Skip to content

TRUESPAN ORTHOCORD 228162Recalls & reported problems

Medos International Sàrl · TRUESPAN Meniscal Repair System PLGA 24 DEGREES ORTHOCORD Violet braided composite suture Size 2-0, (2) PLGA Backstops, and (1) Applier with 24 DEGREES needle Depth Stop: 10mm-20mm Plus or Minus 1mm; Needle: 10mm Plus or Minus .13mm · Class 2 · Implant

All models of TRUESPAN ORTHOCORD (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medos TRUESPAN meniscal repair system is a kit used to repair torn knee cartilage with sutures. Many of its 451 reports describe parts that do not fit with other components, parts that break, and difficulty advancing or deploying implants. The reports list 4 injuries, no deaths and 447 malfunctions, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
460
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 460 adverse event reports that list this exact device (DI 10886705026050), received between Dec 9, 2019 and Sep 21, 2026 — classified as 4 injuries, 456 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (242 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 9 reportsDec 2023: 11 reportsJan 2024: 4 reportsFeb 2024: 5 reportsMar 2024: 11 reportsApr 2024: 3 reportsMay 2024: 6 reportsJun 2024: 5 reportsJul 2024: 0 reportsAug 2024: 7 reportsSep 2024: 8 reportsOct 2024: 19 reportsNov 2024: 10 reportsDec 2024: 8 reportsJan 2025: 9 reportsFeb 2025: 6 reportsMar 2025: 3 reportsApr 2025: 13 reportsMay 2025: 10 reportsJun 2025: 14 reportsJul 2025: 15 reportsAug 2025: 6 reportsSep 2025: 7 reportsOct 2025: 9 reportsNov 2025: 10 reportsDec 2025: 5 reportsJan 2026: 7 reportsFeb 2026: 9 reportsMar 2026: 6 reportsApr 2026: 9 reportsMay 2026: 11 reportsJun 2026: 9 reportsJul 2026: 0 reportsAug 2026: 6 reportsSep 2026: 11 reportsOct 2026: 0 reports20242025202619
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility242
  • Break118
  • Difficult to Advance34
  • Detachment of Device or Device Component33
  • Migration27
  • Premature Activation24
  • Material Deformation20
  • Physical Resistance/Sticking15

Patient problems

  • No Clinical Signs, Symptoms or Conditions438
  • No Consequences Or Impact To Patient20
  • Foreign Body In Patient5
  • Not Applicable2
  • Swelling/ Edema2
  • Abscess1
  • Hematoma1
  • Inflammation1

Recent adverse event reports

The 25 most recent of 460 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 21, 2026Difficult to Advance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50846835 ↗
  • MalfunctionSep 21, 2026Break; Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50839550 ↗
  • MalfunctionSep 21, 2026Use of Device Problem; Device-Device Incompatibility

    It was reported that during an unspecified surgical procedure, when the deploying the first anchor of the TRUESPAN 24 DEGREE PLGA anchor device, it was observed that the click of the release was heard, but no anchor or sutures were seen in the joint space. It was reported that the anchor did not fire, and the device did not work as intended. It was reported that a new anchor device was used and the device worked as intended. It was not reported if there was a delay in the procedure due to the event. It was reported that there was no harm to the patient or user. No additional information was provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50837646 ↗
  • MalfunctionSep 16, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50794680 ↗
  • MalfunctionSep 10, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50722998 ↗
  • MalfunctionSep 10, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50722996 ↗
  • MalfunctionSep 10, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50722328 ↗
  • MalfunctionSep 10, 2026Device-Device Incompatibility; Noise, Audible

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50717942 ↗
  • MalfunctionSep 10, 2026Device-Device Incompatibility; Noise, Audible

