HEALIX ADVANCE ORTHOCORD 222303Recalls & reported problems

Medos International Sàrl · HEALIX ADVANCE PEEK ANCHOR w/ORTHOCORD PEEK Anchor (1) Violet (1) Blue strand, Size 2 (5 metric) ORTHOCORD braided composite suture, 36 Inches (91cm) 4.5mm · Class 2 · Implant

All models of HEALIX ADVANCE ORTHOCORD (21)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medos HEALIX ADVANCE PEEK bone anchor with its Orthocord suture is used to fasten tissue to bone in procedures like rotator cuff and knee ligament repair. The 45 reports mostly describe the anchor or suture breaking, cracking, fraying or pulling out during surgery, with another device used to finish the operation in most cases. There are no deaths and 1 injury. No recalls name this device, and the 9 reports in each of the last two 12-month periods show no change.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
48
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 48 adverse event reports that list this exact device (DI 10886705021390), received between Jul 5, 2017 and Sep 29, 2026 — classified as 1 injury, 47 malfunctions.

The most frequently reported specific device problem is “Break” (28 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 4 reportsAug 2024: 5 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 3 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 3 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break28
  • Migration6
  • Crack5
  • Material Deformation3
  • Appropriate Device Problem Term/Code Not Available2
  • Detachment of Device or Device Component2
  • Device-Device Incompatibility2
  • Mechanical Jam2

Patient problems

  • No Clinical Signs, Symptoms or Conditions37
  • No Consequences Or Impact To Patient9
  • Device Embedded In Tissue or Plaque1
  • Not Applicable1

Recent adverse event reports

The 25 most recent of 48 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Migration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933489 ↗
  • MalfunctionSep 29, 2026Migration

    This is report 2 of 2 for (b)(4)It was reported that during a Rotator Cuff Repair procedure, it was discovered that the anchor on the 4.5 HEALIX ADVANCE PEEK W/CORD device pulled out. A replacement device was used to continue the surgery, but the same issue occurred again. It was not reported if there were any delays to the surgical procedure. It was reported that a spare device was used to complete the surgery. There were no reports of injuries, medical intervention or prolonged hospitalization. All available information has been disclosed.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933487 ↗
  • MalfunctionSep 25, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50903505 ↗
  • MalfunctionApr 3, 2026Crack

    THIS IS REPORT 2 OF 2 OF (B)(4). IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE HEALIX ADVANCE ORTHOCORD DEVICE SUTURE BROKE AND FRAYED INTRAOPERATIVELY. THE SURGEON CONTINUED TO USE HEALIX ADVANCE BUT WOULD NOT USE THE ORTHOCORD DEVICE WITH THE PERMACORD. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24775420 ↗
  • MalfunctionApr 1, 2026Crack

    THIS IS REPORT 1 OF 2 OF (B)(4). IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE IT WAS OBSERVED THAT THE HEALIX ADVANCE ORTHOCORD DEVICE SUTURE BROKE AND FRAYED INTRAOPERATIVELY. THE SURGEON CONTINUED TO USE HEALIX ADVANCE BUT WOULD NOT USE THE ORTHOCORD DEVICE WITH THE WITH PERMACORD. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24758570 ↗
  • MalfunctionMar 31, 2026Crack

    THIS IS REPORT 2 OF 3 OF (B)(4). IT WAS REPORTED THAT, DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE HEALIX ADVANCE ORTHOCORD DEVICE SUTURE BROKE AND FRAYED INTRAOPERATIVELY. THE SURGEON CONTINUED TO USE HEALIX ADVANCE BUT WOULD NOT USE THE ORTHOCORD DEVICE WITH THE WITH PERMACORD. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24742211 ↗
  • MalfunctionMar 31, 2026Crack

    THIS IS REPORT 1 OF 3 OF (B)(4). IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE IT WAS OBSERVED THAT THE HEALIX ADVANCE ORTHOCORD DEVICE SUTURE BROKE AND FRAYED INTRAOPERATIVELY. THE SURGEON CONTINUED TO USE HEALIX ADVANCE BUT WOULD NOT USE THE ORTHOCORD DEVICE WITH THE WITH PERMACORD. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24742180 ↗
  • MalfunctionMar 16, 2026Migration

