MILAGRO 231814Recalls & reported problems

Medos International Sàrl · MILAGRO BR INTERFERENCE SCREW Absorbable - TCP/PLGA 5 x 12mm · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Medos MILAGRO absorbable interference screw is a soft-tissue fixation screw used in arthroscopic ligament and rotator cuff repair. Most of the 58 reports describe the screw breaking during insertion or during surgery, with broken pieces then removed from the patient. There are no deaths and one injury; 57 reports are malfunctions. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
58
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 58 adverse event reports that list this exact device (DI 10886705021963), received between Apr 23, 2018 and Jun 11, 2026 — classified as 1 injury, 57 malfunctions.

The most frequently reported specific device problem is “Break” (54 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 1 reportsDec 2024: 4 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 4 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break54
  • Crack1
  • Detachment of Device or Device Component1
  • Material Frayed1
  • Migration1
  • Migration or Expulsion of Device1
  • Scratched Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions39
  • No Consequences Or Impact To Patient17
  • Not Applicable2
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 58 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 11, 2026Break

    IT WAS REPORTED THAT DURING AN UNSPECIFIED ARTHROSCOPIC PROCEDURE THE 5X12 MILAGRO INTERERENCE SCREW DEVICE BROKE. ALL THE BROKEN PARTS WERE REMOVED. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25477885 ↗
  • MalfunctionApr 24, 2026Break

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC LIGAMENT REPAIR SURGERY, THE SCREW (5X12 MILAGRO INTERERENCE SCREW) BROKE. ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24981548 ↗
  • MalfunctionMar 13, 2026Break

    IT WAS REPORTED THAT DURING AN UNKNOWN ARTHROSCOPIC PROCEDURE THAT THE 5X12 MILAGRO INTERERENCE SCREW DEVICE BROKE DURING INSERTION. ALL THE PIECES WERE REMOVED FROM THE PATIENT. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24590568 ↗
  • MalfunctionMar 10, 2026Break

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR PROCEDURE, THE 5X12 MILAGRO INTERERENCE SCREW SURGEON TRIED TO INSERT THE SCREW INTO THE BONE, IT BROKE AND BECAME UNSTABLE. IT SEEMED TO HAVE LACKED STRENGTH. THERE WAS NO SURGICAL DELAY AND NO HARM TO THE PATIENT USING DIFFERENT PRODUCT. NO REMAINS WERE LEFT INSIDE THE BODY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24561258 ↗
  • MalfunctionJan 2, 2026Break

    THIS IS REPORT 4 OF 4 FOR (B)(4). IT WAS REPORTED THAT DURING ARTHROSCOPY PROCEDURE, IT WAS DISCOVERED THAT THE MILAGRO INTERFRENCE SCREW DEVICE BROKE. ALL THE BROKEN PARTS WERE REMOVED FROM THE PATIENT. THREE REPLACEMENT DEVICES WERE USED TO CONTINUE THE SURGERY, BUT THE SAME ISSUE OCCURRED AGAIN. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23944910 ↗
  • MalfunctionJan 2, 2026Break

    THIS IS REPORT 3 OF 4 FOR (B)(4). IT WAS REPORTED THAT DURING ARTHROSCOPY PROCEDURE, IT WAS DISCOVERED THAT THE MILAGRO INTERFRENCE SCREW DEVICE BROKE. ALL THE BROKEN PARTS WERE REMOVED FROM THE PATIENT. THREE REPLACEMENT DEVICES WERE USED TO CONTINUE THE SURGERY, BUT THE SAME ISSUE OCCURRED AGAIN. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23944908 ↗
  • MalfunctionJan 2, 2026Break

    THIS IS REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, IT WAS DISCOVERED THAT WHEN THE 5X12 MILAGRO INTERERENCE SCREW DEVICE WAS INSERTED, IT BROKE OFF. ALL THE BROKEN PARTS WERE REMOVED FROM THE PATIENT. A REPLACEMENT DEVICE WAS USED TO CONTINUE THE SURGERY, BUT THE SAME ISSUE OCCURRED AGAIN. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN JUNE 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23943828 ↗
  • MalfunctionJan 2, 2026Break

