RIGIDLOOP 232447Recalls & reported problems
Medos International Sàrl · RIGIDLOOP Adjustable Cortical Implant STANDARD · Class 2 · Implant
All models of RIGIDLOOP (16)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Medos RIGIDLOOP adjustable implant is a fastener used to fix soft tissue during surgery. Of the 612 reports naming it, most describe the suture breaking when it is tightened, sometimes forcing surgeons to switch to a spare device. The reports include 30 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 612
- Deaths reported
- 0
- Injuries reported
- 30
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 612 adverse event reports that list this exact device (DI 10886705024094), received between Jan 16, 2018 and Sep 29, 2026 — classified as 30 injuries, 582 malfunctions.
The most frequently reported specific device problem is “Break” (526 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break526
- Material Frayed21
- Material Split, Cut or Torn21
- Adverse Event Without Identified Device or Use Problem9
- Loosening of Implant Not Related to Bone-Ingrowth9
- Migration8
- Detachment of Device or Device Component7
- Unintended Movement7
Patient problems
- No Clinical Signs, Symptoms or Conditions482
- No Consequences Or Impact To Patient71
- Not Applicable33
- Unspecified Tissue Injury14
- Tissue Damage6
- Discomfort4
- No Code Available4
- Injury2
Recent adverse event reports
The 25 most recent of 612 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 29, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50938199 ↗MalfunctionSep 29, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50938186 ↗MalfunctionSep 28, 2026Delivered as Unsterile Product; Packaging Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50923036 ↗MalfunctionSep 24, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50890315 ↗MalfunctionSep 24, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50890304 ↗MalfunctionAug 27, 2026Break
It was reported that during an Anterior Cruciate Ligament Reconstruction surgery the suture on the RGDLOOP ADJUSTABLE STND device broke off when it was tightened. It was not reported if there were any delays to the surgical procedure. It was reported that a spare device was used to complete the surgery. There were no reports of injuries, medical intervention or prolonged hospitalization. All available information has been disclosed.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50477525 ↗MalfunctionAug 27, 2026Break
It was reported that during an Anterior Cruciate Ligament Reconstruction surgery, the suture on the RGDLOOP ADJUSTABLE STND device broke when tightened. It was not reported if there were any delays to the surgical procedure. It was reported that a spare device was used to complete the surgery. There were no reports of injuries, medical intervention or prolonged hospitalization. All available information has been disclosed.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50474389 ↗MalfunctionAug 25, 2026Break
It was reported that during arthroscopic surgery using a RGDLOOP ADJUSTABLE STND System, it was observed that when the suture was tightened, it broke off. It was reported that another device was used to complete the procedure. It was reported that there was a 10-minute delay in the procedure due to the event. There were no adverse consequences to the patient. No additional information could be provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50437486 ↗MalfunctionAug 24, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50407548 ↗MalfunctionJul 29, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50074039 ↗MalfunctionJul 28, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50056273 ↗MalfunctionJul 28, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50051984 ↗MalfunctionJun 18, 2026Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ANTERIOR CRUCIATE LIGAMENT REPAIR PROCEDURE THE RGDLOOP ADJUSTABLE STND DEVICE SUTURE BROKE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25552347 ↗MalfunctionJun 12, 2026Break
IT WAS REPORTED THAT DURING AN UNSPECIFIED ARTHROSCOPIC PROCEDURE THE RGDLOOP ADJUSTABLE STND DEVICE SUTURE BROKE OFF. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25487407 ↗MalfunctionJun 11, 2026Break
IT WAS REPORTED THAT DURING AN UNSPECIFIED ARTHROSCOPIC PROCEDURE THE RGDLOOP ADJUSTABLE STND ANCHOR DEVICE WHEN TIGHTENED THE SUTURE, THE SUTURE WAS BROKE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25476580 ↗MalfunctionJun 11, 2026Break
IT WAS REPORTED THAT DURING ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION, THE RIGIDLOOP SUTURE WAS BROKEN OFF. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25469287 ↗MalfunctionMay 28, 2026Break
