTED BY THE SALES REP THAT DURING A KNEE SCOPE PROCEDURE ON (B)(6) 2024 IT WAS OBSERVED THAT THE IRRIGATION TUBESET DEVICE FMS FLUID MANAGEMENT SYSTEM/INFLOW TUBING (FMS DUO+ OR FMS SOLO) WAS CAUSING THE PUMP TO READ HIGH PRESSURE AND THE DEVICE HAD BAD TUBING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A SLIGHT DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19352379 ↗MalfunctionMay 20, 2024Increase in Pressure; Increase in Pressure
IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL THAT DURING KNEE SCOPE PROCEDURE ON (B)(6) 2024 THE IRRIGATION TUBESET DEVICE FMS FLUID MANAGEMENT SYSTEM/INFLOW TUBING (FMS DUO+ OR FMS SOLO) WAS CAUSING THE PUMP TO READ HIGH PRESSURE AND HAD BAD TUBING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19352295 ↗MalfunctionMay 20, 2024Increase in Pressure; Increase in Pressure
IT WAS REPORTED BY THE SALES REP THAT DURING A SHOULDER SCOPE PROCEDURE ON (B)(6) 2024 IT WAS OBSERVED THAT THE IRRIGATION TUBESET DEVICE FMS FLUID MANAGEMENT SYSTEM/INFLOW TUBING (FMS DUO+ OR FMS SOLO) WAS CAUSING THE PUMP TO READ HIGH PRESSURE AND THE DEVICE HAD BAD TUBING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A SLIGHT DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19351324 ↗MalfunctionApr 24, 2024Overfill; Overfill
IT WAS REPORTED BY THE SALES REP THAT DURING A SHOULDER SCOPE PROCEDURE ON 3/10/2024 IT WAS OBSERVED THAT THE IRRIGATION TUBESET DEVICE CHAMBERS OVERFILLED AND WATER WENT TO THE SENSOR VALVE ON TUBING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A THREE MINUTE DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19171308 ↗MalfunctionMar 13, 2024Overfill; Overfill
IT WAS REPORTED BY THE SALES REP THAT DURING AN UNKNOWN SURGICAL PROCEDURE ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE FMS FLUID MANAGEMENT SYSTEM/INFLOW TUBING (FMS DUO+ OR FMS SOLO) HAD THE PUMP FILL CHAMBER KEPT FILLING UP AND SHUTTING THE PUMP THEN GOING TO LOW PRESSURE MODE. THEY CHANGED THE PUMP WITH NO RESOLUTION. ANOTHER LIKE DEVICE WAS USED TO COMPLETE PROCEDURE WITH A COUPLE OF MINUTES SURGICAL DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18896302 ↗MalfunctionMar 6, 2024Overfill; Overfill
IT WAS REPORTED BY THE SALES REP THAT DURING AN UNKNOWN SURGICAL PROCEDURE ON (B)(6) 2024, IT WAS OBSERVED THAT THE FMS FLUID MANAGEMENT SYSTEM INFLOW TUBING (FMS DUO+ OR FMS SOLO) HAD THE PUMP FILL CHAMBER KEPT FILLING UP AND SHUTTING THE PUMP THEN GOING TO LOW PRESSURE MODE. THEY CHANGED THE PUMP WITH NO RESOLUTION. ANOTHER LIKE DEVICE WAS USED TO COMPLETE PROCEDURE WITH A COUPLE OF MINUTES SURGICAL DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18849511 ↗MalfunctionMar 6, 2024Overfill; Overfill
IT WAS REPORTED BY THE SALES REP THAT DURING AN UNKNOWN SURGICAL PROCEDURE ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE FMS FLUID MANAGEMENT SYSTEM/INFLOW TUBING (FMS DUO+ OR FMS SOLO) HAD THE PUMP FILL CHAMBER KEPT FILLING UP AND SHUTTING THE PUMP THEN GOING TO LOW PRESSURE MODE. THEY CHANGED THE PUMP WITH NO RESOLUTION. ANOTHER LIKE DEVICE WAS USED TO COMPLETE PROCEDURE WITH A COUPLE OF MINUTES SURGICAL DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18849505 ↗MalfunctionJan 25, 2024Overfill; Overfill
