MILAGRO 231812Recalls & reported problems
Medos International Sàrl · MILAGRO BR INTERFERENCE SCREW Absorbable - TCP/PLGA 8 x 15mm · Class 2 · Implant
All models of MILAGRO (26)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886705020423), received between Aug 6, 2018 and Mar 15, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 15, 2024Break; Break
REPORT 3 OF 4 FOR (B)(4). IT WAS REPORTED BY SALES REP THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE ON (B)(6) 2024, IT WAS OBSERVED THAT A 8X15MM MILAGRO INTERFERENCE SCREW, 5X12MM MILAGRO INTERFERENCE SCREW, 8X23MM MILAGRO ADVANCE INTERFERENCE SCREW AND 10X23MM MILAGRO ADVANCE INTERFERENCE SCREW DEVICES WERE BROKEN. IT WAS REPORTED THAT THE PATIENT HAD EXTREMELY HARD BONE, AND IT WAS A REVISION CASE. IT WAS REPORTED THAT THE SAME LARGE TUNNELS WERE USED AND THE DOCTOR WANTED TO START WITH A LARGER SCREW BUT IT WOULD NOT DRIVE EVEN AFTER NOTCHING AND TAPPING FOUR TIMES. ACCORDING TO THE REPORTER, MULTIPLE SCREWS WERE TRIED, THEY ALL ENDED UP BREAKING AND THE BROKEN SCREW STAYED INSIDE THE PATIENT AS IT WASN'T BADLY BROKEN SINCE IT WAS STILL HOLDING THE GRAFT IN PLACE. IT WAS REPORTED THAT THE DOCTOR LEFT THE LEFT BROKEN SCREW INSIDE PATIENT AS IT HELD THE GRAFT BUT BROKE IN HALF. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18911160 ↗MalfunctionAug 6, 2018Break; Break
IT WAS REPORTED BY THE SALES REP THAT THE CUSTOMERS MILAGRO BR SCREW 8X15MM BROKE ONCE INSERTED 3/4 WAY IN, SO THE SURGEON BACKED THE SCREW OUT, AND USED ANOTHER LIKE DEVICE IN THE SAME BONE HOLE TO COMPLETE THE PROCEDURE, WITH NO PATIENT CONSEQUENCES AND NO DELAY. THE SALES REP IS REQUEST THAT ONLY HIS TRUCK STOCK BE ADJUSTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7753456 ↗
Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices
- GRYPHON ORTHOCORD 210813Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE712 reports
- HEALIX ADVANCE 222330Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall607 reports
- HEALIX ADVANCE ORTHOCORD 222295Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE640 reports
- HEALICOIL 72205137Smith & Nephew, Inc. · Fastener, fixation, biodegradable, soft tissue427 reports
- HEALIX ADVANCE ORTHOCORD 222296Medos International Sàrl · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE320 reports
- TWINFIX 72202597Smith & Nephew, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE288 reports
- Arthrex® AR-2324BCCArthrex, Inc. · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE254 reports
- HEALIX ADVANCE 222331Depuy Mitek, LLC · FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE1 recall196 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page