    THIS IS REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, IT WAS DISCOVERED THAT WHEN THE 5X12 MILAGRO INTERERENCE SCREW DEVICE WAS INSERTED, IT BROKE OFF. ALL THE BROKEN PARTS WERE REMOVED FROM THE PATIENT. A REPLACEMENT DEVICE WAS USED TO CONTINUE THE SURGERY, BUT THE SAME ISSUE OCCURRED AGAIN. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN JUNE 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23943816 ↗
  • MalfunctionJul 9, 2025Break; Break

    IT WAS REPORTED THAT DURING A LIGAMENT RECONSTRUCTION SURGICAL PROCEDURE IT WAS OBSERVED THAT THE 5X12 MILAGRO INTERERENCE SCREW DEVICE ANCHOR BROKE OFF. ALL BROKEN PARTS WERE REMOVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22461165 ↗
  • MalfunctionJul 7, 2025Detachment of Device or Device Component; Migration; Detachment of Device or Device Component; Migration

    IT WAS REPORTED THAT DURING AN ANKLE RECONSTRUCTION SURGERY, IT WAS OBSERVED THAT THE 5X12 MILAGRO INTERERENCE SCREW DEVICE WAS PULLED OUT OF THE BONE TUNNEL. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22412033 ↗
  • MalfunctionMay 23, 2025Break; Break

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT SURGICAL PROCEDURE, THE 5X12 MILAGRO INTERERENCE SCREW DEVICE ANCHOR WAS BROKEN OFF, REMOVED ALL THE BROKEN PARTS. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22077813 ↗
  • MalfunctionDec 31, 2024Break; Break

    REPORT 2 OF 2 FOR (B)(4). IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE THE MILAGRO INTERERENCE SCREW DEVICE ANCHOR WAS BROKEN OFF. ALL THE BROKEN PARTS WERE REMOVED. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21047370 ↗
  • MalfunctionDec 31, 2024Break; Break

    REPORT 1 OF 2 FOR (B)(4). IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE THE MILAGRO INTERERENCE SCREW DEVICE ANCHOR WAS BROKEN OFF. ALL THE BROKEN PARTS WERE REMOVED. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21047368 ↗
  • MalfunctionDec 31, 2024Break; Break

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, THE 5X12 MILAGRO INTERFERENCE SCREW DEVICE ANCHOR WAS BROKEN OFF. IT WAS REPORTED THAT ALL THE BROKEN PARTS WERE REMOVED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21047065 ↗
  • MalfunctionDec 5, 2024Break; Break

    IT WAS REPORTED THAT DURING A LIGAMENT RECONSTRUCTION SURGICAL PROCEDURE THE 5X12 MILAGRO INTERERENCE SCREW DEVICE ANCHOR WAS BROKEN OFF. ALL BROKEN PARTS WERE REMOVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20858002 ↗
  • MalfunctionNov 21, 2024Break; Break

    IT WAS REPORTED THAT DURING AN ANKLE FUSION OSTEOTOMY SURGICAL PROCEDURE THE MILAGRO INTERERENCE SCREW DEVICE ANCHOR WAS BROKEN OFF. ALL THE BROKEN PARTS WERE REMOVED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20745805 ↗
  • MalfunctionOct 30, 2024Break; Break

    IT WAS REPORTED FROM (B)(6) THAT DURING A LIGAMENT RECONSTRUCTION PROCEDURE, IT WAS DISCOVERED THAT THE ANCHOR ON THE 5X12 MILAGRO INTERFERENCE SCREW DEVICE BROKE OFF. ALL THE BROKEN PARTS WERE REMOVED. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN (B)(6) 2024. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20568885 ↗
  • MalfunctionOct 30, 2024Break; Break

    IT WAS REPORTED FROM CHINA THAT DURING A LIGAMENT RECONSTRUCTION PROCEDURE, IT WAS DISCOVERED THAT THE ANCHOR ON THE 5X12 MILAGRO INTERFERENCE SCREW DEVICE BROKE OFF. ALL THE BROKEN PARTS WERE REMOVED. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN (B)(6) 2024. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20568637 ↗
  • MalfunctionAug 28, 2024Break; Break