IT WAS REPORTED BY SALES REP THAT DURING A SHOULDER REPAIR PROCEDURE ON (B)(6) 2024, IT WAS OBSERVED THAT THE CHAMBER ON THE FMS FLUID MANAGEMENT SYSTEM/INFLOW TUBING (FMS DUO+ OR FMS SOLO) DEVICE FILLED UP. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18579834 ↗MalfunctionNov 8, 2023Overfill; Overfill
IT WAS REPORTED BY THE SALES REP IN ISRAEL THAT DURING A SHOULDER ARTHROSCOPY SURGICAL PROCEDURE ON AN UNKNOWN DATE, WHILE USING AN FMS DUO PUMP DEVICE, THE SALINE CHAMBER FILLED ALL WAY UP TWO DIFFERENT SETS OF INFLOW TUBING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITHOUT DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 2 OF 2 FOR THE SAME EVENT IN (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18094239 ↗MalfunctionNov 8, 2023Overfill; Overfill
IT WAS REPORTED BY THE SALES REP IN ISRAEL THAT DURING A SHOULDER ARTHROSCOPY SURGICAL PROCEDURE ON AN UNKNOWN DATE, WHILE USING AN FMS DUO PUMP DEVICE, THE SALINE CHAMBER FILLED ALL WAY UP TWO DIFFERENT SETS OF INFLOW TUBING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITHOUT DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 2 OF 2 FOR THE SAME EVENT IN (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18093557 ↗MalfunctionNov 8, 2023Overfill; Overfill
IT WAS REPORTED BY THE SALES REP IN ISRAEL THAT DURING A SHOULDER ARTHROSCOPY SURGICAL PROCEDURE ON AN UNKNOWN DATE, WHILE USING AN FMS DUO PUMP DEVICE, THE SALINE CHAMBER FILLED ALL WAY UP TWO DIFFERENT SETS OF INFLOW TUBING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITHOUT DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 1 OF 2 FOR THE SAME EVENT IN (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18093492 ↗MalfunctionJun 27, 2023Improper Flow or Infusion; Improper Flow or Infusion
IT WAS REPORTED BY THE DISTRIBUTOR IN BRAZIL THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE FMS FLUID MANAGEMENT SYSTEM/INFLOW TUBING (FMS DUO+ OR FMS SOLO) DEVICE DID NOT WORK PROPERLY. ACCORDING TO THE REPORT, IT WAS NOT GIVING PRESSURE AND THE SERUM FLOW. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17212686 ↗MalfunctionMar 10, 2023Increase in Pressure; Difficult or Delayed Separation; Increase in Pressure; Difficult or Delayed Separation
IT WAS REPORTED BY SALES REP THAT DURING A LEFT KNEE ARTHROSCOPY PROCEDURE WITH PATELLAR CONDULARPLASTY ON (B)(6)2023, IT WAS OBSERVED THAT WHEN THEY WENT TO PLUG IN THE FMS FLUID MANAGEMENT SYSTEM/INFLOW TUBING (FMS DUO+ OR FMS SOLO) DEVICE INTO THE MACHINE, THERE WAS TOO MUCH PRESSURE AND THE CORD WAS NOT HOLDING IN PLACE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH A DELAY OF 15 MINUTES. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16523994 ↗MalfunctionSep 6, 2022Break; Break
IT WAS REPORTED BY THE SALES REP IN HONG KONG THAT THE FMS FLUID MANAGEMENT SYSTEM/INFLOW TUBING (FMS DUO+ OR FMS SOLO) DEVICE WAS BROKEN. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15364750 ↗MalfunctionAug 11, 2022Overfill; Output Problem; Overfill; Output Problem