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ANTERIOR CRUCIATE LIGAMENT REPAIR PROCEDURE THAT THE 4.5 HEALIX ADVANCE PEEK W/CORD ANCHOR DEVICE PULLED OUT. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24607209 ↗
  • MalfunctionFeb 12, 2026Migration

    IT WAS REPORTED THAT DURING ROTATOR CUFF REPAIR SURGERY, THE ANCHOR WAS PULLED OUT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24320860 ↗
  • MalfunctionDec 16, 2025Difficult to Advance; Material Deformation; Difficult to Advance; Material Deformation

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGERY, WHEN INSERTING THE 4.5 HEALIX ADVANCE PEEK W/CORD DEVICE ANCHOR, IT WAS OBSERVED THAT THE THREAD OF ANCHOR WAS PEELED AND THE ANCHOR WOULD NOT SCREW IN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23819044 ↗
  • MalfunctionDec 10, 2025Break; Break

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE, WHEN INSERTING THE 4.5 HEALIX ADVANCE PEEK W/CORD ANCHOR DEVICE, IT WAS OBSERVED THAT THE ANCHOR WAS BROKEN OFF. IT WAS REPORTED THAT ALL THE BROKEN PARTS WERE REMOVED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23770274 ↗
  • MalfunctionNov 5, 2025Break; Break

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGERY, IT WAS OBSERVED THAT THE SURGEON WAS PASSING SUTURES THROUGH THE ROTATOR CUFF AFTER IMPLANTING ANCHOR AND THE 4.5 HEALIX ADVANCE PEEK W/CORD DEVICE SUTURE BROKE IN HALF. IT WAS REPORTED THAT TENSION WAS FELT ON THE DEVICE AND AND THE SUTURE WAS EASED HOWEVER, THE SUTURE STILL BROKE. ACCORDING TO THE REPORTER, THIS LEAD TO THE SURGEON HAVING ONLY HAVE ONE POINT OF FIXATION VERSUS TWO. IT WAS REPORTED THAT BESIDES THE SUTURE BEING TOO WEAK OR TOO THIN, THE SURGEON DID NOT HAVE ENOUGH SPACE TO PUT A SECOND ANCHOR IN AND THEREFORE, IT POTENTIALLY COMPROMISED THE INTEGRITY OF THE PATIENT'S REPAIR. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23475339 ↗
  • MalfunctionSep 2, 2025Break; Break

    REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGERY, IT WAS OBSERVED THAT A 4.5 HEALIX ADVANCE PEEK W/CORD DEVICE ANCHOR WAS BROKEN OFF AND ALL THE BROKEN PARTS WERE REMOVED. ACCORDING TO THE REPORTER, A 5.5 HEALIX ADVANCE KNTLSS BR DEVICE ANCHOR WAS ALSO BROKEN OFF AND ALL THE BROKEN PARTS WERE REMOVED. ANOTHER DEVICES WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22931935 ↗
  • MalfunctionJul 24, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, WHEN THE 4.5 HEALIX ADVANCE PEEK W/CORD ANCHOR DEVICE WAS INSERTED, IT WAS OBSERVED THAT THE SUTURE DETACHED FROM THE ANCHOR. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22588873 ↗
  • MalfunctionMay 5, 2025Break; Break

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE 4.5 HEALIX ADVANCE PEEK W/CORD ANCHOR DEVICE WAS BROKEN. IT WAS REPORTED THAT ALL BROKEN PARTS WERE REMOVED FROM THE PATIENT. IT WAS REPORTED THAT THE SURGEON CHANGED TO ANOTHER 4.5 HEALIX ADVANCE PEEK W/CORD ANCHOR DEVICE AND THE SAME PROBLEM OCCURRED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS REPORT IS 1 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21962855 ↗
  • MalfunctionMay 5, 2025Break; Break