    IT WAS REPORTED FROM CHINA THAT DURING AN ARTHROSCOPIC CARTILAGE AND LIGAMENT REPAIR SURGERY FOR THE RIGHT ANKLE, THE UNSPECIFIED SURGICAL PROCEDURE THE 5X12 MILAGRO INTERERENCE SCREW DEVICE ANCHOR WAS BROKEN OFF. ALL BROKEN PARTS WERE REMOVED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20086088 ↗
  • MalfunctionJul 12, 2024Break; Break

    IT WAS REPORTED BY THE SALES REP THAT DURING A RECONSTRUCTION OF THE TALOFIBULAR LIGAMENT SURGICAL PROCEDURE ON(B)(6)2024 THE 5X12 MILAGRO INTERERENCE SCREW DEVICE ANCHOR WAS BROKEN. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19731140 ↗
  • MalfunctionMay 22, 2024Break; Break

    IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT PROCEDURE, THE 5X12 MILAGRO INTERERENCE SCREW DEVICE ANCHOR WAS BROKEN OFF, REMOVED ALL THE BROKEN PARTS. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THE EXACT DATE OF THIS EVENT IS UNKNOWN. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19369746 ↗
  • MalfunctionApr 25, 2024Break; Break

    IT WAS REPORTED BY THE HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE ON (B)(6) 2024, IT WAS OBSERVED THAT THE 5X12 MILAGRO INTERFERENCE SCREW DEVICE ANCHOR WAS BROKEN OFF, REMOVED ALL THE BROKEN PARTS. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH A DELAY OF FIVE MINUTES. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19179455 ↗
  • MalfunctionMar 15, 2024Break; Break

    REPORT 2 OF 4 FOR (B)(4). IT WAS REPORTED BY SALES REP THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE ON (B)(6) 2024, IT WAS OBSERVED THAT A 8X15MM MILAGRO INTERFERENCE SCREW, 5X12MM MILAGRO INTERFERENCE SCREW, 8X23MM MILAGRO ADVANCE INTERFERENCE SCREW AND 10X23MM MILAGRO ADVANCE INTERFERENCE SCREW DEVICES WERE BROKEN. IT WAS REPORTED THAT THE PATIENT HAD EXTREMELY HARD BONE, AND IT WAS A REVISION CASE. IT WAS REPORTED THAT THE SAME LARGE TUNNELS WERE USED AND THE DOCTOR WANTED TO START WITH A LARGER SCREW BUT IT WOULD NOT DRIVE EVEN AFTER NOTCHING AND TAPPING FOUR TIMES. ACCORDING TO THE REPORTER, MULTIPLE SCREWS WERE TRIED, THEY ALL ENDED UP BREAKING AND THE BROKEN SCREW STAYED INSIDE THE PATIENT AS IT WASN'T BADLY BROKEN SINCE IT WAS STILL HOLDING THE GRAFT IN PLACE. IT WAS REPORTED THAT THE DOCTOR LEFT THE LEFT BROKEN SCREW INSIDE PATIENT AS IT HELD THE GRAFT BUT BROKE IN HALF. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18911215 ↗
  • MalfunctionMar 6, 2024Break; Break

    IT WAS REPORTED BY THE SALES REP IN CHINA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) RECONSTRUCTION SURGICAL PROCEDURE ON (B)(6) 2024 IT WAS OBSERVED THAT THE 5X12 MILAGRO INTERERENCE SCREW DEVICE WAS BROKEN OFF. IT WAS REPORTED THAT ALL BROKEN PARTS WERE REMOVED. IT WAS REPORTED THAT ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18846567 ↗
  • MalfunctionJan 23, 2024Break; Break

    IT WAS REPORTED FROM CHINA BY THE SALES REP THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE ON (B)(6) 2023 THE MILAGRO INTERERENCE SCREW DEVICE ANCHOR WAS BROKEN. ALL THE BROKEN PARTS WERE REMOVED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18563120 ↗

Other FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices

All FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE devices →