THIS IS REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL IN FRANCE THAT DURING AN UNKNOWN SURGERY ON (B)(6) 2022, IT WAS OBSERVED THAT WHILE PURGING THE "GREEN NOTCH" FMS FLUID MANAGEMENT SYSTEM/INFLOW TUBING (FMS DUO+ OR FMS SOLO) DEVICE, THE RESERVOIR FILLED-UP COMPLETELY THEN THE PUMP WENT INTO ALARM. ACCORDING TO THE REPORT, THE PUMP WAS CHANGED AND THE SAME PROBLEM OCCURRED WITH ANOTHER SET OF TUBING DEVICE FROM THE SAME LOT. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15212923 ↗MalfunctionAug 11, 2022Overfill; Output Problem; Overfill; Output Problem
THIS IS REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL IN FRANCE THAT DURING AN UNKNOWN SURGERY ON (B)(6) 2022, IT WAS OBSERVED THAT WHILE PURGING THE "GREEN NOTCH" FMS FLUID MANAGEMENT SYSTEM/INFLOW TUBING (FMS DUO+ OR FMS SOLO) DEVICE, THE RESERVOIR FILLED-UP COMPLETELY THEN THE PUMP WENT INTO ALARM. ACCORDING TO THE REPORT, THE PUMP WAS CHANGED AND THE SAME PROBLEM OCCURRED WITH ANOTHER SET OF TUBING DEVICE FROM THE SAME LOT. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15212712 ↗MalfunctionJun 15, 2022Leak/Splash; Leak/Splash
IT WAS REPORTED BY THE SALES REP IN UNITED KINGDOM THAT DURING AN UNKNOWN ARTHROSCOPIC PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE FMS FLUID MANAGEMENT SYSTEM/INFLOW TUBING (FMS DUO+ OR FMS SOLO) DEVICE WAS FAULTY AND LEAKING. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14697140 ↗MalfunctionJun 15, 2022Leak/Splash; Leak/Splash
IT WAS REPORTED BY THE SALES REP IN UNITED KINGDOM THAT DURING AN UNKNOWN ARTHROSCOPIC PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE FMS FLUID MANAGEMENT SYSTEM/INFLOW TUBING (FMS DUO+ OR FMS SOLO) DEVICE WAS FAULTY AND LEAKING. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14697086 ↗MalfunctionFeb 16, 2022Crack; Leak/Splash; Moisture Damage; Crack; Leak/Splash; Moisture Damage
IT WAS REPORTED BY THE CUSTOMER IN (B)(6) THAT DURING AN ARTHROSCOPY SURGERY ON (B)(6) 2022, IT WAS OBSERVED THAT THE IRRIGATION TUBESET DEVICE WAS PERFORATED ON THE FLEXIBLE PART AT THE PUMP SHAFT AFTER 35 MINUTES OF USE. IT WAS FURTHER REPORTED THAT JUST AFTER AN IRRIGATION BAG REPLACEMENT, THERE WAS SALINE DRIPPING ON SENSITIVE ELECTRICAL EQUIPMENT (ARTHROSCOPY COLUMN). THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13542371 ↗MalfunctionJul 27, 2021Overfill; Overfill
IT WAS REPORTED BY THE SALES REP THAT DURING A ROTATOR CUFF REPAIR PROCEDURE ON (B)(6)2021, IT WAS OBSERVED THAT THE CHAMBER ON THE PUMP STARTED TO FILL UP AND A HIGH PRESSURE CODE WAS DISPLAYED. IT WAS REPORTED THAT A LOT OF SOLUTION WAS SQUIRTING OUT FROM THE JOINT. ACCORDING TO THE REPORT, AFTER REPLACING THE FMS FLUID MANAGEMENT SYSTEM/INFLOW TUBING (FMS DUO+ OR FMS SOLO) THE ERROR CODE WAS CLEARED AND THE ISSUE OF THE CHAMBER GOT SOLVED. THE PROCEDURE WAS COMPLETED WITH A DELAY OF LESS THAN FIVE MINUTES. THERE WERE NO ADVERSE NO PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12233857 ↗MalfunctionJul 12, 2021Break; Leak/Splash; Break; Leak/Splash
IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2021, IT WAS OBSERVED THAT THE IRRIGATION TUBE DEVICE HAD A HOLE AND A TEAR THAT IT WAS LEAKING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITHOUT DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12154652 ↗MalfunctionJun 10, 2021Crack; Leak/Splash; Crack; Leak/Splash
IT WAS REPORTED BY THE CUSTOMER THAT THE FMS FLUID MANAGEMENT SYSTEM/INFLOW TUBING (FMS DUO+ OR FMS SOLO) WAS TORN AND POURING WATER. THERE WAS NO PROCEDURE NOR PATIENT INVOLVEMENT REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11975642 ↗MalfunctionApr 8, 2021Overfill; Overfill
IT WAS REPORTED BY SALES REP THAT DURING A SHOULDER REPAIR PROCEDURE ON (B)(6) 2021, IT WAS OBSERVED THAT THE CHAMBER ON THE IRRIGATION DEVICE OVERFLOWED WHILE SETTING-UP. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITHOUT DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11634316 ↗MalfunctionApr 7, 2021Overfill; Overfill
IT WAS REPORTED BY SALES REP THAT DURING A SHOULDER PROCEDURE ON (B)(6) 2021, IT WAS OBSERVED THAT THE CHAMBER ON THE IRRIGATION TUBING SET DEVICE FILLED UP FULLY. ANOTHER PUMP DEVICE WAS USED TO WITH ANOTHER TUBING DEVICE. THERE WAS NO SURGICAL DELAY OR PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11629105 ↗MalfunctionOct 23, 2020Residue After Decontamination; Residue After Decontamination
IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT PREOPERATIVELY TO A ARTHROSCOPIC SUBACROMIAL DECOMPRESSION (ASAD) PROCEDURE, IT WAS OBSERVED THAT THE TUBING WAS SPLIT WHEN CONNECTED TO MACHINE WHILE THE PACKAGING WAS INTACT. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE TUBING HAD RESIDUES OF SALINE WATER. THERE WAS NO DELAY IN THE SURGERY DUE TO THE EVENT. ANOTHER SET OF TUBING WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10726441 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K951843. One submission often covers many models, so check the product and lot details in each record.
- Z-0191-2014 · Aug 29, 2013 · TerminatedDePuy Mitek FMS Intermediary Tubing with One-Way valve Product Code: 281142 Product Usage: The Intermediary Tubing for FMS Fluid Management Systems connects the Inflow Tubing to
- Z-1946-2011 · Dec 3, 2010 · TerminatedDePuy Mitek FMS Irrigation Intermediary Tube With One-way Valve Arthroscopy tubing Product Code: 281142 Fluid Tissue Management
- Z-1948-2011 · Dec 3, 2010 · TerminatedDePuy Mitek FMS Irrigation Intermediary Tube Without One-way Valve Arthroscopy tubing (100 units/box) (International Distribution Only) Product Code: 281102 Fluid Tissue Management
Other ARTHROSCOPE devices
- FMS 283512Medos International Sàrl · ARTHROSCOPE1,800 reports
- Arthrex® AR-6480Arthrex, Inc. · ARTHROSCOPE949 reports
- FMS VUE 284002Depuy Mitek, LLC · ARTHROSCOPE829 reports
- MITEK 242018Medos International Sàrl · Arthroscope731 reports
- METRx® System 9560100Medtronic Sofamor Danek, Inc. · Arthroscope667 reports
- FMS VUE II 284004Medos International Sàrl · Arthroscope562 reports
- VAPR COOLPULSE 228146Depuy Mitek, LLC · ARTHROSCOPE474 reports
- FMS DUO 284580Depuy Mitek, LLC · ARTHROSCOPE458 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.