    It was reported that during an anterior cruciate ligament/meniscus surgery, the surgeon stated the deployment of the TRUESPAN ORTHOCORD implants felt "different" than the normal. It did not click like normal with deployment - the device either made a faint noise to question if the implant deployed or made a loud clunk. The implants deployed even with the unusual sounds; however, the suture loops appeared to be wound together which made tensioning more difficult than normal. The surgeon expressed that he was not sure if he was able to fully tension the implants, or if the suture was bunched together, causing laxity in the repair. There was no patient or user harm reported. There was no significant delay in surgery. All available information has been disclosed.Report 2 of 3 for (b)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50717935 ↗
  • MalfunctionSep 10, 2026Device-Device Incompatibility; Noise, Audible

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50717928 ↗
  • MalfunctionSep 4, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50615763 ↗
  • MalfunctionAug 28, 2026Crack

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50498970 ↗
  • MalfunctionAug 25, 2026Device-Device Incompatibility

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50438843 ↗
  • MalfunctionAug 20, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50355760 ↗
  • MalfunctionAug 14, 2026Difficult to Advance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50252677 ↗
  • MalfunctionAug 14, 2026Break; Difficult to Advance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50252674 ↗
  • MalfunctionAug 6, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50153828 ↗
  • MalfunctionJun 25, 2026Premature Activation

    IT WAS REPORTED THAT DURING THE MENISCAL REPAIR SURGERY, AFTER 1ST FIRING WITH TRUESPAN 24 DEGREE PLGA, TWO PLATES WERE DEPLOYED AT THE SAME TIME. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25620377 ↗
  • MalfunctionJun 24, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT DURING MEDIAL MENISCUS REPAIR SURGERY, TWO TRUESPAN ORTHOCORD IMPLANTS DID NOT FULLY DEPLOY. THE IMPLANTS WERE CUT AWAY AND REMOVED BY THE SURGEON AND A NEW IMPLANT INSERTED. THERE WAS A DELAY HOWEVER NO HARM TO PATIENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25610721 ↗
  • MalfunctionJun 23, 2026Device Damaged Prior to Receipt by User

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE TRUESPAN ORTHOCORD DEVICE WOULD NOT RELEASE. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE WAS DAMAGED OUT OF THE BOX. IT WAS NOT ABLE TO DEPLOY THE IMPLANTS. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25598819 ↗
  • MalfunctionJun 11, 2026Crack; Use of Device Problem

    IT WAS REPORTED THAT DURING A MENISCAL REPAIR SURGICAL PROCEDURE IT WAS OBSERVED THAT THE TRUESPAN 24 DEGREE PLGA DEVICE DID NOT WORK. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE DEPLOYER WITH THE WHITE SLEEVE CRACKED AT THE DISTAL END. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED; HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2026. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25474143 ↗
  • MalfunctionJun 10, 2026Premature Activation

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC MENISCAL REPAIR PROCEDURE THE TRUESPAN 24 DEGREE PLGA DEVICE AFTER 1ST FIRING (DEPLOYMENT), TWO PLATES WERE DEPLOYED AT THE SAME TIME. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25459902 ↗
  • MalfunctionJun 10, 2026Premature Activation

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC MENISCAL REPAIR PROCEDURE THE TRUESPAN 24 DEGREE PLGA DEVICE AFTER 1ST FIRING (DEPLOYMENT), TWO PLATES WERE DEPLOYED AT THE SAME TIME. CHANGED ANOTHER ONE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25459894 ↗
  • MalfunctionJun 10, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, THE TRUESPAN ORTHOCORD DEVICE NOT DEPLOYED CORRECTLY. IT WAS PULLED OUT OF THE MENISCUS AS THE DEVICE WAS PULLED BACK TO SET THE SECOND ANCHOR. IT HAPPENED WITH 3 TRUESPAN PLGA 24. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25459853 ↗
  • MalfunctionJun 10, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, THE TRUESPAN ORTHOCORD DEVICE NOT DEPLOYED CORRECTLY. IT WAS PULLED OUT OF THE MENISCUS AS THE DEVICE WAS PULLED BACK TO SET THE SECOND ANCHOR. IT HAPPENED WITH 3 TRUESPAN PLGA 24. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25459824 ↗

Other FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE devices

All FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE devices →

Sources and updates