HEALIX ADVANCE 222327Recalls & reported problems
Medos International Sàrl · HEALIX ADVANCE Drill Package Contains: 1 Drill 3.5mm · Class 1
All models of HEALIX ADVANCE (21)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886705021840), received between Sep 13, 2021 and Oct 12, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device-Device Incompatibility3
- Break1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 12, 2022Device-Device Incompatibility; Device-Device Incompatibility
HEALIX DRILL BIT WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? YES, WAS PROCEDURE SUCCESSFULLY COMPLETED? YES. IT WAS REPORTED BY THE SALES REP THAT DURING AN UNKNOWN SURGERY ON (B)(6) 2022, IT WAS OBSERVED THAT THE 3.5MM HEALIX DRILL DEVICE GOT STUCK IN THE GUIDE AND BINDED TO THE GUIDE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH A DELAY OF 2 TO 3 MINUTES. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15587511 ↗MalfunctionSep 9, 2022Break; Device-Device Incompatibility; Material Deformation; Break; Device-Device Incompatibility; Material Deformation
IT WAS REPORTED BY A SALES REP THAT DURING A LATERAL ANKLE STABILITY REPAIR PROCEDURE ON (B)(6)2022, IT WAS OBSERVED THAT THE 3.5MM HEALIX DRILL BIT DEVICE GOT STUCK IN THE GUIDE THEN STARTED TO BEND; AND GENERATED METAL SHAVINGS FROM THE BINDING. THE PROCEDURE WAS COMPLETED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15393871 ↗MalfunctionSep 13, 2021Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED BY THE SALES REP THAT DURING A BICEP TENODESIS PROCEDURE ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE GII/SUPER/RC DRGDE SLOT DEVICE AND THE HEALIX ADVANCE 3.5MM DRILL BIT DEVICE WERE WELDED TOGETHER THAT THEY COULD NOT GET APART. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12463396 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page