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGERY, WHEN TIGHTENED THE SUTURE OF THE RIGIDLOOP, THE SUTURE BROKE OFF. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25314333 ↗MalfunctionMay 27, 2026Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT (ACL) REPAIR SURGICAL PROCEDURE, WHILE USING THE RGDLOOP ADJUSTABLE STND SUTURE DEVICE, IT WAS OBSERVED THAT THE SUTURE WAS BROKEN OFF. IT WAS REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25304517 ↗MalfunctionMay 19, 2026Break
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE IT WAS OBSERVED THAT THE SUTURE OF THE RGDLOOP ADJUSTABLE STAND WAS TORN. THERE WERE NO REPORTED ADVERSE CONSEQUENCES TO THE PATIENT. THE EXACT DATE OF THE EVENT WAS NOT REPORTED; HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN (B)(6) OF 2026. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25219615 ↗MalfunctionMay 19, 2026Break
IT WAS REPORTED THAT DURING ACL (ANT.CRUCIATE LIGAMENT) SURGERY, THE RIGIDLOOP SUTURE DEVICE WAS BROKEN OFF. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25216535 ↗MalfunctionMay 12, 2026Break
IT WAS REPORTED THAT DURING ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION, WHEN THE SUTURE OF THE RIGIDLOOP WAS TIGHTENED, THE SUTURE BROKE OFF. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25147281 ↗MalfunctionMay 11, 2026Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ANTERIOR CRUCIATE LIGAMENT REPAIR PROCEDURE THE RGDLOOP ADJUSTABLE STND DEVICE SUTURE BROKE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THIS IS REPORT 3 OF 3 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25142461 ↗MalfunctionMay 4, 2026Break
THIS IS 1 OF 2 ASSOCIATED WITH (B)(4). IT WAS REPORTED THAT DURING THE KNEE ARTHROSCOPY SURGERY, THE RGDLOOP ADJUSTABLE STND SUTURE BROKEN OFF. THEN, THE 4.75 HEALIX ADVANCE KNTLSS BR ANCHOR BROKE OFF. ALL THE BROKEN PARTS WERE REMOVED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25067683 ↗MalfunctionMay 4, 2026Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY SURGERY, WHEN TIGHTENED THE SUTURE OF THE RIGIDLOOP, THE SUTURE WAS BROKEN OFF. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25066515 ↗MalfunctionApr 29, 2026Break
IT WAS REPORTED THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION SURGERY, WHEN TIGHTENED THE RIGIDLOOP SUTURE, THE SUTURE WAS BROKEN OFF. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25031093 ↗
FDA clearance and classification
Cleared through 510(k) K140324 on Jul 7, 2014 (decision: Substantially Equivalent). Submitted by Depuy Mitek, A Johnson and Johnson Company as “Rigidloop Adjustable Cortical Implant: Standard, Long And Xl”. The FDA review took 147 days. FDA record ↗
The FDA regulates product code MBI (fastener, fixation, nondegradable, soft tissue) under 21 CFR 888.3040 as a Class II device. 21 CFR 888.3040 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the RIGIDLOOP 232447
Has the RIGIDLOOP 232447 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886705024094), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the RIGIDLOOP 232447?
The FDA's MAUDE database lists 612 adverse event reports for this device, received between Jan 16, 2018 and Sep 29, 2026. In those reports, the most common specific device problems are “Break” (526), “Material Frayed” (21) and “Material Split, Cut or Torn” (21); the most common patient problems are “Not Applicable” (33), “Unspecified Tissue Injury” (14) and “Tissue Damage” (6). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the RIGIDLOOP 232447?
None of the reports are classified as deaths, 30 as injuries and 582 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the RIGIDLOOP 232447 FDA approved?
It was cleared through 510(k) K140324 on Jul 7, 2014. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the RIGIDLOOP 232447 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the RIGIDLOOP 232447, and is it still sold?
The FDA's device database lists Medos International Sàrl as the company responsible for it (brand RIGIDLOOP, model 232447). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE devices
- TRUESPAN ORTHOCORD 228151Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1,655 reports
- TRUESPAN ORTHOCORD 228152Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1,338 reports
- OMNISPAN ORTHOCORD 228142Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE893 reports
- FOOTPRINT 72202901Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE719 reports
- Arthrex® AR-3636Arthrex, Inc. · Fastener, fixation, nondegradable, soft tissue617 reports
- OMNISPAN ORTHOCORD 228141Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE599 reports
- TRUESPAN ORTHOCORD 228162Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE445 reports
- TWINFIX 72202595Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE443 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page