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE 4.5 HEALIX ADVANCE PEEK W/CORD ANCHOR DEVICE WAS BROKEN. IT WAS REPORTED THAT ALL BROKEN PARTS WERE REMOVED FROM THE PATIENT. IT WAS REPORTED THAT THE SURGEON CHANGED TO ANOTHER 4.5 HEALIX ADVANCE PEEK W/CORD ANCHOR DEVICE AND THE SAME PROBLEM OCCURRED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS REPORT IS 2 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21962852 ↗
  • MalfunctionFeb 20, 2025Break; Break

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE THE 4.5 HEALIX ADVANCE PEEK W/CORD DEVICE ANCHOR BROKE OFF. ALL BROKEN PARTS WERE REMOVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21426814 ↗
  • MalfunctionDec 13, 2024Material Deformation; Appropriate Device Problem Term/Code Not Available; Material Deformation; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE 4.5 HEALIX ADVANCE PEEK W/CORD DEVICE WAS DEFECTIVE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE ANCHOR TIP OF THE DEVICE WAS DEFORMED DUE TO BONE FRICTION WHILE INSERTING. IT WAS UNKNOWN IF THERE WAS PATIENT INVOLVEMENT. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2024. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20926311 ↗
  • MalfunctionDec 13, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE 4.5 HEALIX ADVANCE PEEK W/CORD DEVICE WAS DEFECTIVE. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE ANCHOR TIP OF THE DEVICE WAS DEFORMED DUE TO BONE FRICTION WHILE INSERTING. IT WAS UNKNOWN IF THERE WAS PATIENT INVOLVEMENT. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2024. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20926310 ↗
  • MalfunctionDec 9, 2024Break; Break

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR PROCEDURE, THE 4.5 HEALIX ADVANCE PEEK W/CORD DEVICE ANCHOR WAS BROKEN OFF, REMOVED ALL THE BROKEN PARTS. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20878838 ↗
  • MalfunctionNov 11, 2024Break; Break

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR PROCEDURE, THE 4.5 HEALIX ADVANCE PEEK W/CORD WAS BROKEN OFF. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20652327 ↗
  • MalfunctionSep 4, 2024Break; Break

    THIS IS REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED FROM CHINA THAT DURING A RECONSTRUCTION OF MEDIAL PATELLAR RETINACULUM PROCEDURE, IT WAS DISCOVERED THAT THE ANCHOR OF THE 4.5 HEALIX ADVANCE PEEK W/CORD DEVICE BROKE OFF. ALL THE BROKEN PARTS WERE REMOVED. A REPLACEMENT DEVICE WAS USED TO CONTINUE THE SURGERY, BUT THE SAME ISSUE OCCURRED AGAIN. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20136112 ↗
  • MalfunctionSep 4, 2024Break; Break

    THIS IS REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED FROM CHINA THAT DURING A RECONSTRUCTION OF MEDIAL PATELLAR RETINACULUM PROCEDURE, IT WAS DISCOVERED THAT THE ANCHOR OF THE 4.5 HEALIX ADVANCE PEEK W/CORD DEVICE BROKE OFF. ALL THE BROKEN PARTS WERE REMOVED. A REPLACEMENT DEVICE WAS USED TO CONTINUE THE SURGERY, BUT THE SAME ISSUE OCCURRED AGAIN. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20136110 ↗
  • MalfunctionAug 27, 2024Break; Break

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE THE 4.5 HEALIX ADVANCE PEEK W/CORD DEVICE WAS BEING USED WHEN THE ANCHOR WAS BROKEN OFF. ALL BROKEN PARTS WERE REMOVED. CHANGED TO ANOTHER ONE TO CONTINUE THE SURGERY AND THE SAME PROBLEM OCCURRED WITH A 2ND 4.5 HEALIX ADVANCE PEEK W/CORD DEVICE. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20077610 ↗
  • MalfunctionAug 15, 2024Break; Break

    IT WAS REPORTED BY THE SALES REP IN HUNGARY THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE ON AN UNKNOWN DATE THE 4.5 HEALIX ADVANCE PEEK W/CORD DEVICE ANCHOR WAS BROKEN OFF, ALL BROKEN PARTS WERE REMOVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19996488 ↗

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →